Validity and limits of the rebound tonometer (ICare®): clinical study

被引:44
作者
Scuderi, Gian Luca [1 ]
Cascone, Nikhil Carlo [1 ]
Regine, Federico [2 ]
Perdicchi, Andrea [1 ]
Cerulli, Angelica [2 ]
Recupero, Santi Maria [1 ]
机构
[1] Univ Roma La Sapienza, St Andrea Hosp, Sch Med 2, Dept Ophthalmol, Rome, Italy
[2] Univ Roma Tor Vergata, Dept Biopathol & Diagnost Imaging, Rome, Italy
关键词
Glaucoma; ICare (R); Intraocular pressure; Rebound tonometry; MEASURING INTRAOCULAR-PRESSURE; GOLDMANN APPLANATION TONOMETER; CORNEAL THICKNESS; TONOPEN-XL; AGREEMENT; IMPACT;
D O I
10.5301/EJO.2010.3712
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. The aim of this study was to evaluate the measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare (R)) compared with Goldmann applanation tonometry (GAT). We also aimed to quantify the systematic and random errors (bias) of the 2 methods, to evaluate the sensitivity and specificity of the ICare (R) tonometer in identifying patients with 21 mmHg or more measured with the GAT, and to study the influence of corneal thickness on IOP measurement with the 2 tonometers. Methods. We compared the IOP values obtained with the 2 instruments in 97 patients. Results. Analysis based on the Bland and Altman method revealed that the IOP values recorded with the ICare (R) tonometer were slightly higher than those obtained with the GAT. The estimated bias for right eye measurements was 0.78 mmHg with 95% limits of agreement +/- 3.55 mmHg. This overestimation, which is not clinically relevant, was confirmed when we used the IOP values corrected according to central corneal thickness for data analysis. The sensitivity and specificity were 0.90 and 0.95, respectively. Conclusions. The ICare (R) tonometer proved to be comparable with other nonconventional tonometers and can be used by nonophthalmologists and paramedical personnel during screening tests of populations. In addition, the ICare (R) tonometer could be considered a valid alternative to GAT when GAT is not available.
引用
收藏
页码:251 / 257
页数:7
相关论文
共 26 条
[1]  
Bhan A, 2002, INVEST OPHTH VIS SCI, V43, P1389
[2]   Applying the right statistics: analyses of measurement studies [J].
Bland, JM ;
Altman, DG .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (01) :85-93
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[5]   Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure [J].
Cervino, A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (12) :1444-1446
[6]   Method for the noninvasive measurement of intraocular pressure in mice [J].
Danias, J ;
Kontiola, AI ;
Filippopoulos, T ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (03) :1138-1141
[7]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209
[8]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[9]   Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings [J].
Feltgen, N ;
Leifert, D ;
Funk, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (01) :85-87
[10]   Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population [J].
Fernandes, P ;
Díaz-Rey, JA ;
Queirós, A ;
Gonzalez-Meijome, JM ;
Jorge, J .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) :436-440