Aerosol delivery through nasal cannulas:: An in vitro study

被引:94
作者
Bhashyam, Abhiram R. [1 ]
Wolf, Matthew T. [2 ]
Marcinkowski, Amy L. [2 ]
Saville, Al [3 ]
Thomas, Kristina [1 ]
Carcillo, Joseph A. [3 ]
Corcoran, Timothy E. [1 ,2 ]
机构
[1] Univ Pittsburgh, Pittsburgh, PA USA
[2] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA USA
[3] UPMC, Childrens Hosp Pittsburgh, Pittsburgh, PA USA
关键词
bench testing; nebulizer; cannula;
D O I
10.1089/jamp.2007.0662
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
In most circumstances, a nasal route for the delivery of pulmonary aerosol medications is rarely considered; however, in specific instances, this route may be quite useful. Consider, for example, the delivery of aerosol treatments during humidified high-flow nasal cannula use in pediatric critical care, or continuous aerosol delivery via cannula for medications with short durations of action. The goal of this study was to evaluate the potential for delivering aerosols via nasal cannula through in vitro studies of aerosol output and size. The system utilized for testing included an Aerogen Solo nebulizer downstream of a heater/humidifier system, followed by a nasal cannula and an aerosol collection apparatus. Adult, pediatric, and infant cannulas were tested with and without an inhalation-only breathing simulator. The cannulas were driven by 3 1pm (50 psig) oxygen flows. Dose quantification was performed using radioisotope techniques. Total cannula output and system losses were measured. Aerosol size measurements were made from the nebulizer, from the heating tube, and from the prongs of the adult and pediatric cannulas, using laser-diffraction techniques. Total cannula output ranged from 8.4-25.1% and 18.6-26.9% of loaded dose, without and with the addition of inhalation flows. Volume median diameters were 2.2 +/- 0.2 mu m from the adult cannula and 1.9 +/- 0.3 mu m from the pediatric cannula. Ninety percent of the aerosol volume was in sizes smaller than 4.2 +/- 0.4 mu m (adult) and 3.8 +/- 0.5 mu m (pediatric). System losses were highest in the nebulizer-humidifier connectors, heated tube, and humidifier. Losses in the nebulizer were very low (2.2-3.5%). This study demonstrates that aerosols can be efficiently delivered through a humidified high-flow nasal cannula system. Further study is required to determine if this route is viable for pulmonary delivery.
引用
收藏
页码:181 / 187
页数:7
相关论文
共 18 条
  • [1] Use of high flow nasal cannula on a pediatric burn patient with inhalation injury and post-extubation stridor
    Byerly, FL
    Haithcock, JA
    Buchanan, IB
    Short, KA
    Cairns, BA
    [J]. BURNS, 2006, 32 (01) : 121 - 125
  • [2] Nasal continuous positive airway pressure from high flow cannula versus Infant Flow for preterm infants
    Campbell, D. M.
    Shah, P. S.
    Shah, V.
    Kelly, E. N.
    [J]. JOURNAL OF PERINATOLOGY, 2006, 26 (09) : 546 - 549
  • [3] Improving drug delivery from medical nebulizers: The effects of increased nebulizer flow rates and reservoirs
    Corcoran, TE
    Dauber, JH
    Chigier, N
    Iacono, AT
    [J]. JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2002, 15 (03): : 271 - 282
  • [4] Comparing two methods of delivering high-flow gas therapy by nasal cannula following endotracheal extubation: a prospective, randomized, masked, crossover trial
    D Woodhead, D.
    Lambert, D. K.
    Clark, J. M.
    Christensen, R. D.
    [J]. JOURNAL OF PERINATOLOGY, 2006, 26 (08) : 481 - 485
  • [5] Pharyngeal pressure in preterm infants receiving nasal continuous positive airway pressure
    De Paoli, AG
    Lau, R
    Davis, PG
    Morley, CJ
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2005, 90 (01): : F79 - F81
  • [6] High-flow nasal cannula versus nasal continuous positive airway pressure in the management of apnea of prematurity
    Dutta, S
    [J]. PEDIATRICS, 2002, 109 (04) : 718 - 719
  • [7] Ferron G.A., 1989, J. Aerosol Med., V2, P271, DOI [DOI 10.1089/JAM.1989.2.271, 10.1089/jam.1989.2.271]
  • [8] DEPOSITION OF PARTICLES IN THE HUMAN RESPIRATORY-TRACT IN THE SIZE RANGE 0.005-15-MU-M
    HEYDER, J
    GEBHART, J
    RUDOLF, G
    SCHILLER, CF
    STAHLHOFEN, W
    [J]. JOURNAL OF AEROSOL SCIENCE, 1986, 17 (05) : 811 - 825
  • [9] MATHEMATICAL-MODEL FOR THE POSTNATAL-GROWTH OF THE HUMAN-LUNG
    HOFMANN, W
    [J]. RESPIRATION PHYSIOLOGY, 1982, 49 (01): : 115 - 129
  • [10] Heated humidified high-flow nasal cannula: use and a neonatal early extubation protocol
    Holleman-Duray, D.
    Kaupie, D.
    Weiss, M. G.
    [J]. JOURNAL OF PERINATOLOGY, 2007, 27 (12) : 776 - 781