Accuracy of 2 Endodontic Rotary Motors with Integrated Apex Locator

被引:19
作者
Cruz, Alessandra Timponi Goes [1 ]
Wichnieski, Caroline [1 ]
Carneiro, Everdan [1 ]
da Silva Neto, Ulisses Xavier [1 ]
Gambarini, Gianluca [2 ]
Piasecki, Lucila [3 ]
机构
[1] Pontificia Univ Catolica Parana, Sch Hlth & Biosci, Dept Endodont, Curitiba, Parana, Brazil
[2] Univ Roma La Sapienza, Dept Endodont, Rome, Italy
[3] SUNY Buffalo, Dept Periodont & Endodont, 240 Squire Hall, Buffalo, NY 14214 USA
关键词
Apical limit; auto apical reverse; electronic apex locator; endodontic motor; ROOT-ZX-II; WORKING LENGTH DETERMINATION; MICRO-COMPUTED TOMOGRAPHY; IN-VIVO EVALUATION; TRI AUTO ZX; CANAL LENGTH; APICAL CONSTRICTION; PROTAPER INSTRUMENTS; MEASUREMENT DEVICES; FORAMEN;
D O I
10.1016/j.joen.2017.05.012
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The purpose of this study was to evaluate in vitro the efficacy of both the electronic apex locator (EAL) and auto apical reverse (AAR) functions of the endodontic motor MM Control (Micro-Mega, Besancon Cedex, France) compared with Root ZX II (J Morita, Tokyo, Japan). Methods: After cervical preflaring, the actual lengths (ALs) of 36 single-rooted teeth were obtained up to the apical foramen. The EAL measurements at the marks "APEX" and "0.5" of both devices were obtained using an alginate model. The teeth were divided randomly into 2 groups (n = 18), and root canal preparation was performed with rotary instruments using the AAR function (0.5 mark) of each motor. The length provided by the AAR was compared with the visual length after preparation (AL2). The differences between the electronic lengths and the respective visual measurements were assigned as negative for lower or positive when higher. The means of the absolute values and the percentages of distribution of the electronic measurements between devices were compared. Results: There was no difference between the devices in terms of the means of the EAL measurements or AAR length (analysis of variance, P>.05). However, the EAL function of MM Control presented a greater percentage of measurements >1.01 mm longer than AL (chi-square, P<.01). The AAR function provided an acceptable apical limit in 83.3% of the cases for Root ZX and 77.8% of the cases for MM Control. Conclusions: The AAR function of both MM Control and Root ZX II provided an adequate apical limit of preparation in vitro. However, the use of only the EAL function of MM Control resulted in significantly more cases of overextended readings.
引用
收藏
页码:1716 / 1719
页数:4
相关论文
共 34 条
[1]   Not All Electronic Foramen Locators Are Accurate in Teeth with Enlarged Apical Foramina: An In Vitro Comparison of 5 Brands [J].
Akisue, Eduardo ;
Gratieri, Simone Domingues ;
Barletta, Fernando Branco ;
Caldeira, Celso Luiz ;
Grazziotin-Soares, Renata ;
Gavini, Giulio .
JOURNAL OF ENDODONTICS, 2014, 40 (01) :109-112
[2]   Effect of canal preparation and residual root filling material on root impedance [J].
Al-bulushi, A. ;
Levinkind, M. ;
Flanagan, M. ;
Ng, Y. -L. ;
Gulabivala, K. .
INTERNATIONAL ENDODONTIC JOURNAL, 2008, 41 (10) :892-904
[3]   Accuracy of Root Length Determination Using Tri Auto ZX and ProTaper Instruments: An In Vitro Study [J].
Carneiro, Everdan ;
Bramante, Clovis Monteiro ;
Picoli, Fabio ;
Letra, Ariadne ;
da Silva Neto, Ulisses Xavier ;
Menezes, Renato .
JOURNAL OF ENDODONTICS, 2006, 32 (02) :142-144
[4]   Influence of Preflaring on the Accuracy of Length Determination With Four Electronic Apex Locators [J].
de Camargo, Ericson Janolio ;
Zapata, Ronald Ordinola ;
Medeiros, Paulo Leal ;
Bramante, Clovis Monteiro ;
Bernardineli, Norberti ;
Garcia, Roberto Brandao ;
de Moraes, Ivaldo Gomes ;
Hungaro Duarte, Marco Antonio .
JOURNAL OF ENDODONTICS, 2009, 35 (09) :1300-1302
[5]   Evaluation of the Maintenance of the Apical Limit during Instrumentation with Hybrid Equipment in Rotary and Reciprocating Modes [J].
de Vasconcelos, Bruno Carvalho ;
Arcanjo Frota, Luciana Maria ;
Souza, Tamara de Abreu ;
Bernardes, Ricardo Affonso ;
Hungaro Duarte, Marco Antonio .
JOURNAL OF ENDODONTICS, 2015, 41 (05) :682-685
[6]   THE POSITION AND TOPOGRAPHY OF THE APICAL CANAL CONSTRICTION AND APICAL FORAMEN [J].
DUMMER, PMH ;
MCGINN, JH ;
REES, DG .
INTERNATIONAL ENDODONTIC JOURNAL, 1984, 17 (04) :192-198
[7]   A simple mounting model for consistent determination of the accuracy and repeatability of apex locators [J].
ElAyouti, A ;
Löst, C .
INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (02) :108-112
[8]   Apical Constriction: Location and Dimensions in Molars-A Micro-Computed Tomography Study [J].
ElAyouti, Ashraf ;
Huelber-J, Maren ;
Judenhofer, Martin S. ;
Connert, Thomas ;
Mannheim, Julia G. ;
Lost, Claus ;
Pichlen, Bernd J. ;
von Ohle, Christiane .
JOURNAL OF ENDODONTICS, 2014, 40 (08) :1095-1099
[9]   Consistency of Apex Locator Function: A Clinical Study [J].
ElAyouti, Ashraf ;
Dima, Eleftheria ;
Ohmer, Julia ;
Sperl, Kristina ;
von Ohle, Christiane ;
Loest, Claus .
JOURNAL OF ENDODONTICS, 2009, 35 (02) :179-181
[10]   An in vivo evaluation of the Auto Apical Reverse function of the Root ZX II [J].
Fadel, G. ;
Piasecki, L. ;
Westphalen, V. P. D. ;
Silva Neto, U. X. ;
Fariniuk, L. F. ;
Carneiro, E. .
INTERNATIONAL ENDODONTIC JOURNAL, 2012, 45 (10) :950-954