Multi-element analysis of size-segregated fine and ultrafine particulate via Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry

被引:15
作者
Rovelli, Sabrina [1 ]
Nischkauer, Winfried [2 ]
Cavallo, Domenico M. [1 ]
Limbeck, Andreas [2 ]
机构
[1] Univ Insubria, Dept Sci & High Technol, Via Valleggio 11, I-22100 Como, Italy
[2] TU Wien, Inst Chem Technol & Analyt, Getreidemarkt 9-164-IAC, A-1060 Vienna, Austria
关键词
LA-ICP-MS; Size-fractionated particulate; Sample fixation; Metal quantification; CARCINOGENIC METAL-COMPOUNDS; EMISSION-SPECTROMETRY; AIRBORNE PARTICLES; OXIDATIVE STRESS; TRACE-ELEMENTS; MATTER; FRACTION; GOTEBORG; FILTERS; SAMPLES;
D O I
10.1016/j.aca.2018.10.026
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study a novel and reliable Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry (LA-ICP-MS) measurement protocol for the elemental characterization of size-segregated particulate was developed. Special efforts were made to improve and optimize sample pre-treatment steps and LA operating conditions to avoid some critical drawbacks encountered during analysis and to make the particulate samples suitable for an accurate and reproducible LA-ICP-MS analysis, regardless of the mass loading on each filter. For example, a new approach for dust-fixation on the sample-carrier was developed using a glycerol coverage, which allowed to overcome problematic sample losses during the ablation process. Under the optimum conditions, dust samples, blank filters and standards for calibration were analyzed by multiple rastering of defined spot areas. Quantitative analysis was accomplished with dried micro-droplets of aqueous standard solutions. Derived method detection limits varied between 0.001 and 0.1 ng m(-3) and allowed even for the smallest particle fraction quantitative measurements. The accuracy of LA-ICP-MS results was verified by comparison with conventional ICP-MS analysis of selected PM samples after sample mineralization. The proposed LA treatment procedure benefits from a simple and fast sample preparation, thus overcoming the laborious pre-treatment steps required for wet chemical digestion. Moreover, the better sensitivity of the LA-ICP-MS approach provided more complete information about the mass concentration and size-distribution of the investigated elements, thus allowing to deeper investigate the composition of the most dangerous PM fractions in terms of health concern. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:11 / 19
页数:9
相关论文
共 50 条
  • [21] High-speed mapping of Hg and Se in biological tissue via laser ablation-inductively coupled plasma-mass spectrometry
    Van Helden, Tom
    Braeuer, Simone
    Van Acker, Thibaut
    Leroux, Olivier
    Van Der Straeten, Dominique
    Vanhaecke, Frank
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (07) : 1455 - 1461
  • [22] Direct Quantitative Determination of Trace Elements in Fine-Grained Whole Rocks by Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry
    Li, Ming
    Hu, Zhaochu
    Gao, Shan
    Liu, Yongsheng
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2011, 35 (01) : 7 - 22
  • [23] High spatial resolution quantitative elemental imaging of foraminifer by laser ablation-inductively coupled plasma-mass spectrometry
    Ke, Yuqiu
    Zhou, Jianzong
    Qiao, Lei
    Zhang, Muhui
    Guo, Wei
    Jin, Lanlan
    Hu, Shenghong
    ANALYTICAL METHODS, 2019, 11 (16) : 2129 - 2137
  • [24] A new technique to determine element amounts down to femtograms in dust using femtosecond laser ablation-inductively coupled plasma-mass spectrometry
    Macholdt, Dorothea S.
    Jochum, Klaus Peter
    Stoll, Brigitte
    Weis, Ulrike
    Andreae, Meinrat O.
    CHEMICAL GEOLOGY, 2014, 383 : 123 - 131
  • [25] Natural Obsidian Glass as an External Accuracy Reference Material in Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry
    Ulrich, Thomas
    Kamber, Balz S.
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2013, 37 (02) : 169 - 188
  • [26] Quantification strategies for elemental imaging of biological samples using laser ablation-inductively coupled plasma-mass spectrometry
    Hare, Dominic
    Austin, Christine
    Doble, Philip
    ANALYST, 2012, 137 (07) : 1527 - 1537
  • [27] Solution-based calibration strategy for laser ablation-inductively coupled plasma-mass spectrometry using desolvating nebulizer system
    Zhang, Guoxia
    Li, Qing
    Zhu, Yan
    Wang, Zheng
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 145 : 51 - 57
  • [28] LASER ABLATION-INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY IN ARCHAEOMETRIC RESEARCH
    Resano, Martin
    Garcia-Ruiz, Esperanza
    Vanhaecke, Frank
    MASS SPECTROMETRY REVIEWS, 2010, 29 (01) : 55 - 78
  • [29] Quasi 'non-destructive' laser ablation-inductively coupled plasma-mass spectrometry fingerprinting of sapphires
    Guillong, M
    Günther, D
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (07) : 1219 - 1231
  • [30] The determination of low level trace elements in coals by laser ablation-inductively coupled plasma-mass spectrometry
    Booth, CA
    Spears, DA
    Krause, P
    Cox, AG
    FUEL, 1999, 78 (14) : 1665 - 1670