We report a successful percutaneous revalvulation of a dysfunctional tricuspid bioprothesis in an 8-year-old child. Five years after implanting a 25-mm Carpentier-Edwards valve in the tricuspid position, the prosthesis showed significant dysfunction with clinical right heart failure. A 26-mm Edwards-Sapien XT inverted aortic valve was successfully implanted through a 19F sheath using a jugular approach. Such procedure can significantly postpone the need for surgical replacement of a biological valve. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:692 / 695
页数:4
相关论文
共 5 条
  • [1] Aupart MR, 2006, J HEART VALVE DIS, V15, P768
  • [2] Transcatheter Pulmonary Valve Implantation Using the Edwards SAPIEN™ Transcatheter Heart Valve
    Boone, Robert H.
    Webb, John G.
    Horlick, Eric
    Benson, Lee
    Cao, Qi-Ling
    Nadeem, N.
    Kiess, Maria
    Hijazi, Ziyad M.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 75 (02) : 286 - 294
  • [3] Tricuspid valve replacement:: An analysis of 25 years of experience at a single center
    Carrier, M
    Hébert, Y
    Pellerin, M
    Bouchard, D
    Perrault, LP
    Cartier, R
    Basmajian, A
    Pagé, P
    Poirier, NC
    [J]. ANNALS OF THORACIC SURGERY, 2003, 75 (01) : 47 - 50
  • [4] Pulmonary valve replacement: A comparison of three biological valves
    Fiore, Andrew C.
    Rodefeld, Mark
    Turrentine, Mark
    Vijay, Palaniswamy
    Reynolds, Tyler
    Standeven, John
    Hill, Kirstin
    Bost, Jamie
    Carpenter, Dustin
    Tobin, Courtney
    Brown, John W.
    [J]. ANNALS OF THORACIC SURGERY, 2008, 85 (05) : 1712 - 1718
  • [5] MITROFLOW PERICARDIAL BIOPROSTHESIS - CLINICAL-PERFORMANCE TO 10 YEARS
    POMAR, JL
    JAMIESON, WRE
    PELLETIER, LC
    GEREIN, AN
    CASTELLA, M
    BROWNLEE, RT
    [J]. ANNALS OF THORACIC SURGERY, 1995, 60 (02) : S305 - S310