Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover During Negative Pressure Application

被引:3
|
作者
Brambilla, Enrico [1 ]
Ferguson, Tanner J. [2 ]
Chu, Nathan [1 ]
Ammar, David [3 ]
Yoo, Paul [1 ]
机构
[1] Equinox Ophthalm Inc, Newport Beach, CA USA
[2] Vance Thompson Vis, 3101 W 57th St, Sioux Falls, SD 57108 USA
[3] Lions Eye Inst Transplant & Res, Tampa, FL USA
来源
CLINICAL OPHTHALMOLOGY | 2022年 / 16卷
关键词
tonometry; intraocular pressure; TOP; multi-pressure dial; MPD; ophthalmic techniques; diagnostics; GLAUCOMA PROGRESSION;
D O I
10.2147/OPTH.S359605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This is a 2-part study to investigate the agreement between pneumatonometry and direct pressure transducer intraocular pressure ( IOP) measurements in a perfusion organ culture (POC) model where (1) the perfusion fluid column is open to atmospheric pressure, holding IOP constant to permit evaluation of the impact of negative pressure (NP) on IOP measurements, and (2) the perfusion fluid column is a closed system, allowing IOP to vary with NP application. Methods: The first part incorporated a fluid column open to atmospheric pressure, maintaining IOP constant to permit evaluation of the effect of applied NP on IOP measurement accuracy. In the second part, the POC column was closed, allowing IOP to vary with NP application and permit evaluation of agreement between pneumatonometry and pressure transducer measurements. In each part, four perfused tissues were used in thirteen paired pre-set IOP (10, 20, 25, 30 mmHg) and NP (0, 5, 10, 15, 20 mmHg) combinations, resulting in a total of 1040 paired measurements (520 per study). The difference in IOP measurements (Delta IOP = Excursion tonometry - pressure transducer) was calculated at each paired configuration. Results: During the first part, the mean Delta IOP was -0.7 +/- 1.6 mmHg across all measurements. During the second part, the mean Delta IOP across all measurements was +0.7 +/- 1.4 mmHg. At NP settings of -5, -10, -15, and -20 mmHg, across all pre-set IOPs, the mean TOP reduction via Excursion tonometry was 3.1 +/- 0.3, 5.6 +/- 1.3, 8.5 +/- 1.7 and 11.2 +/- 1.8 mmHg, respectively. Conclusion: Measurement of IOP via Excursion tonometry yields results within the accuracy range of the pneumatonometry device (per manufacturer) and is minimally impacted by NP application. The IOP-lowering results are consistent with previous studies and further support the effectiveness of the Multi-Pressure Dial in lowering IOP relative to atmospheric pressure.
引用
收藏
页码:1289 / 1300
页数:12
相关论文
共 50 条
  • [21] Performance of a tonometer for arterial pressure measurement during anesthesia in the elderly
    Harasawa K.
    Yamane M.
    Morimoto Y.
    Kemmotsu O.
    Journal of Anesthesia, 2004, 18 (4) : 316 - 317
  • [22] Intraocular pressure measurement: A review *
    Da Silva, Filipe
    Lira, Madalena
    SURVEY OF OPHTHALMOLOGY, 2022, 67 (05) : 1319 - 1331
  • [23] Factors affecting intraocular pressure measured by noncontact tonometer
    Yazici, A.
    Sen, E.
    Ozdal, P.
    Aksakal, F. N.
    Altinok, A.
    Oncul, H.
    Koklu, G.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (01) : 61 - 65
  • [24] Intraocular Pressure Measurement Using Rebound Tonometer for Deviated Angles and Positions in Human Eyes
    Takenaka, Joji
    Mochizuki, Hideki
    Kunihara, Eriko
    Tanaka, Junko
    Kiuchi, Yoshiaki
    CURRENT EYE RESEARCH, 2012, 37 (02) : 109 - 114
  • [25] Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study
    Nacaroglu, Senay Asik
    Un, Emine Seker
    Ersoz, Mehmet Giray
    Tasci, Yelda
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2015, 8 (03) : 540 - 543
  • [26] Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study
    Senay Asik Nacaroglu
    Emine Seker Un
    Mehmet Giray Ersoz
    Yelda Tasci
    International Journal of Ophthalmology, 2015, (03) : 540 - 543
  • [27] Dynamic Contour Tonometer and Goldmann Applanation Tonometer Performance in a Developing World Setting: Intraocular Pressure Measurement Acquisition and Precision
    Cheng, Jason
    Salam, Tahrina
    Russell, Paul J.
    Heath, David G.
    Kotecha, Aachal
    JOURNAL OF GLAUCOMA, 2013, 22 (09) : 736 - 739
  • [28] A portable applanation tonometer for accurate intraocular pressure measurements
    Feng, Tingting
    Chen, Xinjian
    Geng, Jialei
    Nie, Baoqing
    SENSORS AND ACTUATORS A-PHYSICAL, 2022, 344
  • [29] Measurement of intraocular pressure (IOP) in chickens using a rebound tonometer: Quantitative evaluation of variance due to position inaccuracies
    Prashar, A.
    Guggenheim, J. A.
    Erichsen, J. T.
    Hocking, Pm.
    Morgan, J. E.
    EXPERIMENTAL EYE RESEARCH, 2007, 85 (04) : 563 - 571
  • [30] Malignant glaucoma during execution of diurnal curve of intraocular pressure with dynamic contour tonometer
    Ferraz Rego, Marcos Antonio
    Lopes Lima, Francisco Eduardo
    de Avila, Marcos Pereira
    REVISTA BRASILEIRA DE OFTALMOLOGIA, 2012, 71 (05) : 285 - 288