Contrast agent comparison for three-dimensional micro-CT angiography: A cadaveric study

被引:23
作者
Kingston, Mitchell J. [1 ,2 ]
Perriman, Diana M. [1 ,2 ]
Neeman, Teresa [3 ]
Smith, Paul N. [1 ,2 ]
Webb, Alexandra L. [1 ]
机构
[1] Australian Natl Univ, Sch Med, Coll Med Biol & Environm, 54 Mills Rd, Canberra, ACT 2601, Australia
[2] Canberra Hosp, TORU, Canberra, ACT, Australia
[3] Australian Natl Univ, Stat Consulting Unit, Canberra, ACT, Australia
关键词
micro-computed tomography; angiography; barium sulfate; lead oxide; contrast media; INJECTION TECHNIQUE; FLAP; ANATOMY;
D O I
10.1002/cmmi.1695
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Barium sulfate and lead oxide contrast media are frequently used for cadaver-based angiography studies. These contrast media have not previously been compared to determine which is optimal for the visualisation and measurement of blood vessels. In this study, the lower limb vessels of 16 embalmed Wistar rats, and four sets of cannulae of known diameter, were injected with one of three different contrast agents (barium sulfate and resin, barium sulfate and gelatin, and lead oxide combined with milk powder). All were then scanned using micro-computed tomography (CT) angiography and 3-D reconstructions generated. The number of branching generations of the rat lower limb vessels were counted and compared between the contrast agents using ANOVA. The diameter of the contrast-filled cannulae, were measured and used to calculate the accuracy of the measurements by comparing the bias and variance of the estimates. Intra-and inter-observer reliability were calculated using intra-class correlation coefficients. There was no significant difference (mean difference [MD] 0.05; MD 95% confidence interval [CI] -0.83 to 0.93) between the number of branching generations for barium sulfate-resin and lead oxide-milk powder. Barium sulfate-resin demonstrated less bias and less variance of the estimates (MD 0.03; standard deviation [SD] 1.96 mm) compared to lead oxide-milk powder (MD 0.11; SD 1.96 mm) for measurements of contrast-filled cannulae scanned at high resolution. Barium sulfate-resin proved to be more accurate than lead oxide-milk powder for high resolution micro-CT scans and is preferred due to its non-toxicity. This technique could be applied to any embalmed specimen model. Copyright (C) 2016 John Wiley & Sons, Ltd.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 50 条
  • [41] Micro-CT versus synchrotron radiation phase contrast imaging of human cochlea
    Elfarnawany, M.
    Alam, S. Riyahi
    Rohani, S. A.
    Zhu, N.
    Agrawal, S. K.
    Ladak, H. M.
    JOURNAL OF MICROSCOPY, 2017, 265 (03) : 349 - 357
  • [42] Examination of inner ear structures: a micro-CT study
    Geneci, Ferhat
    Uzuner, Muhammet Bora
    Bilecenoglu, Burak
    Torun, Bilge I. pek
    Orhan, Kaan
    Ocak, Mert
    ACTA OTO-LARYNGOLOGICA, 2022, 142 (01) : 1 - 5
  • [43] Morphometric micro-CT study of contralateral mandibular incisors
    Sevgi, Usame
    Johnsen, Gaute Floer
    Hussain, Badra
    Piasecki, Lucila
    Nogueira, Liebert Parreiras
    Haugen, Havard Jostein
    CLINICAL ORAL INVESTIGATIONS, 2023, 28 (01)
  • [44] Viewing Three-Dimensional Structure of the Cadaveric Brain Arteries Running Head: Three-Dimensional Structure of the Brain Arteries
    Ozdemir, Mevci
    Comert, Ayhan
    Citisli, Veli
    Kahilogullari, Gokmen
    Tekdemir, Ibrahim
    Acar, Feridun
    Egemen, Nihat
    JOURNAL OF NEUROLOGICAL SCIENCES-TURKISH, 2017, 34 (02): : 162 - 166
  • [45] How Can We Show You, If You Can't See It? Trialing the Use of an Interactive Three-Dimensional Micro-CT Model in Medical Education
    O'Rourke, Joseph C.
    Smyth, Lillian
    Webb, Alexandra L.
    Valter, Krisztina
    ANATOMICAL SCIENCES EDUCATION, 2020, 13 (02) : 206 - 217
  • [46] Diagnostic value of three-dimensional reconstruction with multislice spiral CT angiography in patients with cerebrovascular disease
    Sui, R-D
    Wang, H-T
    Yu, X-Y
    Zhang, X-Q
    Wei, X-L
    Xin, D-Y
    Han, D-W
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2020, 34 (05) : 1799 - 1807
  • [47] Three-dimensional CT angiography for surgical planning in congenital hand malformations: a case series presentation
    Morandi, Evi M.
    Rieger, Michael
    Baur, Eva-Maria
    Piza-Katzer, Hildegunde
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2020, 45 (10) : 1017 - 1022
  • [48] Relationship between Injection Rate and Contrast Enhancement on Three-dimensional Digital Subtraction Angiography of the Cerebral Arteries
    Takagi, Satoshi
    Hanasaki, Naoyuki
    JOURNAL OF THE BELGIAN SOCIETY OF RADIOLOGY, 2018, 102 (01):
  • [49] Takayasu arteritis: Diagnosis with breath-hold contrast-enhanced three-dimensional MR angiography
    Yamada, I
    Nakagawa, T
    Himeno, Y
    Kobayashi, Y
    Numano, F
    Shibuya, H
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 11 (05): : 481 - 487
  • [50] Optical comparison between micro-CT and OCT in imaging of marginal composite adaptation: Observational study
    Bakhsh, Turki A.
    JOURNAL OF MICROSCOPY, 2021, 282 (02) : 136 - 145