TRANSCRANIAL ULTRASOUND TO DETECT ELEVATED INTRACRANIAL PRESSURE: COMPARISON OF SEPTUM PELLUCIDUM UNDULATIONS AND OPTIC NERVE SHEATH DIAMETER

被引:10
作者
Bolesch, Susanne [1 ]
von Wegner, Frederic [1 ]
Senft, Christian [2 ]
Lorenz, Matthias W. [1 ]
机构
[1] Univ Hosp Frankfurt, Dept Neurol, Frankfurt, Germany
[2] Univ Hosp Frankfurt, Dept Neurosurg, Frankfurt, Germany
关键词
Intracranial pressure; Ultrasound; Intracranial hypertension; Papilledema; Bedside test; ULTRASONOGRAPHY; HYPERTENSION;
D O I
10.1016/j.ultrasmedbio.2014.12.023
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Two ultrasound tests that can be used to assess increased intracranial pressure (ICP) at the bedside are described. In outpatients receiving lumbar puncture and in intensive care patients with invasive ICP monitoring, we measured the optic nerve sheath diameter (ONSD) with transbulbar B-mode sonography and septum pellucidum undulation (SPU) induced by repeated passive head rotation with transtemporal M-mode sonography. We assessed the sensitivity and specificity of ONSD and SPU in the prediction of ICP >20 cm H2O. For ONSD, sensitivity was 53% and specificity 100% (n = 35, p < 0.001). The sensitivity of the SPU test was 75% and the specificity 100% (n = 32, p < 0.001). Although the SPU test may not feasible in some patients, it has high sensitivity and specificity comparable to those of ONSD measurement. The SPU test and ONSD may be useful alternatives to fundoscopy in clinical routine, preferably in combination. (C) 2015 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1233 / 1240
页数:8
相关论文
共 18 条
  • [1] Sonographic assessment of the optic nerve sheath in idiopathic intracranial hypertension
    Baeuerle, Jochen
    Nedelmann, Max
    [J]. JOURNAL OF NEUROLOGY, 2011, 258 (11) : 2014 - 2019
  • [2] Transorbital optic nerve sheath ultrasonography in normal children
    Ballantyne, J
    Hollman, AS
    Hamilton, R
    Bradnam, MS
    Carachi, R
    Young, DG
    Dutton, GN
    [J]. CLINICAL RADIOLOGY, 1999, 54 (11) : 740 - 742
  • [3] Becker G, 1994, J Neuroimaging, V4, P17
  • [4] Blaivas M, 2003, ACAD EMERG MED, V10, P376, DOI 10.1197/aemj.10.4.376
  • [5] Nonmydriatic Ocular Fundus Photography in the Emergency Department
    Bruce, Beau B.
    Lamirel, Cedric
    Wright, David W.
    Ward, Antoinette
    Heilpern, Katherine L.
    Biousse, Valerie
    Newman, Nancy J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2011, 364 (04) : 387 - 389
  • [6] Friedman Deborah I, 2008, Expert Rev Neurother, V8, P397, DOI 10.1586/14737175.8.3.397
  • [7] Prospective Analysis of Single Operator Sonographic Optic Nerve Sheath Diameter Measurement for Diagnosis of Elevated Intracranial Pressure
    Frumin, Erica
    Schlang, Joelle
    Wiechmann, Warren
    Hata, Stacy
    Rosen, Sasha
    Anderson, Craig
    Pare, Laura
    Rosen, Mark
    Fox, John Christian
    [J]. WESTERN JOURNAL OF EMERGENCY MEDICINE, 2014, 15 (02) : 217 - 220
  • [8] Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury
    Geeraerts, Thomas
    Launey, Yoann
    Martin, Laurent
    Pottecher, Julien
    Vigue, Bernard
    Duranteau, Jacques
    Benhamou, Dan
    [J]. INTENSIVE CARE MEDICINE, 2007, 33 (10) : 1704 - 1711
  • [9] The role of optic nerve ultrasonography in the diagnosis of elevated intracranial pressure
    Girisgin, Abdullah Sadik
    Kalkan, Erdal
    Kocak, Sedat
    Cander, Basar
    Gul, Mehmet
    Semiz, Mustafa
    [J]. EMERGENCY MEDICINE JOURNAL, 2007, 24 (04) : 251 - 254
  • [10] The subarachnoid space surrounding the optic nerves. An ultrasound study of the optic nerve sheath
    Hansen, HC
    Helmke, K
    [J]. SURGICAL AND RADIOLOGIC ANATOMY, 1996, 18 (04) : 323 - 328