A Field-Programmable Lab-on-a-Chip with Built-in Self-Test Circuit and Low-Power Sensor-Fusion Solution in 0.35 μm Standard CMOS Process

被引:0
作者
Lai, Kelvin Yi-Tse [1 ,2 ]
Shiu, Ming-Feng [1 ,2 ]
Lu, Yi-Wen [1 ,2 ]
Ho, Yingchieh [3 ]
Kao, Yu-Chi [1 ,2 ]
Yang, Yu-Tao [1 ,2 ]
Wang, Gary [1 ,2 ]
Liu, Keng-Ming [3 ]
Chang, Hsie-Chia [1 ,2 ]
Lee, Chen-Yi [1 ,2 ]
机构
[1] Natl Chiao Tung Univ, Dept Elect Engn, Hsinchu, Taiwan
[2] Natl Chiao Tung Univ, Inst Elect, Hsinchu, Taiwan
[3] Natl Dong Hwa Univ, Dept Elect Engn, Hualien, Taiwan
来源
2015 IEEE ASIAN SOLID-STATE CIRCUITS CONFERENCE (A-SSCC) | 2015年
关键词
EWOD; MEDA; FPLOC; BIST; CDC;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents a field programmable lab-on-a-chip (FPLOC) with microfluidics actuation, droplet location/category readback, built-in-self-test (BIST), and bioassays detection. This FPLOC is composed of 1,800 MEDA microelectrodes and integrates a set of microfluidic operations to perform droplet moving/cutting/mixing. Besides, capacitive sensing circuits are considered for system automation, droplets classification, malfunctioning microelectrodes and surface flatness testing, and biomedical detection. Implemented in 0.35 mu m standard CMOS process in 7.4mm(2), the FPLOC demonstrates microfluidic operations under 25V at 1KHz, 1.3fF resolution for droplet location/category/chip BIST maps readback at 1MHz, and 0.34fF sensitivity of capacitive measurement window for biomedical detection at 1KHz.
引用
收藏
页码:281 / 284
页数:4
相关论文
共 13 条
  • [1] Danneels H., 2011, IEEE ESSCIRC
  • [2] Fair R. B., 2007, MICROFLUID NANOFLUID
  • [3] Haeberle S., 2007, LAB CHIP
  • [4] Hugo De Man, 2005, IEEE ISSCC
  • [5] Kelvin Yi-Tse Lai, 2013, IEEE ASSCC
  • [6] Kelvin Yi-Tse Lai, 2014, IEEE ASSCC
  • [7] Lai K. Y.-T., 2014, JSPS
  • [8] Lee Chen-Yi, 2014, IEEE VLSI DAT
  • [9] Lin Luan, 2008, IEEE SENSORS
  • [10] Tan Zhichao, 2012, IEEE JSSC