Mass-based readout for agglutination assays

被引:9
作者
Chunara, Rumi
Godin, Michel
Knudsen, Scott M.
Manalis, Scott R. [1 ]
机构
[1] Harvard Univ, MIT, Div Hlth Sci & Technol, Cambridge, MA 02139 USA
[2] MIT, Biol Engn Dept, Cambridge, MA 02139 USA
[3] MIT, Biol Engn & Mech Engn Dept, Cambridge, MA 02139 USA
基金
美国国家卫生研究院; 加拿大自然科学与工程研究理事会;
关键词
D O I
10.1063/1.2806197
中图分类号
O59 [应用物理学];
学科分类号
摘要
We present a mass-based readout for agglutination assays. The suspended microchannel resonator (SMR) is used to classify monomers and dimers that are formed during early stage aggregation, and to relate the total count to the analyte concentration. Using a model system of streptavidin functionalized microspheres and biotinylated antibody as the analyte, we obtain a dose-response curve over a concentration range of 0.63-630 nM and show that the results are comparable to what has been previously achieved by image analysis and conventional flow cytometry.
引用
收藏
页数:3
相关论文
共 10 条
[1]   Laser light scattering immunoassay: An improved data analysis by CONTIN method [J].
Antony, T ;
Saxena, A ;
Roy, KB ;
Bohidar, HB .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1998, 36 (2-3) :75-85
[2]   Weighing of biomolecules, single cells and single nanoparticles in fluid [J].
Burg, Thomas P. ;
Godin, Michel ;
Knudsen, Scott M. ;
Shen, Wenjiang ;
Carlson, Greg ;
Foster, John S. ;
Babcock, Ken ;
Manalis, Scott R. .
NATURE, 2007, 446 (7139) :1066-1069
[3]   Quantitative detection of C-reactive protein using phosphocholine-labelled enzyme or microspheres [J].
Deegan, O ;
Walshe, K ;
Kavanagh, K ;
Doyle, S .
ANALYTICAL BIOCHEMISTRY, 2003, 312 (02) :175-181
[4]  
Eda S, 1998, J CLIN LAB ANAL, V12, P137, DOI 10.1002/(SICI)1098-2825(1998)12:3<137::AID-JCLA2>3.0.CO
[5]  
2-6
[6]  
FERNANDO SA, 1992, J IMMUNOL METHODS, V151, P47
[7]   Ultraviolet and visible light spectrophotometric approach to blood typing: objective analysis by agglutination index [J].
Narayanan, S ;
Orton, S ;
Leparc, GF ;
Garcia-Rubio, LH ;
Potter, RL .
TRANSFUSION, 1999, 39 (10) :1051-1059
[8]   Counting and sizing of particles and particle agglomerates in a microfluidic device using laser light scattering: application to a particle-enhanced immunoassay [J].
Pamme, N ;
Koyama, R ;
Manz, A .
LAB ON A CHIP, 2003, 3 (03) :187-192
[9]  
PLOTZ CM, 2003, AM J MED, V312, P175
[10]   Fluorescence-microscopy-based image analysis for analyte-dependent particle doublet detection in a single-step immuno agglutination assay [J].
Wiklund, M ;
Nord, O ;
Gothäll, R ;
Chernyshev, AV ;
Nygren, PÅ ;
Hertz, HM .
ANALYTICAL BIOCHEMISTRY, 2005, 338 (01) :90-101