Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software

被引:206
作者
Lee, Young Han [1 ,2 ]
Park, Kwan Kyu [3 ]
Song, Ho-Taek [1 ,2 ]
Kim, Sungjun [1 ,2 ]
Suh, Jin-Suck [1 ,2 ]
机构
[1] Yonsei Univ, Dept Radiol, Res Inst Radiol Sci, Med Convergence Res Inst,Coll Med, Seoul 120752, South Korea
[2] Yonsei Univ, Severance Biomed Sci Inst, Coll Med, Seoul 120752, South Korea
[3] Yonsei Univ, Dept Orthopaed Surg, Coll Med, Seoul 120752, South Korea
关键词
Dual-energy CT; Computed tomography; Metallic artefacts; REFORMATTED PROJECTIONS; COMPUTED-TOMOGRAPHY; HIP; REPLACEMENT; HARDWARE; IMPLANTS;
D O I
10.1007/s00330-011-2370-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To assess the usefulness of gemstone spectral imaging (GSI) dual-energy CT (DECT) with/without metal artefact reduction software (MARs). Methods The DECTs were performed using fast kV-switching GSI between 80 and 140 kV. The CT data were retro-reconstructed with/without MARs, by different displayed fields-of-view (DFOV), and with synthesised monochromatic energy in the range 40-140 keV. A phantom study of size and CT numbers was performed in a titanium plate and a stainless steel plate. A clinical study was performed in 26 patients with metallic hardware. All images were retrospectively reviewed in terms of the visualisation of periprosthetic regions and the severity of beam-hardening artefacts by using a five-point scale. Results The GSI-MARs reconstruction can markedly reduce the metal-related artefacts, and the image quality was affected by the prosthesis composition and DFOV. The spectral CT numbers of the prosthesis and periprosthetic regions showed different patterns on stainless steel and titanium plates. Conclusion Dual-energy CT with GSI-MARs can reduce metal-related artefacts and improve the delineation of the prosthesis and periprosthetic region. We should be cautious when using GSI-MARs because the image quality was affected by the prosthesis composition, energy (in keV) and DFOV. The metallic composition and size should be considered in metallic imaging with GSI-MARs reconstruction. Key Points Metal-related artefacts can be troublesome on musculoskeletal computed tomography (CT). Gemstone spectral imaging (GSI) with dual-energy CT (DECT) offers a novel solution GSI and metallic artefact reduction software (GSI-MAR) can markedly reduce these artefacts. However image quality is influenced by the prosthesis composition and other parameters. We should be aware about potential overcorrection when using GSI-MARs.
引用
收藏
页码:1331 / 1340
页数:10
相关论文
共 50 条
  • [41] Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements
    Lewis, Mark
    Reid, Karen
    Toms, Andoni P.
    SKELETAL RADIOLOGY, 2013, 42 (02) : 275 - 282
  • [42] Quantitative evaluation of beam-hardening artefact correction in dual-energy CT myocardial perfusion imaging
    Bucher, Andreas M.
    Wichmann, Julian L.
    Schoepf, U. Joseph
    Wolla, Christopher D.
    Canstein, Christian
    McQuiston, Andrew D.
    Krazinski, Aleksander W.
    De Cecco, Carlo N.
    Meinel, Felix G.
    Vogl, Thomas J.
    Geyer, Lucas L.
    EUROPEAN RADIOLOGY, 2016, 26 (09) : 3215 - 3222
  • [43] Complementary contrast media for metal artifact reduction in Dual-Energy CT
    Lambert, Jack W.
    Edic, Peter M.
    FitzGerald, Paul
    Torres, Andrew S.
    Yeh, Benjamin M.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [44] Comparison and Combination of Dual-Energy- and Iterative-Based Metal Artefact Reduction on Hip Prosthesis and Dental Implants
    Bongers, Malte N.
    Schabel, Christoph
    Thomas, Christoph
    Raupach, Rainer
    Notohamiprodjo, Mike
    Nikolaou, Konstantin
    Bamberg, Fabian
    PLOS ONE, 2015, 10 (11):
  • [45] Is metal artefact reduction mandatory in cardiac PET/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?
    Pardis Ghafarian
    S. M. R. Aghamiri
    Mohammad R. Ay
    Arman Rahmim
    Thomas H. Schindler
    Osman Ratib
    Habib Zaidi
    European Journal of Nuclear Medicine and Molecular Imaging, 2011, 38 : 252 - 262
  • [46] The impact of a metal artefact reduction algorithm on treatment planning for patients undergoing radiotherapy of the pelvis
    King, John
    Whittam, Shona
    Smith, David
    Al-Qaisieh, Bashar
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 24 : 138 - 143
  • [47] Combination of dual-energy computed tomography and iterative metal artefact reduction to increase general quality of imaging for radiotherapy patients with high dense materials. Phantom study
    Pawalowski, Bartosz
    Panek, Rafal
    Szweda, Hubert
    Piotrowski, Tomasz
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 92 - 99
  • [48] Exploring metal artifact reduction using dual-energy CT with pre-metal and post-metal implant cadaver comparison: are implant specific protocols needed?
    Wellenberg, Ruud H. H.
    Donders, Johanna C. E.
    Kloen, Peter
    Beenen, Ludo F. M.
    Kleipool, Roeland P.
    Maas, Mario
    Streekstra, Geert J.
    SKELETAL RADIOLOGY, 2018, 47 (06) : 839 - 845
  • [49] Comparison of metal artifact reduction using single-energy CT and dual-energy CT with various metallic implants in cadavers
    Barreto, Izabella
    Pepin, Eric
    Davis, Ivan
    Dean, Cooper
    Massini, Tara
    Rees, John
    Olguin, Catherine
    Quails, Nathan
    Correa, Nathalie
    Rill, Lynn
    Arreola, Manuel
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 133
  • [50] Are metal artefact reduction algorithms effective to correct cone beam CT artefacts arising from the exomass?
    Candemil, Amanda Pelegrin
    Salmon, Benjamin
    Freitas, Deborah Queiroz
    Bovi Ambrosano, Glaucia Maria
    Haiter-Neto, Francisco
    Oliveira, Matheus Lima
    DENTOMAXILLOFACIAL RADIOLOGY, 2019, 48 (03)