Homogeneous assay based on anti-Stokes' shift time-resolved fluorescence resonance energy-transfer measurement

被引:25
|
作者
Laitala, V [1 ]
Hemmila, L [1 ]
机构
[1] Wallac Oy, PerkinElmer Life & Analyt Sci, FIN-20101 Turku, Finland
关键词
D O I
10.1021/ac048414o
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report here a novel, time-resolved, lanthanide-based energy-transfer assay utilizing nonoverlapping acceptor fluorophores, which have their absorption energetically at a higher level than the emittive transitions of the donor. The technique was studied by comparing a series of nonoverlapping acceptors in a homogeneous DNA model assay utilizing Eu3+ chelate as a donor. The assay provides strong energy-transfer enhanced acceptor emission and enables the anti-Stokes' shift FRET measurement, in which the induced acceptor emission is at shorter wavelength than the donor emission. This results in high sensitivity, and 0.8 pM detection limit was measured for the DNA target. The acceptor signal of the assay is characterized by exceptional lifetime properties and is not strictly following the Forster's theory. The mechanism of nonoverlapping energy transfer is considered, and we propose that when nonoverlapping acceptors are utilized, the energy transfer arises from the upper D-5(2) and D-5(1), excited states of europium. The assumption was studied using a simplified energy level scheme of the Eu3+ donor and the acceptors, and a correlation between the acceptor emission behavior and the energy level scheme was found.
引用
收藏
页码:1483 / 1487
页数:5
相关论文
共 50 条
  • [1] Homogeneous assay based on low quantum yield Sm(III)-donor and anti-Stokes' shift time-resolved fluorescence resonance energy-transfer measurement
    Laitala, V
    Hemmilä, L
    ANALYTICA CHIMICA ACTA, 2005, 551 (1-2) : 73 - 78
  • [2] Simultaneous use of time-resolved fluorescence and anti-stokes photoluminescence in a bioaffinity assay
    Kuningas, K
    Rantanen, T
    Karhunen, U
    Lövgren, T
    Soukka, T
    ANALYTICAL CHEMISTRY, 2005, 77 (09) : 2826 - 2834
  • [3] TIME-RESOLVED FLUORESCENCE ENERGY-TRANSFER
    VANDERMEER, BW
    CHEN, SY
    BIOPHYSICAL JOURNAL, 1993, 64 (02) : A242 - A242
  • [4] A homogeneous time-resolved fluorescence resonance energy transfer assay for phosphatidylserine exposure on apoptotic cells
    Gasser, Jean-Philippe
    Hehl, Michaela
    Millward, Thomas A.
    ANALYTICAL BIOCHEMISTRY, 2009, 384 (01) : 49 - 55
  • [5] A homogeneous europium cryptate-based assay for the diagnosis of mutations by time-resolved fluorescence resonance energy transfer
    Lopez-Crapez, E
    Bazin, H
    Andre, E
    Noletti, J
    Grenier, J
    Mathis, G
    NUCLEIC ACIDS RESEARCH, 2001, 29 (14) : 70
  • [6] A flow injection kinase assay system based on time-resolved fluorescence resonance energy-transfer detection in the millisecond range
    Hirata, J
    de Jong, CF
    van Dongen, MM
    Buijs, J
    Ariese, F
    Irth, H
    Gooijer, C
    ANALYTICAL CHEMISTRY, 2004, 76 (15) : 4292 - 4298
  • [7] A homogeneous G protein-coupled receptor ligand binding assay based on time-resolved fluorescence resonance energy transfer
    Hu, Liaoyuan A.
    Zhou, Tian
    Hamman, Brian D.
    Liu, Qingyun
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2008, 6 (04) : 543 - 550
  • [8] Homogeneous Time-resolved Forster Resonance Energy Transfer-based Assay for Detection of Insulin Secretion
    Aslanoglou, Despoina
    George, Emily W.
    Freyberg, Zachary
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2018, (135):
  • [9] Use of homogeneous time-resolved fluorescence energy transfer in the measurement of nuclear receptor activation
    Zhou, GC
    Cummings, R
    Hermes, J
    Moller, DE
    METHODS, 2001, 25 (01) : 54 - 61
  • [10] Homogeneous time-resolved IL-2-IL-2Rα assay using fluorescence resonance energy transfer
    Stenroos, K
    Hurskainen, P
    Eriksson, S
    Hemmila, I
    Blomberg, K
    Lindqvist, C