Wavelength modulation based surface plasmon resonance sensor for detection of cardiac marker proteins troponin I and troponin T

被引:32
作者
Sinha, Rajeev K. [1 ]
机构
[1] Manipal Acad Higher Educ, Dept Atom & Mol Phys, Manipal 576104, Karnataka, India
关键词
Surface plasmon resonance; Wavelength modulation; Cardiac troponin I; Cardiac troponin T; SELF-ASSEMBLED MONOLAYERS; IMMUNOSENSOR; SPECTROMETRY; SENSITIVITY; ABILITY;
D O I
10.1016/j.sna.2021.113104
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A wavelength modulation based surface plasmon resonance (SPR) sensor setup was developed with inexpensive components. The developed setup is utilized for the detection of cardiac marker proteins, cardiac troponin I (CTnI), and cardiac troponin T (CTnT). A simple approach was used for the fabrication of CTnI/ CTnT antibody sensor chip. A self-assembled monolayer (SAM) of 1-octanethiol and 11-mercaptoundecanoic acid (11-MUA) was formed on annealed Au thin film. The -COOH group of 11-MUA was modified with EDCNHS for amine coupling to monoclonal antibodies of CTnI and CTnT. The developed SPR setup and the sensor chip with monoclonal antibodies showed good performance for the detection of CTnI and CTnT in direct assay. For CTnI, the lowest concentration detected was 0.03125 ng/mL whereas for CTnT, it was 0.0625 mu g/mL. The detected concentrations are well below the onset concentration of these two proteins in blood following the acute myocardial infarction (AMI). The performance of this direct assay based on the developed setup and sensor chip therefore can be used easily for detection of CTnI and CTnT during their onset period. (c) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:7
相关论文
共 36 条
  • [1] Troponin elevation in coronary vs. non-coronary disease
    Agewall, S.
    Giannitsis, E.
    Jernberg, T.
    Katus, H.
    [J]. EUROPEAN HEART JOURNAL, 2011, 32 (04) : 404 - 411B
  • [2] Development of a specific troponin I detection system with enhanced immune sensitivity using a single monoclonal antibody
    Bozdogan, Anil
    El-Kased, Reham F.
    Jungbluth, Vanessa
    Knoll, Wolfgang
    Dostalek, Jakub
    Kasry, Amal
    [J]. ROYAL SOCIETY OPEN SCIENCE, 2020, 7 (10):
  • [3] Grafting of benzylic amide macrocycles onto acid-terminated self-assembled monolayers studied by XPS, RAIRS, and contact angle measurements
    Cecchet, F
    Pilling, M
    Hevesi, L
    Schergna, S
    Wong, JKY
    Clarkson, GJ
    Leigh, DA
    Rudolf, P
    [J]. JOURNAL OF PHYSICAL CHEMISTRY B, 2003, 107 (39) : 10863 - 10872
  • [4] Molecularly imprinted polymers coupled to matrix assisted laser desorption ionization mass spectrometry for femtomoles detection of cardiac troponin I peptides
    Cenci, Lucia
    Anesi, Andrea
    Busato, Mirko
    Guella, Graziano
    Bossi, Alessandra Maria
    [J]. JOURNAL OF MOLECULAR RECOGNITION, 2016, 29 (01) : 41 - 50
  • [5] Self-referencing SPR biosensing with an ultralow limit-of-detection using long-wavelength excitation
    Chen, Shimeng
    Liu, Yun
    Yu, Qingxu
    Peng, Wei
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2021, 327
  • [6] Detection of cardiac troponin-I by optic biosensors with immobilized anti-cardiac troponin-I monoclonal antibody
    Cimen, Duygu
    Bereli, Nilay
    Gunaydin, Serdar
    Denizli, Adil
    [J]. TALANTA, 2020, 219
  • [7] An SPR immunosensor for human cardiac troponin T using specific binding avidin to biotin at carboxymethyldextran-modified gold chip
    Dutra, Rosa Fireman
    Kubota, Lauro Tatsuo
    [J]. CLINICA CHIMICA ACTA, 2007, 376 (1-2) : 114 - 120
  • [8] Evaluation of PAMAM Dendrimers (G3, G4, and G5) in the Construction of a SPR-based Immunosensor for Cardiac Troponin T
    Giarola, Juliana de F.
    Souto, Denio E. P.
    Kubota, Lauro T.
    [J]. ANALYTICAL SCIENCES, 2021, 37 (07) : 1007 - 1013
  • [9] On the sensitivity of surface plasmon resonance sensors with spectral interrogation
    Homola, J
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1997, 41 (1-3) : 207 - 211
  • [10] Huang Y. H., 2012, Advances in Optical Technologies, DOI 10.1155/2012/471957