Determination of Ultratrace Selenium and Arsenic in Drinking Water by Electrochemical Hydride Generation Coupled with Quartz Furnace - Atomic Absorption Spectrometry

被引:0
|
作者
Hranicek, Jakub [1 ]
Cerveny, Vaclav [1 ]
Rychlovsky, Petr [1 ]
机构
[1] Charles Univ Prague, Prirodovedecka Fak, Katedra Analyt Chem, Prague 12843 2, Czech Republic
来源
CHEMICKE LISTY | 2010年 / 104卷 / 12期
关键词
IN-SITU CONCENTRATION; SAMPLE-INTRODUCTION; VOLATILE COMPOUNDS; NATURAL-WATERS; DESIGN; SYSTEM; CELL; AAS;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The aim of this study was to develop a method of Se and As determination in drinking water by electrochemical hydride generation coupled with AAS. Three types of electrolytic cells were constructed and optimized for the determination. Two cells (thin-layer and tubular) were finally chosen for their low inner volume and a high hydride generation efficiency (ca. 90 %). The limits of detection of As(III) and Se(IV) were 0.07 and 0.37 ng ml(-1), respectively. The accuracy of the method was verified using reference materials.
引用
收藏
页码:1196 / 1203
页数:8
相关论文
共 50 条