Segmented T1 weighted echo-planar imaging in the abdomen and pelvis. Comparison with T1 TSE and T2 UTSE sequences

被引:0
作者
Maubon, A [1 ]
Berger, V [1 ]
Aubas, P [1 ]
Ferru, JM [1 ]
Rouanet, JP [1 ]
机构
[1] CMC Beausoleil, Serv Imagerie Med, F-34000 Montpellier, France
来源
JOURNAL DE RADIOLOGIE | 1999年 / 80卷 / 03期
关键词
fast imaging; MRI; pulse sequences; comparison; abdomen; pelvis;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose:To assess, quantitatively and qualitatively, the diagnostic value of a segmented EPI T1W sequence compared to T1W and T2W TSE sequences. Material and methods: A prospective analysis of abdominal and pelvic MRI examinations of 70 patients (44 women, 26 men, mean age of 61 years), was performed on a 0.5 T supraconductive magnet with 15 mT/m gradients. The sequences were randomized and compared in a blinded fashion by 3 independent reviewers :TSE TIW (TR/TE = 500/ 12 ms, NSA = 6, turbo factor 5, 3 : 49 min), EPI T1W (TR/TE = 500/30 ms, NSA = 6, EPI factor = 7, 2 : 13 min) and UTSE T2W (TR/TE = 1600-2500/100, NSA = 6, turbo factor = 31, 2 : 20 min). Results : Quantitatively, no significant difference was found between T1W sequences for signal to noise ratio. The EPI T1W sequence had lower signal but stronger enhancement after gadolinium injection. Qualitatively, EPI TIW had significantly less flow artefacts (p < 0.001, wilcoxon test), and more chemical shift artifact (p < 0.01). For lesion detection, differences were not statistically significant between T1W sequences or between paired T1W and T2W sequences (sensitivity and specificity 84 and 86 % for TSE T1W 76 and 86 % for EPI T1W, 78 and 79 % for UTSE T2W, 90 and 65 % for TSE T1W-UTSE T2W, 88 and 65 % for EPI T1W-UTSE T2W). Kappa concordance test (0.686) and Mac Nemar symmetry test (3.55) were high between T1W sequences. Conclusion :The segmented EPI T1W sequence used had equivalent results compared to the TSE T1W sequence, it allows a 40 % reduction in acquisition time and this without difference in the diagnostic performances of the reviewers.
引用
收藏
页码:291 / 296
页数:6
相关论文
共 23 条
[1]  
BITTOUN J, 1990, ENCY MED CHIR
[2]   INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
THOMPSON, RM ;
JACK, CR .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) :67-72
[3]   ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
FELMLEE, JP ;
GRIMM, RC .
RADIOLOGY, 1993, 189 (01) :259-264
[4]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[5]   ECHO-PLANAR IMAGING [J].
DELAPAZ, RL .
RADIOGRAPHICS, 1994, 14 (05) :1045-1058
[6]   SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST [J].
EDELMAN, RR ;
WALLNER, B ;
SINGER, A ;
ATKINSON, DJ ;
SAINI, S .
RADIOLOGY, 1990, 177 (02) :515-521
[7]   ECHO-PLANAR MR-IMAGING [J].
EDELMAN, RR ;
WIELOPOLSKI, P ;
SCHMITT, F .
RADIOLOGY, 1994, 192 (03) :600-612
[8]  
FERMANIAN J, 1984, REV EPIDEMIOL SANTE, V32, P140
[9]   LARGE SAMPLE STANDARD ERRORS OF KAPPA AND WEIGHTED KAPPA [J].
FLEISS, JL ;
COHEN, J ;
EVERITT, BS .
PSYCHOLOGICAL BULLETIN, 1969, 72 (05) :323-&
[10]   Liver masses: Replacement of conventional T2-weighted spin-echo MR imaging with breath-hold MR imaging [J].
Gaa, J ;
Hatabu, H ;
Jenkins, RL ;
Finn, JP ;
Edelman, RR .
RADIOLOGY, 1996, 200 (02) :459-464