Blood glucose regulation and control of insulin and glucagon infusion using single model predictive control for type 1 diabetes mellitus

被引:8
作者
Dias, Cifha Crecil [1 ]
Kamath, Surekha [1 ]
Vidyasagar, Sudha [2 ]
机构
[1] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Instrumentat & Control, Manipal, India
[2] Manipal Acad Higher Educ, Kasturba Med Coll, Dept Med, Manipal, India
关键词
predictive control; blood; medical control systems; sugar; diseases; artificial organs; patient treatment; physiological models; biochemistry; blood glucose; longer control time; insulin; single model predictive controller; glucagon infusion; single model predictive control; type 1 diabetes mellitus; model predictive control algorithm; comprehensive physiological model; Sorensen model; ARTIFICIAL PANCREAS; GLYCEMIC CONTROL; PUMP THERAPY; COMPLICATIONS; ALGORITHMS; CHALLENGES;
D O I
10.1049/iet-syb.2019.0101
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
This study elaborates on the design of artificial pancreas using model predictive control algorithm for a comprehensive physiological model such as the Sorensen model, which regulates the blood glucose and can have a longer control time in normal glycaemic region. The main objective of the proposed algorithm is to eliminate the risk of hyper and hypoglycaemia and have a precise infusion of hormones: insulin and glucagon. A single model predictive controller is developed to control the bihormones, insulin, and glucagon for such a development unmeasured disturbance is considered for a random time. The simulation result for the proposed algorithm performed good regulation lowering the hypoglycaemia risk and maintaining the glucose level within the normal glycaemic range. To validate the performance of the tracking of output and setpoint, average tracking error is used and 4.4 mg/dl results are obtained while compared with standard value (14.3 mg/dl).
引用
收藏
页码:133 / 146
页数:14
相关论文
共 50 条
[31]   Influence of GLP-1 receptor agonist on insulin dosage and blood glucose control of patients with type 2 diabetes mellitus [J].
Shen, Yimei ;
Yang, Xiaohua ;
Han, Xiaojun ;
Xi, Wei ;
Jiang, Lihua ;
Wang, Shuqin ;
Zhong, Haifeng ;
Gu, Yunjuan .
AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2021, 13 (10) :11814-11823
[32]   Run-to-run control of blood glucose concentrations for people with type 1 diabetes mellitus [J].
Owens, Camelia ;
Zisser, Howard ;
Jovanovic, Lois ;
Srinivasan, Bala ;
Bonvin, Dorninique ;
Doyle, Francis J., III .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (06) :996-1005
[33]   Habituating Multiple Model Predictive Control for Blood Glucose Regulation in the ICU [J].
Sun, Jing ;
Cameron, Fraser ;
Bequette, B. Wayne .
2012 AMERICAN CONTROL CONFERENCE (ACC), 2012, :3509-3515
[34]   Poor Control of Blood Glucose, Lifestyle, and Cardiometabolic Parameters in Younger Adult Patients with Type 2 Diabetes Mellitus [J].
Nam, Ga Eun ;
Han, Byoungduck ;
Joo, Chae Lin ;
Kang, Seo Young ;
Lim, Jisun ;
Kim, Yang-Hyun ;
Park, Hye Soon .
JOURNAL OF CLINICAL MEDICINE, 2019, 8 (09)
[35]   Model Predictive Control of Blood Glucose for Type 1 Diabetic Rats in a Cyber-Physical System [J].
Ko, Hoo Sang ;
Balouchzadeh, Ramin ;
Uzun, Guney ;
Lee, H. Felix ;
Park, Sarah ;
Kwon, Guim .
25TH INTERNATIONAL CONFERENCE ON PRODUCTION RESEARCH MANUFACTURING INNOVATION: CYBER PHYSICAL MANUFACTURING, 2019, 39 :341-348
[36]   Closed-Loop Control of Artificial Pancreatic β-Cell in Type 1 Diabetes Mellitus Using Model Predictive Iterative Learning Control [J].
Wang, Youqing ;
Dassau, Eyal ;
Doyle, Francis J., III .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (02) :211-219
[37]   Multicenter outpatient dinner/overnight reduction of hypoglycemia and increased time of glucose in target with a wearable artificial pancreas using modular model predictive control in adults with type 1 diabetes [J].
Del Favero, S. ;
Place, J. ;
Kropff, J. ;
Messori, M. ;
Keith-Hynes, P. ;
Visentin, R. ;
Monaro, M. ;
Galasso, S. ;
Boscari, F. ;
Toffanin, C. ;
Di Palma, F. ;
Lanzola, G. ;
Scarpellini, S. ;
Farret, A. ;
Kovatchev, B. ;
Avogaro, A. ;
Bruttomesso, D. ;
Magni, L. ;
DeVries, J. H. ;
Cobelli, C. ;
Renard, E. .
DIABETES OBESITY & METABOLISM, 2015, 17 (05) :468-476
[38]   Blood Glucose Control for Type I Diabetes Mellitus Considering Time Delay [J].
Khalesi, Zahra ;
Dehghani, Maryam .
2017 25TH IRANIAN CONFERENCE ON ELECTRICAL ENGINEERING (ICEE), 2017, :639-644
[39]   Internal model sliding mode control approach for glucose regulation in type 1 diabetes [J].
Abu-Rmileh, Amjad ;
Garcia-Gabin, Winston ;
Zambrano, Darine .
BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2010, 5 (02) :94-102
[40]   Predictors of glucose control in children and adolescents with type 1 diabetes mellitus [J].
Urbach, SL ;
LaFranchi, S ;
Lambert, L ;
Lapidus, JA ;
Daneman, D ;
Becker, TM .