Clinical accuracy of the MedGemt indirect calorimeter for measuring resting energy expenditure in cancer patients

被引:42
|
作者
Reeves, MM
Capra, S
Bauer, J
Davies, PSW
Battistutta, D
机构
[1] Queensland Univ Technol, Ctr Hlth Res, Brisbane, Qld, Australia
[2] Wesley Res Inst, Brisbane, Qld, Australia
[3] Univ Newcastle, Australian Ctr Evidence Based Nutr & Dietet, Newcastle, NSW 2308, Australia
[4] Univ Queensland, Royal Childrens Hosp, Childrens Nutr Res Ctr, Brisbane, Qld, Australia
关键词
REE; oxygen consumption; metabolic rate; cancer; nutrition support;
D O I
10.1038/sj.ejcn.1602114
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: To compare, in patients with cancer and in healthy subjects, measured resting energy expenditure (REE) from traditional indirect calorimetry to a new portable device (MedGem(TM)) and predicted REE. Design: Cross-sectional clinical validation study. Setting: Private radiation oncology centre, Brisbane, Australia. Subjects: Cancer patients (n=18) and healthy subjects (n=17) aged 37-86y, with body mass indices ranging from 18 to 42 kg/m(2). Interventions: Oxygen consumption (VO2) and REE were measured by VMax229 (VM) and MedGem (MG) indirect calorimeters in random order after a 12-h fast and 30-min rest. REE was also calculated from the MG without adjustment for nitrogen excretion (MGN) and estimated from Harris-Benedict prediction equations. Data were analysed using the Bland and Altman approach, based on a clinically acceptable difference between methods of 5%. Results: The mean bias (MGN-VM) was 10% and limits of agreement were -42 to 21% for cancer patients; mean bias -5% with limits of -45 to 35% for healthy subjects. Less than half of the cancer patients (n=7, 46.7%) and only a third (n=5, 33.3%) of healthy subjects had measured REE by MGN within clinically acceptable limits of VM. Predicted REE showed a mean bias (HB-VM) of -5% for cancer patients and 4% for healthy subjects, with limits of agreement of -30 to 20% and -27 to 34%, respectively. Conclusions: Limits of agreement for the MG and Harris Benedict equations compared to traditional indirect calorimetry were similar but wide, indicating poor clinical accuracy for determining the REE of individual cancer patients and healthy subjects.
引用
收藏
页码:603 / 610
页数:8
相关论文
共 50 条
  • [1] Clinical accuracy of the MedGem™ indirect calorimeter for measuring resting energy expenditure in cancer patients
    M M Reeves
    S Capra
    J Bauer
    P S W Davies
    D Battistutta
    European Journal of Clinical Nutrition, 2005, 59 : 603 - 610
  • [2] Accuracy of a Portable Indirect Calorimeter for Measuring Resting Energy Expenditure in Individuals With Cancer
    Purcell, Sarah A.
    Elliott, Sarah A.
    Ryan, Aoife M.
    Sawyer, Michael B.
    Prado, Carla M.
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2019, 43 (01) : 145 - 151
  • [3] Accuracy and reliability of a portable indirect calorimeter compared to whole-body indirect calorimetry for measuring resting energy expenditure
    Purcell, Sarah A.
    Johnson-Stoklossa, Carlene
    Tibaes, Jenneffer Rayane Braga
    Frankish, Alena
    Elliott, Sarah A.
    Padwal, Raj
    Prado, Carla M.
    CLINICAL NUTRITION ESPEN, 2020, 39 : 67 - 73
  • [4] Accuracy and preference of measuring resting energy expenditure using a handheld calorimeter in healthy adults
    Madden, A. M.
    Parker, L. J. F.
    Amirabdollahian, F.
    JOURNAL OF HUMAN NUTRITION AND DIETETICS, 2013, 26 (06) : 587 - 595
  • [5] Accuracy of the MedGem® portable indirect calorimeter for measuring resting energy expenditure in adults with class II or III obesity
    Purcell, Sarah A.
    Johnson-Stoklossa, Carlene
    Braga Tibaes, Jenneffer Rayane
    Frankish, Alena
    Elliott, Sarah A.
    Padwal, Raj
    Prado, Carla M.
    CLINICAL NUTRITION ESPEN, 2020, 40 : 408 - 411
  • [6] Accuracy of Resting Energy Expenditure Predictive Equations in Patients With Cancer
    Purcell, Sarah A.
    Elliott, Sarah A.
    Baracos, Vickie E.
    Chu, Quincy S. C.
    Sawyer, Michael B.
    Mourtzakis, Marina
    Easaw, Jacob C.
    Spratlin, Jennifer L.
    Siervo, Mario
    Prado, Carla M.
    NUTRITION IN CLINICAL PRACTICE, 2019, 34 (06) : 922 - 934
  • [7] Assessing resting energy expenditure in overweight and obese adolescents in a clinical setting: validity of a handheld indirect calorimeter
    Woo, Paula
    Murthy, Gayathri
    Wong, Cindy
    Hursh, Brenden
    Chanoine, Jean-Pierre
    Elango, Rajavel
    PEDIATRIC RESEARCH, 2017, 81 (01) : 51 - 56
  • [8] Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient
    Wang, Xi
    Wang, Yuan
    Ding, Zenghui
    Cao, Guangbei
    Hu, Fusong
    Sun, Yining
    Ma, Zuchang
    Zhou, Duoqi
    Su, Benyue
    ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION, 2018, 27 (01) : 72 - 77
  • [9] Measurement of resting energy expenditure and its accuracy in women with breast cancer
    Ciofi, Iolanda
    Di Vincenzo, Olivia
    Morlino, Delia
    da Silva, Bruna Ramos
    Prado, Carla M.
    Santarpia, Lidia
    Scalfi, Luca
    Giuliano, Mario
    De Angelis, Carmine
    Pasanisi, Fabrizio
    Arpino, Grazia
    Marra, Maurizio
    CLINICAL NUTRITION, 2024, 43 (11) : 54 - 64
  • [10] Validation of the BIOPAC indirect calorimeter for determining resting energy expenditure in healthy free-living older people
    Khalaj-Hedayati, Kerstin
    Bosy-Westphal, Anja
    Mueller, Manfred J.
    Dittmar, Manuela
    NUTRITION RESEARCH, 2009, 29 (08) : 531 - 541