Topography-based screening for previous laser in situ keratomileusis to correct myopia and hyperopia

被引:7
作者
Laliberté, JF
Meunier, J
Hick, S
Brunette, I
机构
[1] Hop Maison Neuve Rosemont, Ophthalmol Res Unit, Montreal, PQ H1T 2H1, Canada
[2] Univ Montreal, Dept Comp Sci & Operat Res, Montreal, PQ, Canada
[3] Univ Montreal, Dept Ophthalmol, Montreal, PQ, Canada
关键词
comeal topography; LASIK; myopia; hyperopia; astigmatism;
D O I
10.1097/00003226-200503000-00008
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To develop a screening tool based on corneal topography to detect previous myopic and hyperopic laser in situ keratomileusis (LASIK). Methods: Clinical data from three private clinics were analyzed in a university setting. A total of 476 topographies (Orbscan II) were randomly selected (1 topography per patient): 338 from unoperated corneas, 81 from corneas that had LASIK to correct myopia, and 57 from corneas that had LASIK to correct hyperopia. The LASIK procedures were performed using a Technolas excimer laser (217C or 217Z) and a Hansatome microkeratome. The first set of algorithms (VESm and VESh) calculated the volume summation between the anterior corneal surface and the best-fit sphere in the central and mid-peripheral regions. The second set of algorithms (DCm and DCh) calculated the mean anterior tangential curvature difference between the central and mid-peripheral regions. Results: The features VESm and DCm for the screening of a myopic LASIK yield, respectively, sensitivity rates of 92.7% and 95.1% and specificity rates of 94.9% and 96.5%. The features VESh and DCh for the screening of a hyperopic LASIK gave sensitivity rates of 93.1% for both and specificity rates of 94.8% and 97.1%. The performance of those features was superior to clinical assessment using topography color maps. Conclusions: Criteria based on Orbscan II corneal topography are proposed for the detection of previous hyperopic and myopic LASIK performed with a Technolas excimer laser (version 217C or 217Z).
引用
收藏
页码:167 / 177
页数:11
相关论文
共 14 条
[1]  
BISHOP C, 1994, NEURAL NETWORK PATTE, P372
[2]   SVMTorch: Support vector machines for large-scale regression problems [J].
Collobert, R ;
Bengio, S .
JOURNAL OF MACHINE LEARNING RESEARCH, 2001, 1 (02) :143-160
[3]   Topography-based screening for previous laser in situ keratomileusis to correct myopia [J].
Destrempes, F ;
Brunette, I ;
Meunier, J ;
Beyrouthy, M ;
Demers, P ;
Fanous, S ;
Doyon, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (09) :1644-1650
[4]  
Hamilton D Rex, 2003, Curr Opin Ophthalmol, V14, P44, DOI 10.1097/00055735-200302000-00008
[5]  
Hart, 2006, PATTERN CLASSIFICATI
[6]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[7]   Screening of myopic photorefractive keratectomy in eye bank eyes by computerized videokeratography [J].
Lim-Bon-Siong, R ;
Williams, JM ;
Samapunphong, S ;
Chuck, RS ;
Pepose, JS .
ARCHIVES OF OPHTHALMOLOGY, 1998, 116 (05) :617-623
[8]   Use of a portable topography machine for screening donor tissue for prior refractive surgery [J].
Ousley, PJ ;
Terry, MA .
CORNEA, 2002, 21 (08) :745-750
[9]   Objective screening methods for prior refractive surgery in donor tissue [J].
Ousley, PJ ;
Terry, MA .
CORNEA, 2002, 21 (02) :181-188
[10]   Topographic detection of photorefractive keratectomy [J].
Schallhorn, SC ;
Reid, JL ;
Kaupp, SE ;
Blanton, CL ;
Zoback, L ;
Goforth, H ;
Flowers, CW ;
McDonnell, PJ .
OPHTHALMOLOGY, 1998, 105 (03) :507-516