High dose rate brachytherapy three-dimensional gel dosimetry using optical computed tomography readout

被引:3
作者
DeVries, D. A. [1 ]
Joshi, C. [1 ,2 ]
Schreiner, L. J. [1 ,2 ]
机构
[1] Queens Univ, Dept Phys, Kingston, ON K7L 3N6, Canada
[2] Kingston Hlth Sci Ctr, Canc Ctr Southeastern Ontario, Dept Med Phys, Kingston, ON K7L 5P9, Canada
来源
10TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE) | 2019年 / 1305卷
基金
加拿大自然科学与工程研究理事会;
关键词
ITERATIVE RECONSTRUCTION ALGORITHM; DISTRIBUTIONS;
D O I
10.1088/1742-6596/1305/1/012051
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The cause and removal of streaking artifacts from catheters within radiochromic 3D dosimeters was investigated to allow the use of these dosimeters for high dose rate (HDR) brachytherapy applications. The OSC-TV iterative reconstruction algorithm's ability to remove the streaking artifacts from one or more catheters was validated using external beam irradiations and leuco crystal violet (LCV) micelle gels. One and two catheter HDR plans were then delivered to LCV gels. Dose volumes from the gels were reconstructed with the OSC-TV algorithm, calibrated, and registered to treatment planning system dose volumes, and were shown to be artifact free and in agreement with the expected values.
引用
收藏
页数:7
相关论文
共 15 条
[1]  
Alexander K, J PHYS C SER, V573, P6
[2]   Radiochromic leuco dye micelle hydrogels: II. Low diffusion rate leuco crystal violet gel [J].
Babic, Steven ;
Battista, Jerry ;
Jordan, Kevin .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (22) :6791-6808
[3]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[4]   TIGRE: a MATLAB-GPU toolbox for CBCT image reconstruction [J].
Biguri, Ander ;
Dosanjh, Manjit ;
Hancock, Steven ;
Soleimani, Manuchehr .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (05)
[5]   Evaluation of an iterative reconstruction algorithm for optical CT radiation dosimetry [J].
Dekker, Kurtis H. ;
Battista, Jerry J. ;
Jordan, Kevin J. .
MEDICAL PHYSICS, 2017, 44 (12) :6678-6689
[6]   High-resolution gel dosimetry of a HDR brachytherapy source using normoxic polymer gel dosimeters: Preliminary study [J].
Hurley, C. ;
McLucas, C. ;
Pedrazzini, G. ;
Baldock, C. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 565 (02) :801-811
[7]   The use of gel dosimetry to measure the 3D dose distribution of a 90Sr/90Y intravascular brachytherapy seed [J].
Massillon-JL, G. ;
Minniti, R. ;
Mitch, M. G. ;
Maryanski, M. J. ;
Soares, C. G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06) :1661-1672
[8]   Evaluation of the OSC-TV iterative reconstruction algorithm for cone-beam optical CT [J].
Matenine, Dmitri ;
Mascolo-Fortin, Julia ;
Goussard, Yves ;
Despres, Philippe .
MEDICAL PHYSICS, 2015, 42 (11) :6376-6386
[9]   Measurement of ultrasonic attenuation coefficient in polymer gel dosimeters [J].
Mather, ML ;
Charles, PH ;
Baldock, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (20) :N269-N275
[10]   Polymer gel dosimetry close to an 125I interstitial brachytherapy seed [J].
Pantelis, E ;
Lymperopoulou, G ;
Papagiannis, P ;
Sakelliou, L ;
Stiliaris, E ;
Sandilos, P ;
Seimenis, I ;
Kozicki, M ;
Rosiak, JM .