Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover During Negative Pressure Application

被引:3
|
作者
Brambilla, Enrico [1 ]
Ferguson, Tanner J. [2 ]
Chu, Nathan [1 ]
Ammar, David [3 ]
Yoo, Paul [1 ]
机构
[1] Equinox Ophthalm Inc, Newport Beach, CA USA
[2] Vance Thompson Vis, 3101 W 57th St, Sioux Falls, SD 57108 USA
[3] Lions Eye Inst Transplant & Res, Tampa, FL USA
来源
CLINICAL OPHTHALMOLOGY | 2022年 / 16卷
关键词
tonometry; intraocular pressure; TOP; multi-pressure dial; MPD; ophthalmic techniques; diagnostics; GLAUCOMA PROGRESSION;
D O I
10.2147/OPTH.S359605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This is a 2-part study to investigate the agreement between pneumatonometry and direct pressure transducer intraocular pressure ( IOP) measurements in a perfusion organ culture (POC) model where (1) the perfusion fluid column is open to atmospheric pressure, holding IOP constant to permit evaluation of the impact of negative pressure (NP) on IOP measurements, and (2) the perfusion fluid column is a closed system, allowing IOP to vary with NP application. Methods: The first part incorporated a fluid column open to atmospheric pressure, maintaining IOP constant to permit evaluation of the effect of applied NP on IOP measurement accuracy. In the second part, the POC column was closed, allowing IOP to vary with NP application and permit evaluation of agreement between pneumatonometry and pressure transducer measurements. In each part, four perfused tissues were used in thirteen paired pre-set IOP (10, 20, 25, 30 mmHg) and NP (0, 5, 10, 15, 20 mmHg) combinations, resulting in a total of 1040 paired measurements (520 per study). The difference in IOP measurements (Delta IOP = Excursion tonometry - pressure transducer) was calculated at each paired configuration. Results: During the first part, the mean Delta IOP was -0.7 +/- 1.6 mmHg across all measurements. During the second part, the mean Delta IOP across all measurements was +0.7 +/- 1.4 mmHg. At NP settings of -5, -10, -15, and -20 mmHg, across all pre-set IOPs, the mean TOP reduction via Excursion tonometry was 3.1 +/- 0.3, 5.6 +/- 1.3, 8.5 +/- 1.7 and 11.2 +/- 1.8 mmHg, respectively. Conclusion: Measurement of IOP via Excursion tonometry yields results within the accuracy range of the pneumatonometry device (per manufacturer) and is minimally impacted by NP application. The IOP-lowering results are consistent with previous studies and further support the effectiveness of the Multi-Pressure Dial in lowering IOP relative to atmospheric pressure.
引用
收藏
页码:1289 / 1300
页数:12
相关论文
共 50 条
  • [31] THE KEELER PULSAIR AIR IMPULSE TONOMETER - COMPARISON WITH THE PERKINS HAND-HELD APPLANATION TONOMETER FOR PERI-OPERATIVE MEASUREMENT OF INTRAOCULAR-PRESSURE
    BRICKER, SRW
    MCGALLIARD, JN
    MOSTAFA, SM
    ANAESTHESIA, 1990, 45 (01) : 36 - 39
  • [32] Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
    Porwal, Amit C.
    Shrishrimal, Meghna
    Punamia, Ravin P.
    Mathew, Bennet Chacko
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 1927 - 1931
  • [33] A comparison of intraocular pressure measurements with the XPERT noncontact tonometer and Goldmann applanation tonometry
    Popovich, KS
    Shields, MB
    JOURNAL OF GLAUCOMA, 1997, 6 (01) : 44 - 46
  • [34] Estimation of intraocular pressure in rabbits using noncontact tonometer: A comparative evaluation with Schiotz tonometer
    Gupta, S. K.
    Saxena, R.
    Agarwal, R.
    Galpalli, N. D.
    Srivastava, S.
    Agrawal, S. S.
    METHODS AND FINDINGS IN EXPERIMENTAL AND CLINICAL PHARMACOLOGY, 2007, 29 (06): : 405 - 409
  • [35] A History of Intraocular Pressure and Its Measurement
    Stamper, Robert L.
    OPTOMETRY AND VISION SCIENCE, 2011, 88 (01) : E16 - E28
  • [36] Intraocular pressure measurement in the conscious rat
    Cabrera, CL
    Wagner, LA
    Schork, MA
    Bohr, DF
    Cohan, BE
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 1999, 77 (01): : 33 - 36
  • [37] Decrease in intraocular pressure following orthokeratology measured with a noncontact tonometer
    Yasuhito Ishida
    Ryoji Yanai
    Takeshi Sagara
    Teruo Nishida
    Hiroshi Toshida
    Akira Murakami
    Japanese Journal of Ophthalmology, 2011, 55 : 190 - 195
  • [38] Intraocular pressure variation by pneumatic tonometer before and after phacoemulsification
    Rodrigues, Francisco W.
    Silva, Camilla de M. N.
    Modesto, Deborah C.
    de Oliveira, Antonio W. S.
    da Silva, Rodrigo Egidio
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2018, 28 (04) : 393 - 397
  • [39] Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children
    Uzlu, Dilek
    Akyol, Nurettin
    Turk, Adem
    Gursoy, Nurcan
    Somuncu, Ahmet Mehmet
    Oruc, Yavuz
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2020, 50 (05): : 271 - 274
  • [40] Decrease in intraocular pressure following orthokeratology measured with a noncontact tonometer
    Ishida, Yasuhito
    Yanai, Ryoji
    Sagara, Takeshi
    Nishida, Teruo
    Toshida, Hiroshi
    Murakami, Akira
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2011, 55 (03) : 190 - 195