Assessment of intraocular pressure in chinchillas of different age groups using rebound tonometry

被引:2
|
作者
Yokoay Claros Chacaltana, Flor Diana [1 ]
Tadeu Pigatto, Joao Antonio [1 ]
Denardin, Ione Terezinha [2 ]
机构
[1] Univ Fed Rio Grande do Sul, Fac Vet, Dept Med Anim, Ave Bento Goncalves 9090, BR-91540000 Porto Alegre, RS, Brazil
[2] Univ Fed Santa Maria, Colegio Politecn, Santa Maria, RS, Brazil
来源
CIENCIA RURAL | 2016年 / 46卷 / 08期
关键词
Chinchillas; Chinchilla lanigera; intraocular pressure; tonometry; TonoVet; OPHTHALMIC EXAMINATION FINDINGS; APPLANATION TONOMETER; REFERENCE VALUES; TONOPEN XL(R); BODY LENGTH; TONOVET(R); COLONY; EYE; ANESTHESIA; THICKNESS;
D O I
10.1590/0103-8478cr20151206
中图分类号
S3 [农学(农艺学)];
学科分类号
0901 ;
摘要
The aim of this research was to measure the intraocular pressure (IOP) of normal chinchilla eyes using the rebound tonometer. A further aim was to assess whether there were differences in the values of intraocular pressure in relation to animals age, gender and time of day. Thirty-six chinchillas were divided into three groups of 12 chinchillas each, by age: Group I (2-6-month-old), Group II (20 and 34 months) and Group III (37 and 135 months). Ophthalmic examination was performed previously by Schirmer tear test, slit lamp biomicroscopy, indirect ophthalmoscopy and fluorescein test in all chinchillas. Three measurements of intraocular pressure were assessed on the same day (7, 12 and 19h). Tonometry was performed on both eyes using the rebound tonometer after calibration in "p" mode. Statistical analysis was performed with SigmaPlot for Windows. The mean IOP for groups I, II and III were 2.47 +/- 0.581mmHg, 2.47 +/- 0.581mmHg and 2.51 +/- 0.531mmHg, respectively. No significant differences were reported between age and IOP and no significant differences were reported between the time of day and IOP. The IOP in chinchillas did not differ significantly between genders or ages of the animals, and did not change with time of day.
引用
收藏
页码:1466 / 1471
页数:6
相关论文
共 50 条
  • [1] Determination of Intraocular Pressure in Clinically Healthy Turkish Eastern Anatolian Red Cattle of Different Age Groups Using Rebound Tonometry
    Yakan, S.
    Isler, C. T.
    Denk, H.
    ISRAEL JOURNAL OF VETERINARY MEDICINE, 2021, 76 (04): : 150 - 155
  • [2] ASSESSMENT OF INTRAOCULAR PRESSURE BY APPLANATION AND REBOUND TONOMETRY IN GUINEA PIGS OF DIFFERENT AGES
    Cairo, Marti
    Teresa Pena, Maria
    Rios, Jose
    Melero, Adria
    Martorell, Jaume
    Leiva, Marta
    JOURNAL OF EXOTIC PET MEDICINE, 2018, 27 (01) : 25 - 31
  • [3] Intraocular Pressure Measurements Using Rebound Tonometry in Eight Different Species of Companion Birds
    Karimi, Vahid
    Asadi, Faezeh
    Rajaei, Seyed Mehdi
    Golabdar, Salar
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2020, 34 (04) : 338 - 342
  • [4] Assessing the Reliability of Intraocular Pressure Measurements Using Rebound Tonometry
    Realini, Tony
    McMillan, Brian
    Gross, Ronald L.
    Devience, Eva
    Balasubramani, Goundappa K.
    JOURNAL OF GLAUCOMA, 2021, 30 (08) : 629 - 633
  • [5] Validation of rebound tonometry for intraocular pressure measurement in the rabbit
    Zhang, Hong
    Yang, Dong
    Ross, Craig M.
    Wigg, Jonathan P.
    Pandav, Surinder
    Crowston, Jonathan G.
    EXPERIMENTAL EYE RESEARCH, 2014, 121 : 86 - 93
  • [6] Technical Note:: A comparison of central and peripheral intraocular pressure using rebound tonometry
    Queiros, A.
    Gonzalez-Meijome, J. M.
    Fernandes, P.
    Jorge, J.
    Montes-Mico, R.
    Almeida, J. B.
    Parafita, M. A.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2007, 27 (05) : 506 - 511
  • [7] Intraocular Pressure Measured by Rebound Tonometry in Seven Raptor Species
    Yonez, Muhammed Kaan
    Erol, M.
    Erol, H.
    Atalan, G.
    Dogan, Z.
    Kibar, M.
    JOURNAL OF THE HELLENIC VETERINARY MEDICAL SOCIETY, 2022, 73 (02): : 3929 - 3934
  • [8] Intraocular pressure with rebound tonometry and effects of topical intraocular pressure reducing medications in guinea pigs
    Di, Yue
    Luo, Xiu-Mei
    Qiao, Tong
    Lu, Na
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2017, 10 (02) : 186 - 190
  • [9] Intraocular pressure with rebound tonometry and effects of topical intraocular pressure reducing medications in guinea pigs
    Yue Di
    Xiu-Mei Luo
    Tong Qiao
    Na Lu
    International Journal of Ophthalmology, 2017, 10 (02) : 186 - 190
  • [10] Monitoring intraocular pressure changes after intravitreal Ranibizumab injection using rebound tonometry
    Fuest, Matthias
    Kotliar, Konstantin
    Walter, Peter
    Plange, Niklas
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2014, 34 (04) : 438 - 444