Chip-based solid-phase extraction pretreatment for direct electrospray mass spectrometry analysis using an array of monolithic columns in a polymeric substrate

被引:95
|
作者
Tan, AM [1 ]
Benetton, S [1 ]
Henion, JD [1 ]
机构
[1] Cornell Univ, New York State Coll Vet Med, Ithaca, NY 14850 USA
关键词
D O I
10.1021/ac030196w
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An array of eight porous monolithic columns, prepared in a Zeonor polymeric chip by UV-initiated polymerization of butyl methacrylate and ethylene dimethacrylate, was tested for solid-phase extraction (SPE) cleanup of biological samples prior to directly coupled electrospray mass spectrometry (ESI-MS). The chip, fabricated by hot embossing and thermal bonding, consists of eight parallel channels (10 mm long, 360 mum i.d.) connected via external fused-silica capillaries. The monomer mixture was aspirated simultaneously into the eight channels using a homemade vacuum manifold device and polymerized in parallel for 20 min under UV irradiation. The porous monolithic columns were then characterized by scanning electron microscopy and evaluated by ESI-MS applications with respect to sample capacity, recovery, reproducibility of peak area or peak height ratios, and linearity between peak height ratio and concentration using imipramine as a pharmaceutical test compound. The average sample capacity was estimated to be 0.30 mug with a relative standard deviation (RSD) of 26.5% for the eight monolithic columns on the same polymeric chip. For two chips prepared using the same monomer mixture, the difference in average sample capacity was 7.0%. The average recovery for the eight monolithic SPE columns on the same chip was 79.1% with an RSD of 7.9%. Using imipramine-d3 as an internal standard, the RSD of peak height ratios for the eight different columns was 2.0% for a standard solution containing 1 mug/mL imipramine. A linear calibration curve (R-2 = 0.9995) was obtained for standard aqueous solutions of imipramine in the range from 0.025 to 10 mug/mL. To demonstrate the analytical potential of the chip-based SPE system, two different types of real-world samples including human urine sample and P450 drug metabolism incubation mixture were tested. Similar to standard aqueous solution, a linear correlation (R2 = 0.9995) was also found for human urine sample spiked with imipramine in the range of 0.025-10 mug/mL When aliquots of a human urine sample spiked with 1 mug/mL imipramine were loaded onto eight different monolithic columns, the RSD of peak height ratios was 3.8%. For a P450-imipramine incubation mixture, the formation of the N-demethylated metabolite (m/z 267.2) and the monohydroxylated metabolite (m/z 297.2) of imipramine was observed following chip-based monolithic SPE sample cleanup and preconcentration.
引用
收藏
页码:5504 / 5511
页数:8
相关论文
共 50 条
  • [1] Coupling on-chip solid-phase extraction to electrospray mass spectrometry through an integrated electrospray tip
    Yang, YN
    Li, C
    Lee, KH
    Craighead, HG
    ELECTROPHORESIS, 2005, 26 (19) : 3622 - 3630
  • [2] Liquid extraction surface analysis (LESA) of food surfaces employing chip-based nano-electrospray mass spectrometry
    Eikel, Daniel
    Henion, Jack
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (16) : 2345 - 2354
  • [3] Rapid Analysis of Biological Samples Using Monolithic Polymer-Based In-Tube Solid-Phase Microextraction with Direct Mass Spectrometry
    Hu, Wei
    Zhou, Wei
    Wang, Chenlu
    Liu, Zichun
    Chen, Zilin
    ACS APPLIED BIO MATERIALS, 2021, 4 (08): : 6236 - 6243
  • [4] Analysis of Biological Fluids by Direct Combination of Solid Phase Extraction and Desorption Electrospray Ionization Mass Spectrometry
    Denes, Julia
    Katona, Maria
    Hosszu, Adam
    Czuczy, Noemi
    Takats, Zoltan
    ANALYTICAL CHEMISTRY, 2009, 81 (04) : 1669 - 1675
  • [5] Polymeric monolithic microcartridges with gold nanoparticles for the analysis of protein biomarkers by on-line solid-phase extraction capillary electrophoresis-mass spectrometry
    Pont, Laura
    Marin, Gemma
    Vergara-Barberan, Maria
    Gagliardi, Leonardo G.
    Sanz-Nebot, Victoria
    Herrero-Martinez, Jose M.
    Benavente, Fernando
    JOURNAL OF CHROMATOGRAPHY A, 2020, 1622 (1622)
  • [6] A device for automated direct sampling and quantitation from solid-phase sorbent extraction cards by electrospray tandem mass spectrometry
    Wachs, T
    Henion, J
    ANALYTICAL CHEMISTRY, 2003, 75 (07) : 1769 - 1775
  • [7] Polymeric Materials in Speciation Analysis Based on Solid-Phase Extraction
    Dakova, Ivanka
    Yordanova, Tanya
    Karadjova, Irina
    MOLECULES, 2024, 29 (01):
  • [8] Solid-phase mini-extraction coupled to nano-electrospray ionization mass spectrometry
    Qi, LM
    Danielson, ND
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2003, 225 : U118 - U118
  • [9] Direct sampling and analysis from solid-phase extraction cards using an automated liquid extraction surface analysis nanoelectrospray mass spectrometry system
    Walworth, Matthew J.
    ElNaggar, Mariam S.
    Stankovich, Joseph J.
    Witkowski, Chuck
    Norris, Jeremy L.
    Van Berkel, Gary J.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (17) : 2389 - 2396
  • [10] Isotope tracing-assisted chip-based solid-phase extraction mass spectrometry for monitoring metabolic changes and vitamin D3 regulation in cells
    Xu, Ning
    Lin, Haifeng
    Ding, Xiaodan
    Wang, Peilong
    Lin, Jin-Ming
    TALANTA, 2025, 288