Multiparametric MR Imaging with High-Resolution Dynamic Contrast-enhanced and Diffusion-weighted Imaging at 7 T Improves the Assessment of Breast Tumors: A Feasibility Study

被引:37
作者
Pinker, Katja [1 ]
Baltzer, Pascal [1 ]
Bogner, Wolfgang [2 ]
Leithner, Doris [1 ]
Trattnig, Siegfried [2 ]
Zaric, Olgica [2 ]
Dubsky, Peter [3 ]
Bago-Horvath, Zsuzsanna [4 ]
Rudas, Margaretha [4 ]
Gruber, Stephan [2 ]
Weber, Michael [1 ]
Helbich, Thomas H. [1 ]
机构
[1] Med Univ Vienna, Dept Biomed Imaging & Imageguided Therapy, Div Mol & Gender Imaging, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Biomed Imaging & Imageguided Therapy, MR Ctr Excellence, A-1090 Vienna, Austria
[3] Med Univ Vienna, Dept Surg, A-1090 Vienna, Austria
[4] Med Univ Vienna, Dept Pathol, A-1090 Vienna, Austria
关键词
ROBUST FAT-SUPPRESSION; DUCTAL CARCINOMA; BILATERAL COIL; RESONANCE; TESLA; SPECTROSCOPY; DIFFERENTIATION; DIAGNOSIS; LESIONS; BENIGN;
D O I
10.1148/radiol.15141905
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To ascertain whether multiparametric magnetic resonance (MR) imaging of the breast in combination with dynamic contrast material-enhanced (DCE) imaging and diffusion-weighted imaging (DWI) at 7 T is feasible and improves diagnostic accuracy. Materials and Methods: From December 2011 to December 2013, 40 patients with suspicious breast lesions were included in this institutional review board-approved prospective study. Before bilateral multiparametric MR imaging of the breast at 7 T, all patients gave written informed consent. Lesions were classified according to Breast Imaging Reporting and Data System (BI-RADS) and assessed for apparent diffusion coefficient (ADC) values by two readers independently. For combined analysis of DCE MR imaging and DWI, the BI-RADS-adapted reading algorithm, which adapted ADC thresholds to the BI-RADS assessment category, was used. Diagnostic values of multiparametric, DCE MR imaging, and DWI were calculated. Receiver operating characteristic curve analysis was performed. Image quality and interreader agreement were assessed. Histopathologic results were used as the highest standard. Results: There were 29 malignant and 17 benign lesions (range, 6-95 mm; mean, 23.3 mm). Multiparametric MR imaging yielded a sensitivity of 100% (29 of 29 lesions), a specificity of 88.2% (16 of 18 lesions), and an area under the curve of 0.941, which was greater than for DCE MR imaging (P = .003), which had a sensitivity of 100% (29 of 29 lesions), a specificity of 53.2% (nine of 17 lesions), and an area under the curve of 0.765. DWI had a sensitivity of 93.1% (27 of 29 lesions), a specificity of 88.2% (15 of 17 lesions), and an area under the curve of 0.907. Multiparametric MR imaging at 7 T of the breast eliminated all false-negative findings and reduced false-positive findings, from eight false-positive findings with DCE MR imaging to two false-positive findings. Thus, if used clinically, 7-T multiparametric MR imaging may have potentially obviated unnecessary breast biopsies in six of eight lesions (P = .031). Multiparametric MR imaging demonstrated either excellent or good image quality and interreader agreement (k = 0.89-1.00). Conclusion: The clinical use of 7-T multiparametric MR imaging is feasible, provides good or excellent image quality, and has the potential to improve diagnostic accuracy. (C)RSNA, 2015
引用
收藏
页码:360 / 370
页数:11
相关论文
共 35 条
[1]   Breast Lesions: Diagnosis by Using Proton MR Spectroscopy at 1.5 and 3.0 T-Systematic Review and Meta-Analysis [J].
Baltzer, Pascal A. T. ;
Dietzel, Matthias .
RADIOLOGY, 2013, 267 (03) :735-746
[2]   Bilateral Diffusion-weighted MR Imaging of Breast Tumors with Submillimeter Resolution Using Readout-segmented Echoplanar Imaging at 7 T [J].
