A converging wall-jet instrument for ion selective electrode dynamic response studies

被引:3
作者
Clay, ML
Pierce, RE
Young, VY
Hoke, S
机构
[1] Univ Florida, Dept Chem, Gainesville, FL 32611 USA
[2] USA, Ctr Hlth Promot & Prevent Med, Aberdeen Proving Ground, MD 21010 USA
关键词
D O I
10.1080/10739149808000914
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ion selective electrodes are used to conduct real-time measurements of gaseous thermal degradation products from halon replacement experiments. Response characteristics of these ion selective electrodes are needed to address potential interferences,and surface poisoning experienced under field conditions. The design and the fabrication of an in situ, converging wall-jet instrument for the measurement of the dynamic responses of ion selective electrodes are described. Various time delays associated with the generation and detection of ion activity steps have been quantitatively determined. Dynamic response measurements for iodide steps on a Coming iodide/cyanide ion selective electrode are comparable to those obtained on the best of the earlier instrument designs, the in situ, switched wall-jet instrument. The advantages of the converging wall-jet instrument over the switched wall-jet instrument are identified.
引用
收藏
页码:461 / 472
页数:12
相关论文
共 17 条
[1]   SIMULTANEOUS DETERMINATION OF BROMIDE AND CHLORIDE IN NATURAL-WATERS BY ION-EXCHANGE CHROMATOGRAPHY AND DIRECT POTENTIOMETRY WITH AN ION-SELECTIVE ELECTRODE [J].
AKAIWA, H ;
KAWAMOTO, H ;
OSUMI, M .
TALANTA, 1982, 29 (08) :689-690
[2]   COUPLED DIFFUSION ADSORPTION MODEL FOR RESPONSE OF PRECIPITATE-BASED IODIDE-SELECTIVE ELECTRODES TO PRIMARY-ION ACTIVITY STEPS [J].
BERUBE, TR ;
BUCK, RP ;
LINDNER, E ;
TOTH, K ;
PUNGOR, E .
ANALYTICAL CHEMISTRY, 1991, 63 (10) :946-953
[3]   THEORETICAL INTERPRETATION OF TRANSIENT SIGNALS OBTAINED WITH PRECIPITATE-BASED ION-SELECTIVE ELECTRODES IN THE PRESENCE OF INTERFERING IONS [J].
GRATZL, M ;
LINDNER, E ;
PUNGOR, E .
ANALYTICAL CHEMISTRY, 1985, 57 (08) :1506-1511
[4]   FLOW INJECTION ANALYSIS OF ENVIRONMENTAL SAMPLES FOR NITRATE USING AN ION-SELECTIVE ELECTRODE [J].
HANSEN, EH ;
GHOSE, AK ;
RUZICKA, J .
ANALYST, 1977, 102 (1219) :705-713
[5]  
HOKE SH, 1996, P HAL ALT TECHN WORK, P577
[6]  
JONES EA, 1989, S AFR J CHEM, V41, P25
[7]   FLOW-INJECTION APPROACH FOR THE DETERMINATION OF THE DYNAMIC-RESPONSE CHARACTERISTICS OF ION-SELECTIVE ELECTRODES .1. THEORETICAL CONSIDERATIONS [J].
KOLEV, SD ;
TOTH, K ;
LINDNER, E ;
PUNGOR, E .
ANALYTICA CHIMICA ACTA, 1990, 234 (01) :49-56
[8]  
LINDER E, 1988, DYNAMIC CHARACTERIST
[9]   RESPONSE-TIME STUDIES ON NEUTRAL CARRIER ION-SELECTIVE MEMBRANE ELECTRODES [J].
LINDNER, E ;
TOTH, K ;
PUNGOR, E ;
MORF, WE ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1978, 50 (12) :1627-1631
[10]   SWITCHED WALL JET FOR DYNAMIC-RESPONSE MEASUREMENTS [J].
LINDNER, E ;
TOTH, K ;
PUNGOR, E ;
BERUBE, TR ;
BUCK, RP .
ANALYTICAL CHEMISTRY, 1987, 59 (17) :2213-2216