6-DOF Force Sensing for the Master Tool Manipulator of the da Vinci Surgical System

被引:23
|
作者
Black, David G. [1 ]
Hosseinabadi, Amir H. Hadi [1 ]
Salcudean, Septimiu E. [1 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Robot & Control Lab, Vancouver, BC V6T 1Z4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Haptics and haptic interfaces; surgical robotics; laparoscopy; force and tactile sensing; telerobotics and teleoperation; dVRK; HAPTIC FEEDBACK; ROBOT; SURGERY; DESIGN;
D O I
10.1109/LRA.2020.2970944
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
We integrated a force/torque sensor into the wrist of the Master Tool Manipulator (MTM) of the da Vinci Standard Surgical system. The added sensor can be used to monitor the surgeon interaction forces and to improve the haptic experience. The proposed mechanical design is expected to have little effect on the surgeon's operative experience and is simple and inexpensive to implement. We also developed a software package that allows for seamless integration of the force sensor into the da Vinci Research Kit (dVRK) and the Robot Operating System (ROS). The complete mechanical and electrical modifications, as well as the software packages are discussed. Two example applications of impedance control at the MTM and joystick control of the PSM are presented to demonstrate the successful integration of the sensor into the MTM and the interface to the dVRK.
引用
收藏
页码:2264 / 2271
页数:8
相关论文
共 50 条
  • [21] Visual Servo of a 6-DOF Robotic Stereo Flexible Endoscope Based on da Vincix Research Kit (dVRK) System
    Ma, Xin
    Song, Chengzhi
    Chiu, Philip Waiyan
    Li, Zheng
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2020, 5 (02) : 820 - 827
  • [22] Suture damage after grasping with EndoWrist of the da Vinci Surgical System
    Hirano, Yasumitsu
    Ishikawa, Norihiko
    Watanabe, Go
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2010, 19 (3-4) : 203 - 206
  • [23] 6-DOF Loading System with the Simulated Border
    Li, Bing
    Wei, Yulan
    Wang, Yuezhan
    Yan, Qibo
    CHEMICAL ENGINEERING AND MATERIAL PROPERTIES II, 2012, 549 : 894 - 898
  • [24] Surgical techniques: Robotic bladder diverticulectomy with the da Vinci-S surgical system
    Rao R.
    Nayyar R.
    Panda S.
    Hemal A.K.
    Journal of Robotic Surgery, 2007, 1 (03) : 217 - 220
  • [25] TRAJECTORY PLANNING OF 6-DOF MANIPULATOR BASED ON GAUSSIAN PROCESS REGRESSION METHOD
    Zhao, Jinzhi
    Wang, Shizhao
    Jiang, Aibing
    Xiao, Jin
    Wang, Bin
    INTERNATIONAL JOURNAL OF ROBOTICS & AUTOMATION, 2020, 35 (03) : 209 - 220
  • [26] 6-DOF Hybrid Cable-Driven Parallel Robot with an Articulated Manipulator
    Sakurai, Shunichi
    Katsura, Seiichiro
    2024 33RD INTERNATIONAL SYMPOSIUM ON INDUSTRIAL ELECTRONICS, ISIE 2024, 2024,
  • [27] Kinetostatic analysis of a novel 6-DoF 3UPS parallel manipulator with multi-fingers
    Lu, Yi
    Wang, Peng
    Hou, Zhuolei
    Hu, Bo
    Sui, Chunping
    Han, Jianda
    MECHANISM AND MACHINE THEORY, 2014, 78 : 36 - 50
  • [28] A Series-Parallel Hybrid Pelvic Fracture Reduction Surgical Robotic System Based on Novel 6-DOF Force Amplification Mechanism
    Wang, Qianxin
    Lu, Shaolin
    Liu, Pengyun
    Yang, Yuanyuan
    Wang, Qiong
    Li, Bing
    Cai, Shibo
    Zhang, Lihai
    Zhang, Jianwei
    Hu, Ying
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2024, 6 (03): : 1042 - 1053
  • [29] Minimally invasive resection of thymomas with the da Vinci® Surgical System
    Schneiter, Didier
    Tomaszek, Sandra
    Kestenholz, Peter
    Hillinger, Sven
    Opitz, Isabelle
    Inci, Ilhan
    Weder, Walter
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (02) : 288 - 292
  • [30] The da Vinci® Surgical System Overcomes Innate Hand Dominance
    Mucksavage, Phillip
    Kerbl, David C.
    Lee, Jason Y.
    JOURNAL OF ENDOUROLOGY, 2011, 25 (08) : 1385 - 1388