The uncertainty of UV-Vis spectrophotometric and FAAS analysis for the determination of iron in pharmaceutical products

被引:4
作者
Jurgens, Jaana [1 ]
Paama, Lilli [1 ]
Leito, Ivo [1 ]
机构
[1] Univ Tartu, Inst Chem Phys, EE-51014 Tartu, Estonia
关键词
uncertainty; ISO GUM; UV-Vis spectrophotometry; FAAS; iron;
D O I
10.1007/s00769-007-0299-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
UV-Vis spectrophotometric (UV-Vis) and flame atomic absorption spectrometric (FAAS) analysis for iron determination in a pharmaceutical product were compared in terms of uncertainty budgets. Both methods are selective for iron; no interference due to other components present in the matrix was found. The results for the UV-Vis and FAAS methods were 11.4 +/- 0.5 mg and 11.9 +/- 0.8 mg, respectively, of iron per tablet (at a 95% confidence level). In both methods, the main uncertainty contribution is that due to the calibration function's non-linearity (0.32 mg and 0.46 mg or 42% and 31% of the standard uncertainty in the case of UV-Vis and FAAS, respectively). This finding encourages us to take uncertainty due to non-linearity into account in uncertainty estimations, even for highly linear methods like UV-Vis. In the ranking of uncertainty contributors, non-linearity is followed by instrument drift in the FAAS method and the uncertainty of volumetric measurements in UV-Vis. In particular, pipetting contributes about 16% of the uncertainty for UV-Vis spectrometry and 10% for FAAS, which is somewhat larger than that usually assumed. The uncertainty contributions of the two methods are compared and discussed in terms of experimental conditions.