Distribution of Absorbed Dose in Cone-Beam Breast Computed Tomography: A Phantom Study With Radiochromic Films

被引:8
|
作者
Russo, Paolo [1 ,2 ]
Coppola, Teresa [1 ,2 ]
Mettivier, Giovanni [1 ,2 ]
机构
[1] Univ Naples Federico II, Dipartimento Sci Fis, Lab Med Phys, I-80126 Naples, Italy
[2] Ist Nazl Fis Nucl, Sez Napoli, I-80126 Naples, Italy
关键词
Computed tomography imaging; dosimetry; CT; DOSIMETRY; GEOMETRY; MAMMOTOMOGRAPHY; COEFFICIENTS; FEASIBILITY;
D O I
10.1109/TNS.2010.2050005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Cone-Beam Breast Computed Tomography (CBBCT) of the pendant breast with dedicated scanners is an experimental 3D X-ray imaging technique for breast cancer diagnosis under evaluation in comparison to conventional two-view 2-D mammography of the compressed breast. In CBBCT it is generally assumed that a more uniform distribution of the radiation dose to the breast volume can be obtained, with respect to mammography, at equal Mean Glandular Dose (MGD) levels. In fact, in CBBCT the X-ray beam rotates for 360 deg around the breast, while in each mammography view the breast is irradiated from one side only. Using a CBBCT laboratory scanner developed by our group, we have measured the distribution of the radiation dose in a hemi-ellipsoidal PMMA breast phantom of 14 cm diameter simulating the average uncompressed breast, using radiochromic films type XR-SP inserted at mid-plane in the phantom. The technique factors were 80 kVp (5.6 mm Al Half Value Layer), tube load in the range 23-100 mAs, for an air kerma at isocenter in the range 4.7-20 mGy, for a calculated MGD in the range 3.5-15 mGy for a 14 cm diameter breast of 50% glandularity. Results indicate that the dose decreases from the periphery to the center of the phantom, and that along a transverse profile, the relative dose variation Delta = ((edge-center)/center) is up to (25 +/- 4) % at a distance of 80 mm from the nipple. As for the relative dose variation along the phantom longitudinal axis, the maximum value at middle of the phantom measured is delta = ((nipple-chest wall)/chest wall) = -(15 +/- 4) %, indicating that the dose decreases from the chest wall toward the nipple. The values of the parameters Delta and delta depend also on the height of the X-ray tube focal spot with respect to the phantom vertex (nipple). Results are in rough agreement with similar previous determinations using thermoluminescence dosimeters.
引用
收藏
页码:2220 / 2229
页数:10
相关论文
共 50 条
  • [1] Distribution of Absorbed Dose in Cone-Beam Breast Computed Tomography: a Phantom Study with Radiochromic Films
    Russo, Paolo
    Coppola, Teresa
    Mettivier, Giovanni
    Montesi, Maria Cristina
    Lauria, Adele
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3463 - 3470
  • [2] Dose Distribution in Cone-Beam Breast Computed Tomography: An Experimental Phantom Study
    Russo, Paolo
    Lauria, Adele
    Mettivier, Giovanni
    Montesi, Maria Cristina
    Villani, Natalia
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 366 - 374
  • [3] Dose Distribution in Cone-Beam Breast Computed Tomography: an experimental Phantom Study
    Russo, P.
    Lauria, A.
    Mettivier, G.
    Montesi, M. C.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 5382 - 5387
  • [4] EVALUATION OF DOSE HOMOGENEITY IN CONE-BEAM BREAST COMPUTED TOMOGRAPHY
    Mettivier, Giovanni
    Costa, Matilde
    Lanconelli, Nico
    Ianiro, Anna
    Pugliese, Mariagabriella
    Quarto, Maria
    Russo, Paolo
    RADIATION PROTECTION DOSIMETRY, 2017, 175 (04) : 473 - 481
  • [5] X-ray Cone-Beam Breast Computed Tomography: phantom studies
    Russo, P.
    Lauria, A.
    Mettivier, G.
    Montesi, M. C.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4076 - 4083
  • [6] X-ray Cone-Beam Breast Computed Tomography: Phantom Studies
    Russo, Paolo
    Mettivier, Giovanni
    Lauria, Adele
    Montesi, Maria Cristina
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 160 - 172
  • [7] Evaluation of Scattering in Cone-Beam Breast Computed Tomography: a Monte Carlo and experimental phantom study
    Mettivier, Giovanni
    Russo, Paolo
    Lanconelli, Nico
    Lo Meo, Sergio
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2930 - +
  • [8] Evaluation of Scattering in Cone-Beam Breast Computed Tomography: A Monte Carlo and Experimental Phantom Study
    Mettivier, Giovanni
    Russo, Paolo
    Lanconelli, Nico
    Lo Meo, Sergio
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (05) : 2510 - 2517
  • [9] A Phantom Study on Target Localization Accuracy Using Cone-Beam Computed Tomography
    Yan, Hui
    Zhang, Liwei
    Yin, Fang-Fang
    CLINICAL MEDICINE INSIGHTS-ONCOLOGY, 2008, 2 : 501 - 510
  • [10] Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography
    Trivedi, Gaurav
    Dixit, Chandra K.
    Oinam, Arun S.
    Kapoor, Rakesh
    Bahl, Amit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (03) : 470 - 474