摘要
The purpose of this study was to seek broad verification and validation of human lumbar spine finite element models created using a previously published automated algorithm. The automated algorithm takes segmented CT scans of lumbar vertebrae, automatically identifies important landmarks and contact surfaces, and creates a finite element model. Mesh convergence was evaluated by examining changes in key output variables in response to mesh density. Semi-direct validation was performed by comparing experimental results for a single specimen to the automated finite element model results for that specimen with calibrated material properties from a prior study. Indirect validation was based on a comparison of results from automated finite element models of 18 individual specimens, all using one set of generalized material properties, to a range of data from the literature. A total of 216 simulations were run and compared to 186 experimental data ranges in all six primary bending modes up to 7.8 Nm with follower loads up to 1000 N. Mesh convergence results showed less than a 5% difference in key variables when the original mesh density was doubled. The semi-direct validation results showed that the automated method produced results comparable to manual finite element modeling methods. The indirect validation results showed a wide range of outcomes due to variations in the geometry alone. The studies showed that the automated models can be used to reliably evaluate lumbar spine biomechanics, specifically within our intended context of use: in pure bending modes, under relatively low non-injurious simulated in vivo loads, to predict torque rotation response, disc pressures, and facet forces. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2669 / 2676
页数:8
相关论文
共 38 条
  • [1] Anderson Andrew E., 2007, Computer Methods in Biomechanics and Biomedical Engineering, V10, P171, DOI 10.1080/10255840601160484
  • [2] Parametric convergence sensitivity and validation of a finite element model of the human lumbar spine
    Ayturk, Ugur M.
    Puttlitz, Christian M.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 (08) : 695 - 705
  • [3] Incorporating Population-Level Variability in Orthopedic Biomechanical Analysis: A Review
    Bischoff, Jeffrey E.
    Dai, Yifei
    Goodlett, Casey
    Davis, Brad
    Bandi, Marc
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (02):
  • [4] Brinckmann P., 1991, SPINE
  • [5] An Automated Method for Landmark Identification and Finite-Element Modeling of the Lumbar Spine
    Campbell, Julius Quinn
    Petrella, Anthony J.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (11) : 2709 - 2716
  • [6] Cegofiino J., 2014, J APPL MATH
  • [7] Influence of an auxiliary facet system on intervertebral discs and adjacent facet joints
    Charles, Yann Philippe
    Lima, Lucas Venancio Pires Carvalho
    Persohn, Sylvain
    Rouch, Philippe
    Steib, Jean-Paul
    Skalli, Wafa
    [J]. SPINE JOURNAL, 2013, 13 (10) : 1293 - 1300
  • [8] STEPWISE VALIDATED FINITE ELEMENT MODEL OF THE HUMAN LUMBAR SPINE
    Coombs, Dana
    Bushelow, Michael
    Laz, Peter
    Rao, Milind
    Rullkoetter, Paul
    [J]. ASME 2013 CONFERENCE ON FRONTIERS IN MEDICAL DEVICES: APPLICATIONS OF COMPUTER MODELING AND SIMULATION, 2013,
  • [9] Comparison of eight published static finite element models of the intact lumbar spine: Predictive power of models improves when combined together
    Dreischarf, M.
    Zander, T.
    Shirazi-Adl, A.
    Puttlitz, C. M.
    Adam, C. J.
    Chen, C. S.
    Goel, V. K.
    Kiapour, A.
    Kim, Y. H.
    Labus, K. M.
    Little, J. P.
    Park, W. M.
    Wang, Y. H.
    Wilke, H. J.
    Rohlmann, A.
    Schmidt, H.
    [J]. JOURNAL OF BIOMECHANICS, 2014, 47 (08) : 1757 - 1766
  • [10] Multi-segment FEA of the human lumbar spine including the heterogeneity of the annulus fibrosus
    Eberlein, R
    Holzapfel, GA
    Fröhlich, M
    [J]. COMPUTATIONAL MECHANICS, 2004, 34 (02) : 147 - 163