Bogner, Wolfgang ;
Pinker, Katja ;
Zaric, Olgica ;
Baltzer, Pascal ;
Minarikova, Lenka ;
Porter, David ;
Bago-Horvath, Zsuzsanna ;
Dubsky, Peter ;
Helbich, Thomas H. ;
Trattnig, Siegfried ;
Gruber, Stephan .
RADIOLOGY, 2015, 274 (01) :74-84
[3]   Readout-segmented Echo-pIanar Imaging Improves the Diagnostic Performance of Diffusion-weighted MR Breast Examinations at 3.0 T [J].
Bogner, Wolfgang ;
Pinker-Domenig, Katja ;
Bickel, Hubert ;
Chmelik, Marek ;
Weber, Michael ;
Helbich, Thomas H. ;
Trattnig, Siegfried ;
Gruber, Stephan .
RADIOLOGY, 2012, 263 (01) :64-76
[4]   Diffusion-weighted MR for Differentiation of Breast Lesions at 3.0 T: How Does Selection of Diffusion Protocols Affect Diagnosis? [J].
Bogner, Wolfgang ;
Gruber, Stephan ;
Pinker, Katja ;
Grabner, Guenther ;
Stadlbauer, Andreas ;
Weber, Michael ;
Moser, Ewald ;
Helbich, Thomas H. ;
Trattnig, Siegfried .
RADIOLOGY, 2009, 253 (02) :341-351
[5]   Breast MRI at 7 Tesla with a bilateral coil and robust fat suppression [J].
Brown, Ryan ;
Storey, Pippa ;
Geppert, Christian ;
McGorty, KellyAnne ;
Leite, Ana Paula Klautau ;
Babb, James ;
Sodickson, Daniel K. ;
Wiggins, Graham C. ;
Moy, Linda .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (03) :540-549
[6]   Breast MRI at 7 Tesla with a bilateral coil and T1-weighted acquisition with robust fat suppression: image evaluation and comparison with 3 Tesla [J].
Brown, Ryan ;
Storey, Pippa ;
Geppert, Christian ;
McGorty, KellyAnne ;
Leite, Ana Paula Klautau ;
Babb, James ;
Sodickson, Daniel K. ;
Wiggins, Graham C. ;
Moy, Linda .
EUROPEAN RADIOLOGY, 2013, 23 (11) :2969-2978
[7]   COMPARING THE AREAS UNDER 2 OR MORE CORRELATED RECEIVER OPERATING CHARACTERISTIC CURVES - A NONPARAMETRIC APPROACH [J].
DELONG, ER ;
DELONG, DM ;
CLARKEPEARSON, DI .
BIOMETRICS, 1988, 44 (03) :837-845
[8]   Diffusion-weighted Imaging Improves the Diagnostic Accuracy of Conventional 3.0-T Breast MR Imaging [J].
El Khouli, Riham H. ;
Jacobs, Michael A. ;
Mezban, Sarah D. ;
Huang, Peng ;
Kamel, Ihab R. ;
Macura, Katarzyna J. ;
Bluemke, David A. .
RADIOLOGY, 2010, 256 (01) :64-73
[9]   Dynamic Contrast-Enhanced Magnetic Resonance Imaging of Breast Tumors at 3 and 7 T A Comparison [J].
Gruber, Stephan ;
Pinker, Katja ;
Zaric, Olgica ;
Minarikova, Lenka ;
Chmelik, Marek ;
Baltzer, Pascal ;
Boubela, Roland N. ;
Helbich, Thomas ;
Bogner, Wolfgang ;
Trattnig, Siegfried .
INVESTIGATIVE RADIOLOGY, 2014, 49 (05) :354-362
[10]   Three-dimensional Proton MR Spectroscopic Imaging at 3 T for the Differentiation of Benign and Malignant Breast Lesions [J].
Gruber, Stephan ;
Debski, Bogumil-Krystian ;
Pinker, Katja ;
Chmelik, Marek ;
Grabner, Guenther ;
Helbich, Thomas ;
Trattnig, Siegfried ;
Bogner, Wolfgang .
RADIOLOGY, 2011, 261 (03) :752-761