Characterisation of flattening filter free (FFF) beam properties for initial beam set-up and routine QA, independent of flattened beams

被引:4
作者
Paynter, D. [1 ]
Weston, S. J. [1 ]
Cosgrove, V. P. [1 ]
Thwaites, D. I. [2 ,3 ]
机构
[1] Leeds Teaching Hosp NHS Trust, Leeds Canc Ctr, Med Phys & Engn, Leeds LS9 7TF, W Yorkshire, England
[2] Univ Leeds, LIGHT Inst, Div Med Phys, Leeds LS2 9JT, W Yorkshire, England
[3] Univ Sydney, Inst Med Phys, Sch Phys, Sydney, NSW 2006, Australia
关键词
quality assurance; flattening filter free; radiotherapy; PHOTON BEAMS;
D O I
10.1088/1361-6560/aa9a1f
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Flattening filter free (FFF) beams have reached widespread use for clinical treatment deliveries. The usual methods for FFF beam characterisation for their quality assurance (QA) require the use of associated conventional flattened beams (cFF). Methods for QA of FFF without the need to use associated cFF beams are presented and evaluated against current methods for both FFF and cFF beams. Inflection point normalisation is evaluated against conventional methods for the determination of field size and penumbra for field sizes from 3 cm x 3 cm to 40 cm x 40cm at depths from dmax to 20 cm in water for matched and unmatched FFF beams and for cFF beams. A method for measuring symmetry in the cross plane direction is suggested and evaluated as FFF beams are insensitive to symmetry changes in this direction. Methods for characterising beam energy are evaluated and the impact of beam energy on profile shape compared to that of cFF beams. In-plane symmetry can be measured, as can cFF beams, using observed changes in profile, whereas cross-plane symmetry can be measured by acquiring profiles at collimator angles 0 and 180. Beam energy and 'unflatness' can be measured as with cFF beams from observed shifts in profile with changing beam energy. Normalising the inflection points of FFF beams to 55% results in an equivalent penumbra and field size measurement within 0.5 mm of conventional methods with the exception of 40 cm x 40 cm fields at a depth of 20 cm. New proposed methods are presented that make it possible to independently carry out set up and QA measurements on beam energy, flatness, symmetry and field size of an FFF beam without the need to reference to an equivalent flattened beam of the same energy. The methods proposed can also be used to carry out this QA for flattened beams, resulting in universal definitions and methods for MV beams. This is presented for beams produced by an Elekta linear accelerator, but is anticipated to also apply to other manufacturers' beams.
引用
收藏
页数:8
相关论文
共 9 条
  • [1] The cyberknife: A frameless robotic system for radiosurgery
    Adler, JR
    Chang, SD
    Murphy, MJ
    Doty, J
    Geis, P
    Hancock, SL
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 69 (1-4) : 124 - 128
  • [2] IPEM topical report 1: guidance on implementing flattening filter free (FFF) radiotherapy
    Budgell, Geoff
    Brown, Kirstie
    Cashmore, Jason
    Duane, Simon
    Frame, John
    Hardy, Mark
    Paynter, David
    Thomas, Russell
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (23) : 8360 - 8394
  • [3] Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy
    Fogliata, A.
    Garcia, R.
    Knoos, T.
    Nicolini, G.
    Clivio, A.
    Vanetti, E.
    Khamphan, C.
    Cozzi, L.
    [J]. MEDICAL PHYSICS, 2012, 39 (10) : 6455 - 6464
  • [4] Current status and future perspective of flattening filter free photon beams
    Georg, Dietmar
    Knoos, Tommy
    McClean, Brendan
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1280 - 1293
  • [5] Radiation characteristics of helical tomotherapy
    Jeraj, R
    Mackie, TR
    Balog, J
    Olivera, G
    Pearson, D
    Kapatoes, J
    Ruchala, K
    Reckwerdt, P
    [J]. MEDICAL PHYSICS, 2004, 31 (02) : 396 - 404
  • [6] Beam characteristics of energy-matched flattening filter free beams
    Paynter, D.
    Weston, S. J.
    Cosgrove, V. P.
    Evans, J. A.
    Thwaites, D. I.
    [J]. MEDICAL PHYSICS, 2014, 41 (05)
  • [7] Properties of unflattened photon beams shaped by a multileaf collimator
    Ponisch, Falk
    Titt, Uwe
    Vassiliev, Oleg N.
    Kry, Stephen F.
    Mohan, Radhe
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1738 - 1746
  • [8] Performance of a multi-axis ionization chamber array in a 1.5 T magnetic field
    Smit, K.
    Kok, J. G. M.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (07) : 1845 - 1855
  • [9] Dosimetric properties of photon beams from a flattening filter free clinical accelerator
    Vassiliev, ON
    Titt, U
    Pönisch, F
    Kry, SF
    Mohan, R
    Gillin, MT
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07) : 1907 - 1917