摘要
An implant fracture is an unusual reason for total hip arthroplasty (THA) revision; however, it may cause serious clinical complications. The stress in femoral stems may increase considerably with the loss of bone support around the implant, leading to implant failure. This paper investigates the main reasons for the in vivo fracture of a Ti6Al4V revision femoral ZMR stem. Various methods were used: examination of the patient's by radiographies; optical emission spectrometry; and Scanning Electron Microscopy (SEM). The titanium (Ti) alloy was also characterized through optical microscopy and a Vickers hardness test was carried out. A finite element analysis (FEA) was conducted to evaluate the stress distribution occurring in the femoral stem revision under loading. The chemical composition and microstructure of the prosthesis did not reveal any non-conformity. The SEM analysis revealed fretting damage near the crack initiation site and the fractured surfaces exhibited characteristic ratchet marks, poorly formed striations and a small final fracture area. The FEA showed a stress concentration located at the proximal-distal taper junction, with a smaller diameter, coinciding with the crack initiation site. The conclusion was that fatigue failure was due to a lack of bone sustainability that increased the stress, associated with design characteristics that caused a localized concentration of stress, and these were potentialized by a fretting mechanism, leading to the failure.
引用
收藏
页数:10
相关论文
共 50 条
[31]   Epidemiology of Revision Total Hip Arthroplasty: An Indian Experience [J].
Mittal, Gaurav ;
Kulshrestha, Vikas ;
Kumar, Santhosh ;
Datta, Barun .
INDIAN JOURNAL OF ORTHOPAEDICS, 2020, 54 (05) :608-615
[32]   Revision Total Hip Arthroplasty in Jehovah's Witnesses [J].
Harwin, Steven F. ;
Pivec, Robert ;
Johnson, Aaron J. ;
Naziri, Qais ;
Mont, Michael A. .
ORTHOPEDICS, 2012, 35 (08) :E1145-E1151
[33]   Deep rolling of titanium rods for application in modular total hip arthroplasty [J].
Schuh, Alexander ;
Zeller, Christian ;
Holmarth, Ulrich ;
Kachler, Werner ;
Wilcke, Gerhard ;
Zeiler, Guenther ;
Eigenmann, Bernd ;
Bigoney, Jean .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (02) :330-335
[34]   Patient-reported Outcome Measures of Revision Total Hip Arthroplasty for Prosthetic Joint Infection is not Inferior to Aseptic Revision Total Hip Arthroplasty [J].
Lim, J. B. T. ;
Pang, H. N. ;
Tay, K. J. D. ;
Chia, S. L. ;
Yea, S. J. ;
Lo, N. N. .
MALAYSIAN ORTHOPAEDIC JOURNAL, 2020, 14 (03) :73-81
[35]   15-year survivorship of a unique dual-modular femoral stem in primary hip arthroplasty [J].
Scott, David F. ;
Eppich, Kade ;
Mehic, Edin ;
Gray, Celeste ;
Smith, Crystal Lederhos ;
Johnston, Michael .
BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
[36]   Bimodular femoral stems in primary total hip arthroplasty [J].
Zajc, Jan ;
Fokter, Samo K. .
EXPERT REVIEW OF MEDICAL DEVICES, 2023, 20 (12) :1051-1064
[37]   Risk Factors for Subsidence of Modular Fluted Tapered Stems Used During Revision Total Hip Arthroplasty for Periprosthetic Hip Fractures [J].
Parry, Joshua A. ;
Hernandez, Nicolas M. ;
Berry, Daniel J. ;
Abdel, Matthew P. ;
Yuan, Brandon J. .
JOURNAL OF ARTHROPLASTY, 2018, 33 (09) :2967-2970
[38]   Revision Total Hip Arthroplasty With a Monoblock Splined Tapered Grit-Blasted Titanium Stem [J].
Hellman, Michael D. ;
Kearns, Sean M. ;
Bohl, Daniel D. ;
Haughom, Bryan D. ;
Levine, Brett R. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (12) :3698-3703
[39]   Primary total hip arthroplasty failure: aseptic loosening remains the most common cause of revision [J].
Feng, Xiao ;
Gu, Jianming ;
Zhou, Yixin .
AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2022, 14 (10) :7080-7089
[40]   Does stem profile have an impact on the failure patterns in revision total knee arthroplasty? [J].
Maslaris, Alexander ;
Tsiridis, Eleftherios ;
Schoeneberg, Carsten ;
Pass, Bastian ;
Spyrou, Georgios ;
Maris, Alexandros ;
Matziolis, Georg .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (03) :1549-1569