Microstrip Transmission Line RF Coil for a PET/MRI Insert

被引:8
作者
Akram, Md Shahadat Hossain [1 ]
Obata, Takayuki [1 ]
Yamaya, Taiga [1 ]
机构
[1] Natl Inst Quantum & Radiol Sci & Technol, Natl Inst Radiol Sci, Chiba 2638555, Japan
基金
日本学术振兴会;
关键词
parallel transmission RF coil; microstrip transmission line; PET-strip coil; magnetic resonance imaging; PET insert; ARRAY; PET/CT; BRAIN; CHALLENGES; PROTOTYPE; MRI;
D O I
10.2463/mrms.mp.2019-0137
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: We proposed and developed a new microstrip transmission line radiofrequency (RF) coil for a positron emission tomography (PET) insert for MRI, which has low electrical interactions with PET shield boxes. We performed imaging experiments using a single-channel and a four-channel proposed RF coils for proof-of-concept. Methods: A conventional microstrip coil consists of a microstrip conductor, a ground conductor, and a dielectric between the two conductors. We proposed a microstrip coil for the PET insert that replaced the conventional single-layer ground conductor with the RF shield of the PET insert. A dielectric material, which could otherwise attenuate gamma photons radiated from the PET imaging tracer, was not used. As proof-of-concept, we compared conventional and the proposed single-channel coils. To study multichannel performance, we further developed a four-channel proposed RF coil. Since the MRI system had a single-channel transmission port, an interfacing four-way RF power division circuit was designed. The coils were implemented as both RF transmitters and receivers in a cylindrical frame of diameter 150 mm. Coil bench performances were tested with a network analyzer (Rohde & Schwarz, Germany), and a homogeneous phantom study was conducted for gradient echo imaging and RF field (B-1) mapping in a 3T clinical MRI system (Verio, Siemens, Erlangen, Germany). Results: For all coils, the power reflection coefficient was below -30 dB, and the transmission coefficients in the four-channel configuration were near or below -20 dB. The comparative single-channel coil study showed good similarity between the conventional and proposed coils. The gradient echo image of the four-channel coil showed expected flashing image intensity near the coils and no phase distortion was visible. Transmit B-1 field map resembled the image performance. Conclusion: The proposed PET-microstrip coil performed similarly to the conventional microstrip transmission line coil and is promising for the development of a compact coil-PET system capable of simultaneous PET/MRI analysis with an existing MRI system.
引用
收藏
页码:147 / 153
页数:7
相关论文
共 50 条
[31]   A head coil system with an integrated orbiting transmission point source mechanism for attenuation correction in PET/MRI [J].
Renner, A. ;
Rausch, I. ;
Gonzalez, J. Cal ;
Frass-Kriegl, R. ;
de lara, L. Navarro ;
Sieg, J. ;
Laistler, E. ;
Glanzer, M. ;
Dungl, D. ;
Moser, E. ;
Beyer, T. ;
Figl, M. ;
Birkfellner, W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (22)
[32]   Vibration and acoustic noise characterization of a gradient coil insert in a 4 T MRI [J].
Mechefske, CK ;
Yao, G ;
Wang, FL .
JOURNAL OF VIBRATION AND CONTROL, 2004, 10 (06) :861-880
[33]   Initial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology [J].
Schug, David ;
Lerche, Christoph ;
Weissler, Bjoern ;
Gebhardt, Pierre ;
Goldschmidt, Benjamin ;
Wehner, Jakob ;
Dueppenbecker, Peter Michael ;
Salomon, Andre ;
Hallen, Patrick ;
Kiessling, Fabian ;
Schulz, Volkmar .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (07) :2851-2878
[34]   Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil [J].
Aizaz, Mueez ;
van der Pol, Jochem A. J. ;
Wierts, Roel ;
Zwart, Hans ;
van der Werf, Abe J. ;
Wildberger, Joachim E. ;
Bucerius, Jan A. ;
Moonen, Rik P. M. ;
Kooi, Marianne Eline .
JOURNAL OF CLINICAL MEDICINE, 2022, 11 (09)
[35]   Variation of MRI Coil Uniformity Caused by μPET Shieldings [J].
Solis-Najera, S. E. ;
Tomasi, D. ;
Rodriguez, A. O. .
MEDICAL PHYSICS, 2012, 1494 :50-53
[36]   Microstrip transmission line for soil moisture measurement [J].
Chen, XM ;
Li, J ;
Liang, R ;
Sun, RJ ;
Liu, CR ;
Rogers, R ;
Claros, G .
SENSORS FOR HARSH ENVIRONMENTS, 2004, 5590 :84-91
[37]   SAFIR-I: Design and Performance of a High-Rate Preclinical PET Insert for MRI [J].
Bebie, Pascal ;
Becker, Robert ;
Commichau, Volker ;
Debus, Jan ;
Dissertori, Gunther ;
Djambazov, Lubomir ;
Eleftheriou, Afroditi ;
Fischer, Jannis ;
Fischer, Peter ;
Ito, Mikiko ;
Khateri, Parisa ;
Lustermann, Werner ;
Ritzer, Christian ;
Ritzert, Michael ;
Roser, Ulf ;
Tsoumpas, Charalampos ;
Warnock, Geoffrey ;
Weber, Bruno ;
Wyss, Matthias T. ;
Zagozdzinska-Bochenek, Agnieszka .
SENSORS, 2021, 21 (21)
[38]   PET performance and MRI compatibility evaluation of a digital, ToF-capable PET/MRI insert equipped with clinical scintillators [J].
Schug, David ;
Wehner, Jakob ;
Dueppenbecker, Peter Michael ;
Weissler, Bjoern ;
Gebhardt, Pierre ;
Goldschmidt, Benjamin ;
Salomon, Andre ;
Kiessling, Fabian ;
Schulz, Volkmar .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (18) :7045-7067
[39]   Quadrature RF coil for in Vivo brain MRI of a macaque monkey in a stereotaxic head frame [J].
Roopnariane, Colin A. ;
Ryu, Yeun-Chul ;
Tofighi, Mohammad-Reza ;
Miller, Patti A. ;
Oh, Sukhoon ;
Wang, Jianli ;
Park, Bu Sik ;
Ansel, Lukas ;
Lieu, Christopher A. ;
Subramanian, Thyagarajan ;
Yang, Qing X. ;
Collins, Christopher M. .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2012, 41B (01) :22-27
[40]   Towards Brain MRI Adaptable to Head Size: Bowing RF Coil Phased Arrays [J].
Mathieu, William ;
Popovic, Milica ;
Farivar, Reza .
IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2022, 6 (03) :305-310