Functional haemodynamic monitoring: The value of SVV as measured by the LiDCORapid™ in predicting fluid responsiveness in high risk vascular surgical patients

被引:27
作者
Willars, C. [1 ]
Dada, A. [2 ]
Hughes, T. [1 ]
Green, D. [2 ]
机构
[1] Kings Coll Hosp London, Inst Liver Studies, Liver Intens Care Unit, London SE5 9RS, England
[2] Kings Coll Hosp London, Dept Anaesthesia, London SE5 9RS, England
关键词
Goal-directed therapy; Functional haemodynamic monitoring; Fluid challenge; High-risk surgery; Patient outcomes; STROKE VOLUME VARIATION; MECHANICALLY VENTILATED PATIENTS; GUT MUCOSAL HYPOPERFUSION; PULSE PRESSURE VARIATION; HOSPITAL STAY; ESOPHAGEAL DOPPLER; CARDIAC-OUTPUT; OPTIMIZATION; SURGERY; COMPLICATIONS;
D O I
10.1016/j.ijsu.2012.02.003
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: There is growing evidence that optimal peri-operative fluid management in high-risk surgical patients improves their post-operative outcome. Functional haemodynamic parameters such as SVV (stroke volume variation), PPV (pulse pressure variation) and SPV (systolic pressure variation) have been shown to be superior to CVP (central venous pressure) and Delta CVP in predicting fluid responsiveness. The aim of this study was to determine the accuracy and threshold values of these dynamic parameters using the minimally invasive LiDCORapid (TM) in high-risk surgical patients. Methods: Fluid therapy in 70 patients undergoing vascular surgery was based on parameters derived from the LiDCORapid (TM). All patients received TIVA (total intravenous anaesthesia) and were ventilated with >= 7 ml/kg tidal volume. Haemodynamic data before and after fluid boluses were recorded and analyzed retrospectively. Delta SVI (stroke volume index) >= 10% was determined to be a positive response to a fluid challenge. The AUROCs (area under the receiver-operator curves) and confidence intervals thereof were used to assess the ability of each parameter to predict fluid responsiveness. Results: 32/43 fluid challenges were positive (74.4%). The correlation coefficients between baseline SVV, PPV, and SPV with Delta SVI were: 0.27 (p = 0.08), 0.18 (p = 0.25) and -0.01 (p = 0.96) whilst the AUROCs were 0.75 (0.57-0.93), 0.67 (0.48-0.85) and 0.58 (0.35-0.81) respectively. The best cut-off for SVV using Youden's index was 13.5%, with J = 0.48. At this level, the LR+ (positive likelihood ratio) = 2.74, LR-(negative likelihood ratio) = 0.34 and the DOR (diagnostic odds ratio) = 8.06. Conclusion: Only the SVV was an adequate predictor of fluid responsiveness in this cohort of high risk surgical patients. Whereas PPV and SPV may be obtained from the arterial trace, estimation of the SVV requires a cardiac output monitor which is able to convert an arterial pressure trace into an estimation of stroke volume. (c) 2012 Surgical Associates Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:148 / 152
页数:5
相关论文
共 44 条
[1]  
Avery SA, 2010, PULSECO DERIVED STRO
[2]   Assessment of fluid-responsiveness parameters for off-pump coronary artery bypass surgery: A comparison among LiDCO, transesophageal echochardiography, and pulmonary artery catheter [J].
Belloni, Luigi ;
Pisano, Antonio ;
Natale, Armando ;
Piccirillo, Maria Rosario ;
Piazza, Luigi ;
Ismeno, Gennaro ;
De Martino, Giovanni .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2008, 22 (02) :243-248
[3]   Intraoperative fluid optimization using stroke volume variation in high risk surgical patients: results of prospective randomized study [J].
Benes, Jan ;
Chytra, Ivan ;
Altmann, Pavel ;
Hluchy, Marek ;
Kasal, Eduard ;
Svitak, Roman ;
Pradl, Richard ;
Stepan, Martin .
CRITICAL CARE, 2010, 14 (03)
[4]  
Berkenstadt H, 2001, ANESTH ANALG, V92, P984
[5]   PREOPERATIVE OPTIMIZATION OF CARDIOVASCULAR HEMODYNAMICS IMPROVES OUTCOME IN PERIPHERAL VASCULAR-SURGERY - A PROSPECTIVE, RANDOMIZED CLINICAL-TRIAL [J].
BERLAUK, JF ;
ABRAMS, JH ;
GILMOUR, IJ ;
OCONNOR, SR ;
KNIGHTON, DR ;
CERRA, FB .
ANNALS OF SURGERY, 1991, 214 (03) :289-299
[6]   Uncalibrated pulse contour-derived stroke volume variation predicts fluid responsiveness in mechanically ventilated patients undergoing liver transplantation [J].
Biais, M. ;
Nouette-Gaulain, K. ;
Cottenceau, V. ;
Revel, P. ;
Sztark, F. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 101 (06) :761-768
[7]   Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery [J].
Broch, Ole ;
Renner, Jochen ;
Hoecker, Jan ;
Gruenewald, Matthias ;
Meybohm, Patrick ;
Schoettler, Jan ;
Steinfath, Markus ;
Bein, Berthold .
CRITICAL CARE, 2011, 15 (01)
[8]   EFFECT OF LUNG-INFLATION ON LUNG BLOOD-VOLUME AND PULMONARY VENOUS FLOW [J].
BROWER, R ;
WISE, RA ;
HASSAPOYANNES, C ;
BROMBERGERBARNEA, B ;
PERMUTT, S .
JOURNAL OF APPLIED PHYSIOLOGY, 1985, 58 (03) :954-963
[9]   EFFECT OF INTRA-THORACIC PRESSURE ON LEFT-VENTRICULAR PERFORMANCE [J].
BUDA, AJ ;
PINSKY, MR ;
INGELS, NB ;
DAUGHTERS, GT ;
STINSON, EB ;
ALDERMAN, EL .
NEW ENGLAND JOURNAL OF MEDICINE, 1979, 301 (09) :453-459
[10]   Monitoring of peri-operative fluid administration by individualized goal-directed therapy [J].
Bundgaard-Nielsen, M. ;
Holte, K. ;
Secher, N. H. ;
Kehlet, H. .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2007, 51 (03) :331-340