Determination of biochemical species on electrophoresis chips with an external contactless conductivity detector

被引:53
|
作者
Abad-Villar, EM [1 ]
Kubán, P [1 ]
Hauser, PC [1 ]
机构
[1] Univ Basel, Dept Chem, CH-4004 Basel, Switzerland
关键词
amino acids; contactless conductivity detection; DNA; electrophoresis chip; enzymatic digestion; immunoglobulin G; peptides; peptide map; proteins; proteomics; systems biology;
D O I
10.1002/elps.200500149
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Contactless conductivity measurements were found to be suitable for the direct detection, i.e., without needing any labels, of a range of biochemically relevant species, namely amino acids, peptides, proteins, immunoglobulin, and DNA. It was also possible to monitor the products of the enzymatic digestion of HSA with pepsin. Detection was carried out on bare electrophoresis chips made from poly(methyl methacrylate) by probing the conductivity in the channel with a pair of external electrodes, which are fixed on the chip holder. Separation efficiencies up to 15 000 plates could be obtained and LODs are in the low mu m-range, except for immunoglobulin G (IgG) which could be determined down to 0.4 nM. Linear dynamic ranges of two to three orders of magnitude were obtained for the peptides as examples.
引用
收藏
页码:3609 / 3614
页数:6
相关论文
共 50 条
  • [31] Determination of creatine and phosphocreatine in muscle biopsy samples by capillary electrophoresis with contactless conductivity detection
    See, Hong Heng
    Schmidt-Marzinkowski, Julia
    Pormsila, Worapan
    Morand, Rejane
    Kraehenbuehl, Stephan
    Hauser, Peter C.
    ANALYTICA CHIMICA ACTA, 2012, 727 : 78 - 82
  • [32] Determination of heavy metal ions by microchip capillary electrophoresis coupled with contactless conductivity detection
    Liu, Benyan
    Zhang, Yi
    Mayer, Dirk
    Krause, Hans-Joachim
    Jin, Qinghui
    Zhao, Jianlong
    Offenhaeusser, Andreas
    Xu, Yuansen
    ELECTROPHORESIS, 2012, 33 (08) : 1247 - 1250
  • [33] Determination of artificial sweeteners by capillary electrophoresis with contactless conductivity detection optimized by hydrodynamic pumping
    Stojkovic, Marko
    Thanh Duc Mai
    Hauser, Peter C.
    ANALYTICA CHIMICA ACTA, 2013, 787 : 254 - 259
  • [34] Towards a radiation-tolerant contactless conductivity detector for use with capillary electrophoresis systems in spaceflight applications
    Santos, M. S. Ferreira
    Metz, B. C.
    da Costa, E. T.
    do Lago, C. L.
    Willis, P. A.
    Mora, M. F.
    Noell, A. C.
    ACTA ASTRONAUTICA, 2022, 190 : 299 - 307
  • [35] Frequency-Tuned Contactless Conductivity Detector
    Tuma, Petr
    XXXVII MODERNI ELEKTROCHEMICKE METODY, 2017, : 239 - 243
  • [36] A dual spectrophotometric/contactless conductivity detector for CE determination of incompletely separated amino acids
    Zikmundova, Jana
    Tuma, Petr
    Opekar, Frantisek
    JOURNAL OF SEPARATION SCIENCE, 2008, 31 (02) : 353 - 355
  • [37] Capillary and microchip electrophoresis with contactless conductivity detection for analysis of foodstuffs and beverages
    Tuma, Petr
    Opekar, Frantisek
    Dlouhy, Pavel
    FOOD CHEMISTRY, 2022, 375
  • [38] A simplified poly(dimethylsiloxane) capillary electrophoresis microchip integrated with a low-noise contactless conductivity detector
    Liu, Benyan
    Zhang, Yi
    Mayer, Dirk
    Krause, Hans-Joachim
    Jin, Qinghui
    Zhao, Jianlong
    Offenhaeusser, Andreas
    ELECTROPHORESIS, 2011, 32 (6-7) : 699 - 704
  • [39] Application of capillary electrophoresis with capacitively contactless conductivity detection for biomedical analysis
    Elbashir, Abdalla A.
    Osman, Abdelbagi
    Elawad, Mohammed
    Ziyada, Abobakr K.
    Aboul-Enein, Hassan Y.
    ELECTROPHORESIS, 2024, 45 (5-6) : 400 - 410
  • [40] Dual detector with fluorescence/contactless conductometry for capillary electrophoresis
    Yang, BC
    Tan, F
    Guan, YF
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (05) : 740 - 742