A simplified poly(dimethylsiloxane) capillary electrophoresis microchip integrated with a low-noise contactless conductivity detector

被引:25
|
作者
Liu, Benyan [1 ,2 ,3 ,4 ]
Zhang, Yi [2 ,3 ]
Mayer, Dirk [2 ,3 ]
Krause, Hans-Joachim [2 ,3 ]
Jin, Qinghui [1 ,3 ]
Zhao, Jianlong [1 ,3 ]
Offenhaeusser, Andreas [2 ,3 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Microsyst & Informat Technol, State Key Lab Transducer Technol, Shanghai 200050, Peoples R China
[2] Forschungszentrum Julich, Inst Bioelect PGI 8, D-52425 Julich, Germany
[3] Collaborat CAS Shanghai, Joint Res Lab Supercond Bioelect, Shanghai, Peoples R China
[4] Chinese Acad Sci, Grad Sch, Shanghai 200050, Peoples R China
关键词
CE microchip; Contactless conductivity detection; Lock-in amplifier; PDMS; CHIP; ELECTRODES;
D O I
10.1002/elps.201000562
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A contactless conductivity detector integrated into a poly(dimethylsiloxane) microchip for electrophoresis is presented. It adopted the simplest configuration of electrodes commonly used in this detection mode for capillary electrophoresis microchips. Although the chip is based on a simple and effective design, it is able to obtain low detection levels due to the low noise of the detection circuit. A circuit based on a lock-in amplifier was designed on printed circuit boards to read out the signal. The property of the detection cell was studied by applying excitation signals of different frequencies and different amplitudes. It was found that the best detection limit could be achieved with a frequency of 50 kHz and amplitude of 20 V. The performance of the detector was demonstrated by successfully separating and detecting several inorganic ions and also a mixture of heavy metal ions. An average detection limit of 0.4 mu M was obtained for inorganic cations. This value is significantly improved compared to similar microchip-based detectors. The presented detector could be promising for mass production due to its properties, such as simple construction, high degree of integration, high performance and low cost.
引用
收藏
页码:699 / 704
页数:6
相关论文
共 50 条
  • [1] A simple poly(dimethylsiloxane) electrophoresis microchip with an integrated contactless conductivity detector
    Liu, Benyan
    Jin, Qinghui
    Zhang, Yi
    Mayer, Dirk
    Krause, Hans-Joachim
    Zhao, Jianlong
    Offenhaeusser, Andreas
    MICROCHIMICA ACTA, 2011, 172 (1-2) : 193 - 198
  • [2] A simple poly(dimethylsiloxane) electrophoresis microchip with an integrated contactless conductivity detector
    Benyan Liu
    Qinghui Jin
    Yi Zhang
    Dirk Mayer
    Hans-Joachim Krause
    Jianlong Zhao
    Andreas Offenhäusser
    Microchimica Acta, 2011, 172 : 193 - 198
  • [3] Contactless conductivity detector for microchip capillary electrophoresis
    Pumera, M
    Wang, J
    Opekar, F
    Jelínek, I
    Feldman, J
    Löwe, H
    Hardt, S
    ANALYTICAL CHEMISTRY, 2002, 74 (09) : 1968 - 1971
  • [4] Design of Movable Contactless Conductivity Detector for Microchip Capillary Electrophoresis
    Zhang, Hai-feng
    Liui, Xiao-wei
    Peng, Zhi-cheng
    Wang, Xi-lian
    Wang, Wei
    Chen, Yu-feng
    2009 4TH IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS, VOLS 1 AND 2, 2009, : 190 - +
  • [5] Screen-Printed Contactless Conductivity Detector for Microchip Capillary Electrophoresis
    Wang, Joseph
    Chen, Gang
    Chatrathi, Madhu Prakash
    Wang, Mei
    Rinehart, Robert
    Muck, Alexander
    ELECTROANALYSIS, 2008, 20 (22) : 2416 - 2421
  • [6] Contactless conductivity detector integrated in silicon-based electrophoresis microchip
    Gu, Wen-Wen
    Wen, Zhi-Yu
    Wen, Zhong-Quan
    Xu, Yi
    Liang, Feng-Fei
    Liu, Hai-Tao
    Nami Jishu yu Jingmi Gongcheng/Nanotechnology and Precision Engineering, 2010, 8 (04): : 362 - 367
  • [7] Application of an external contactless conductivity detector for the analysis of beverages by microchip capillary electrophoresis
    Kubán, P
    Hauser, PC
    ELECTROPHORESIS, 2005, 26 (16) : 3169 - 3178
  • [8] Integrated Capacitively Coupled Contactless Conductivity Detector for Capillary Electrophoresis
    Wang Lixin
    Fu Chonggang
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOL. 3, 2008, : 1493 - 1495
  • [9] Measurement of Nitrogen Mustard Degradation Products by Poly(dimethylsiloxane) Microchip Electrophoresis with Contactless Conductivity Detection
    Ding, Yongsheng
    Rogers, Kim
    ELECTROANALYSIS, 2008, 20 (20) : 2192 - 2198
  • [10] Low-noise amperometric detector for use in capillary electrophoresis
    Fu, CG
    Wang, LX
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 1999, 27 (03) : 199 - 205