Evaluation of a New CT-Guided Robotic System for Percutaneous Needle Insertion for Thermal Ablation of Liver Tumors: A Prospective Pilot Study

被引:48
作者
de Baere, Thierry [1 ]
Roux, Charles [1 ]
Deschamps, Frederic [1 ]
Tselikas, Lambros [1 ]
Guiu, Boris [2 ]
机构
[1] Univ Paris Saclay, Dept Intervent Radiol, BIOTHERIS, Gustave Roussy, 114 Rue Edouard Vaillant, F-94805 Villejuif, France
[2] St Eloi Univ Hosp, Dept Radiol, Montpellier, France
关键词
Robotic navigation; Needle navigation; Computed tomography; Liver cancer; Thermal ablation; RADIOFREQUENCY ABLATION; MICROWAVE ABLATION; CLASSIFICATION; ASSISTANCE; STANDARDS;
D O I
10.1007/s00270-022-03267-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose To assess the feasibility and safety of a robotic system for percutaneous needle insertion during thermal ablation of liver tumors. Materials and Methods This study analyzed the CT-guided percutaneous needle insertion using the EPIONE robotic device (Quantum Surgical, Montpellier, France) for radiofrequency or microwave liver ablation. The main criteria of the study were feasibility (possibility to perform the thermal ablation after needle insertion), the number of needle adjustments (reiteration of robotically assisted needle insertion when initial needle positioning is considered insufficient to perform ablation), and robotic-guided procedure safety (complications related to the needle insertion). Patients were followed up at 6 months post-intervention to assess local tumor control. Results Twenty-one patients with 24 tumors, including 6 HCC and 18 metastases measuring 15.6 +/- 7.2 mm, were enrolled. One patient (with one tumor) was excluded for protocol deviation. Robotic assisted thermal ablation was feasible for 22/23 lesions (95.7%) and 19/20 patients (95.0%), as validated by a data safety monitoring Board (95% CI [76.39%; 99.11%]) for the per-protocol population. The mean number of needle adjustments per tumor treated was 0.4 (SD: 0.7), with 70.8% of tumors requiring no adjustment. No adverse events were depicted. Rate of local tumor control was 83.3% for patients and 85.7% for tumors, at 6 months. Conclusion This bicentric first-in-human pilot study suggests both feasibility and safety of a stereotactic CT-guided EPIONE device for the percutaneous needle insertion during liver tumor thermal ablation.
引用
收藏
页码:1701 / 1709
页数:9
相关论文
共 27 条
[1]   Robotic-assisted thermal ablation of liver tumours [J].
Abdullah, Basri Johan Jeet ;
Yeong, Chai Hong ;
Goh, Khean Lee ;
Yoong, Boon Koon ;
Ho, Gwo Fuang ;
Yim, Carolyn Chue Wai ;
Kulkarni, Anjali .
EUROPEAN RADIOLOGY, 2015, 25 (01) :246-257
[2]   Local Treatment of Breast Cancer Liver Metastasis [J].
Bale, Reto ;
Putzer, Daniel ;
Schullian, Peter .
CANCERS, 2019, 11 (09)
[3]   Stereotactic Radiofrequency Ablation of Hepatocellular Carcinoma: a Histopathological Study in Explanted Livers [J].
Bale, Reto ;
Schullian, Peter ;
Eberle, Gernot ;
Putzer, Daniel ;
Zoller, Heinz ;
Schneeberger, Stefan ;
Manzl, Claudia ;
Moser, Patrizia ;
Oberhuber, Georg .
HEPATOLOGY, 2019, 70 (03) :840-850
[4]   Stereotactically navigated percutaneous microwave ablation (MWA) compared to conventional MWA: a matched pair analysis [J].
Beyer, L. P. ;
Luerken, L. ;
Verloh, N. ;
Haimerl, M. ;
Michalik, K. ;
Schaible, J. ;
Stroszczynski, C. ;
Wiggermann, P. .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2018, 13 (12) :1991-1997
[5]   Stereotactically-navigated percutaneous Irreversible Electrop oration (IRE) compared to conventional IRE: a prospective trial [J].
Beyer, Lukas P. ;
Pregler, Benedikt ;
Niessen, Christoph ;
Schicho, Andreas ;
Haimerl, Michael ;
Jung, Ernst Michael ;
Stroszczynski, Christian ;
Wiggermann, Philipp .
PEERJ, 2016, 4
[6]   CIRSE Standards of Practice on Thermal Ablation of Liver Tumours [J].
Crocetti, Laura ;
de Baere, Thierry ;
Pereira, Philippe L. ;
Tarantino, Francesco Paolo .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 43 (07) :951-962
[7]   Robotic assistance for percutaneous needle insertion in the kidney: preclinical proof on a swine animal model [J].
de Baere, Thierry ;
Roux, Charles ;
Noel, Guillaume ;
Delpla, Alexandre ;
Deschamps, Frederic ;
Varin, Eloi ;
Tselikas, Lambros .
EUROPEAN RADIOLOGY EXPERIMENTAL, 2022, 6 (01)
[8]   Computer assisted electromagnetic navigation improves accuracy in computed tomography guided interventions: A prospective randomized clinical trial [J].
Durand, Pierre ;
Moreau-Gaudry, Alexandre ;
Silvent, Anne-Sophie ;
Frandon, Julien ;
Chipon, Emilie ;
Medici, Maud ;
Bricault, Ivan .
PLOS ONE, 2017, 12 (03)
[9]   Stereotactic CT-Guided Percutaneous Microwave Ablation of Liver Tumors With the Use of High-Frequency Jet Ventilation: An Accuracy and Procedural Safety Study [J].
Engstrand, Jennie ;
Toporek, Grzegorz ;
Harbut, Piotr ;
Jonas, Eduard ;
Nilsson, Henrik ;
Freedman, Jacob .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 208 (01) :193-200
[10]   Cirse Quality Assurance Document and Standards for Classification of Complications: The Cirse Classification System [J].
Filippiadis, D. K. ;
Binkert, C. ;
Pellerin, O. ;
Hoffmann, R. T. ;
Krajina, A. ;
Pereira, P. L. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 40 (08) :1141-1146