Role of tube size and intranasal compression of the nasotracheal tube in respiratory pressure loss during nasotracheal intubation: a laboratory study

被引:3
作者
Futagawa, Koichi [1 ]
Takasugi, Yoshihiro [2 ]
Kobayashi, Takeharu [3 ]
Morishita, Satoshi [3 ]
Okuda, Takahiko [1 ]
机构
[1] Kindai Univ, Dept Anesthesiol, Nara Hosp, Fac Med, 1248-1 Otodacho, Nara 6300293, Japan
[2] Kindai Univ, Dept Anesthesiol, Fac Med, 377-2 Ohno Higashi, Osaka, Osaka 5898511, Japan
[3] Daiken Med Co Ltd, Prod Dev Res Ctr, 2-6-2,Ayumino, Izumi, Osaka 5941157, Japan
关键词
Endotracheal tube; Pressure loss; Slip joint; Turbulent flow; Nasotracheal intubation; PEDIATRIC ENDOTRACHEAL-TUBES; INTRATRACHEAL PRESSURE; ADDITIONAL WORK; RESISTANCE; FLOW; VENTILATORS; CIRCUITS; DESIGN;
D O I
10.1186/s12871-017-0432-1
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Small nasotracheal tubes (NTTs) and intranasal compression of the NTT in the nasal cavity may contribute to increasing airway resistance. Since the effects of size, shape, and partial compression of the NTT on airway resistance have not been investigated, values of airway resistance with partial compression of preformed NTTs of various sizes were determined. Methods: To determine the factors affecting the respiratory pressure loss during the nasotracheal intubation, physical and fluid dynamics simulations were used. The internal minor axes of NTTs in the nasal cavity of intubated patients were measured using dial calipers. In physical and fluid dynamics simulations, pressure losses through the tubular parts, compressed parts, and slip joints of NTTs with internal diameters (IDs) of 6.0, 6.5, 7.0, 7.5, and 8.0 mm were estimated under partial compression. Results: The median internal minor axes of the 7.0-and 7.5-mm ID NTTs in the nasal cavity were 5.2 (4.3-5.6) mm and 6. 0 (4.2-7.0) mm, respectively. With a volumetric air flow rate of 30 L/min, pressure losses through uncompressed NTTs with IDs of 6.0-, 6.5-, 7.0-, 7.5-and 8.0-mm were 651.6 +/- 5.7 (6.64 +/- 0.06), 453.4 +/- 3.9 (4.62 +/- 0.04), 336.5 +/- 2.2 (3.43 +/- 0.02), 225. 2 +/- 0.2 (2.30 +/- 0.00), and 179.0 +/- 1.1 Pa (1.82 +/- 0.01 cmH2O), respectively; the pressure losses through the slip joints were 220.3 (2.25), 131.1 (1.33), 86.8 (0.88), 57.1 (0.58), and 36.1 Pa (0.37 cmH2O), respectively; and the pressure losses through the curvature of the NTT were 71.6 (0.73), 69.0 (0.70), 64.8 (0.66), 32.5 (0.33), and 41.6 Pa (0.42 cmH(2)O), respectively. A maximum compression force of 34.1 N increased the pressure losses by 82.0 (0.84), 38.0 (0.39), 23.5 (0.24), 16.6 (0.17), and 9.3 Pa (0.09 cmH(2)O), respectively. Conclusion: Pressure losses through NTTs are in inverse proportion to the tubes' IDs; greater pressure losses due to slip joints, acute bending, and partial compression of the NTT were obvious in small NTTs. Pressure losses through NTTs, especially in small NTTs, could increase the work of breathing to a greater extent than that through standard tubes; intranasal compression further increases the pressure loss.
引用
收藏
页数:10
相关论文
共 32 条
[11]   Estimation of inspiratory pressure drop in neonatal and pediatric endotracheal tubes [J].
Jarreau, PH ;
Louis, B ;
Dassieu, G ;
Desfrere, L ;
Blanchard, PW ;
Moriette, G ;
Isabey, D ;
Harf, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 87 (01) :36-46
[12]  
Kamat SS, 2016, PRACTICAL APPL MECH, P372
[13]  
Katz I. M., 2012, ACP-APPL CARDIOPUL P, V16, P5
[14]   Thermosoftening treatment of the nasotracheal tube before intubation can reduce epistaxis and nasal damage [J].
Kim, YC ;
Lee, SH ;
Noh, GJ ;
Cho, SY ;
Yeom, JH ;
Shin, WO ;
Lee, DH ;
Ryu, JS ;
Park, YS ;
Cha, KJ ;
Lee, SC .
ANESTHESIA AND ANALGESIA, 2000, 91 (03) :698-701
[15]   The influence of endotracheal tube tip design on nasal trauma during nasotracheal intubation: Magill-tip versus Murphy-tip [J].
Lee, JH ;
Kim, CH ;
Bahk, JH ;
Park, KS .
ANESTHESIA AND ANALGESIA, 2005, 101 (04) :1226-1229
[16]   Good short-term agreement between measured and calculated tracheal pressure [J].
Lichtwarck-Aschoff, M ;
Helmer, A ;
Kawati, R ;
Lattuada, M ;
Sjöstrand, UH ;
Zügel, N ;
Guttmann, J ;
Hedenstierna, G .
BRITISH JOURNAL OF ANAESTHESIA, 2003, 91 (02) :239-248
[17]   The use of a nasogastric tube to facilitate nasotracheal intubation: a randomised controlled trial [J].
Lim, C. -W. ;
Min, S. -W. ;
Kim, C. -S. ;
Chang, J. -E. ;
Park, J. -E. ;
Hwang, J. -Y. .
ANAESTHESIA, 2014, 69 (06) :591-597
[18]   USE OF THE BLASIUS RESISTANCE FORMULA TO ESTIMATE THE EFFECTIVE DIAMETER OF ENDOTRACHEAL-TUBES [J].
LOFASO, F ;
LOUIS, B ;
BROCHARD, L ;
HARF, A ;
ISABEY, D .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1992, 146 (04) :974-979
[19]  
Lu P P, 1998, Acta Anaesthesiol Sin, V36, P193
[20]   Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? [J].
Maeda, Y ;
Fujino, Y ;
Uchiyama, A ;
Taenaka, N ;
Mashimo, T ;
Nishimura, M .
CRITICAL CARE, 2003, 7 (05) :R92-R97