Robust design for acetabular cup stability accounting for patient and surgical variability

被引:10
|
作者
Ong, Kevin L. [1 ]
Santner, Thomas J. [2 ]
Bartel, Donald L. [3 ]
机构
[1] Exponent Inc, Philadelphia, PA 19104 USA
[2] Ohio State Univ, Dept Stat, Columbus, OH 43210 USA
[3] Cornell Univ, Sibley Sch Mech & Aerosp Engn, Ithaca, NY 14853 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2008年 / 130卷 / 03期
基金
美国国家科学基金会;
关键词
robust analysis; total hip arthroplasty; acetabular cup; stability; cementless;
D O I
10.1115/1.2907764
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The stability of cementless acetabular cups depends on a close fit between the components and reamed acetabular cavities to promote bone ingrowth. Cup performance and stability are affected by both design and environmental (patient-dependent and surgical) factors. This study used a statistically based metamodel to determine the relative influences of design and environmental factors on acetabular cup stability by incorporating a comprehensive set of patient-dependent and surgical variables. Cup designs with 2 mm or 3 mm intended equatorial bone-implant interferences appeared to perform the best, improving implant stability with smaller mean and variability in cup relative motions and greater mean and smaller variability in ingrowth areas. Cup eccentricity was found to have no effect on implant performance. Design variables did not contribute as much to the variation in performance measures compared to the environmental variables, except for potential ingrowth areas.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] The effect of dorsal rim loss on the initial stability of the BioMedtrix cementless acetabular cup
    Meredith L Montgomery
    Stanley E Kim
    Jonathan Dyce
    Antonio Pozzi
    BMC Veterinary Research, 11
  • [22] Initial stability of a highly porous titanium cup in an acetabular bone defect model
    Yoshimoto, Kensei
    Nakashima, Yasuharu
    Wakiyama, Miyo
    Hara, Daisuke
    Nakamura, Akihiro
    Iwamoto, Mikio
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2018, 23 (04) : 665 - 670
  • [23] Effects of Eccentric Screwing on the Initial Stability of the Acetabular Cup in Bone of Different Qualities
    Hsu, Jui-Ting
    Huang, Heng-Li
    Tsai, Ming-Tzu
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 231 - 234
  • [24] The number of screws, bone quality, coefficient affect acetabular cup and friction stability
    Hsu, Jui-Ting
    Chang, Chih-Han
    Huang, Heng-Li
    Zobitz, Mark E.
    Chen, Weng-Pin
    Lai, Kuo-An
    An, Kai-Nan
    MEDICAL ENGINEERING & PHYSICS, 2007, 29 (10) : 1089 - 1095
  • [25] Use of a revision cup for treatment of Zurich cementless acetabular cup loosening Surgical technique and clinical application in 31 cases
    Vezzoni, L.
    Montinaro, V.
    Vezzoni, A.
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2013, 26 (05) : 408 - 415
  • [26] In silico analysis of the patient-specific acetabular cup anteversion safe zone
    Aubert, Thomas
    Gerard, Philippe
    Galanzino, Giacomo
    Marmor, Simon
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2024, 110 (06)
  • [27] Ex Vivo Evaluation of Cementless Acetabular Cup Stability Using Impact Analyses with a Hammer Instrumented with Strain Sensors
    Tijou, Antoine
    Rosi, Giuseppe
    Hernigou, Philippe
    Flouzat-Lachaniette, Charles-Henri
    Haiat, Guillaume
    SENSORS, 2018, 18 (01):
  • [28] Reconstruction of Paprosky type IIIB acetabular bone defects using a cup-on-cup technique: A surgical technique and case series
    Du, Yin-Qiao
    Liu, Yu-Ping
    Sun, Jing-Yang
    Ni, Ming
    Zhou, Yong-Gang
    WORLD JOURNAL OF CLINICAL CASES, 2020, 8 (07) : 1223 - 1231
  • [29] Reconstruction of Paprosky type ⅢB acetabular bone defects using a cup-on-cup technique: A surgical technique and case series
    Yin-Qiao Du
    Yu-Ping Liu
    Jing-Yang Sun
    Ming Ni
    Yong-Gang Zhou
    World Journal of Clinical Cases, 2020, (07) : 1223 - 1231
  • [30] Design considerations for patient-specific surgical templates for total hip arthroplasty with respect to acetabular cartilage
    Mueller, Samuel
    Ahmad, Intsar
    Kraemer, Manuel
    Utz, Michael
    Gaa, Johannes
    Kahrs, Lueder A.
    Ortmaier, Tobias