MR thermometry;
proton-resonance frequency shift;
thermal calibration;
ex vivo tissues;
D O I:
10.1002/mrm.1910400316
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The temperature sensitivity of the proton-resonance frequency (PRF) has proven valuable for the monitoring of MR image-guided thermal coagulation therapy. However, there is significant inconsistency in reported values of the PRF-thermal coefficient, as measured from experiments encompassing a range of in vivo and ex vivo tissue types and experimental conditions. A method of calibrating the temperature dependence of the PRF is described and results are presented that indicate a tissue-type independence. To this end, other possible mechanisms for variations in the PRF-thermal coefficient are suggested, including physiological perturbations and volume magnetic susceptibility effects from geometry and orientation.