Improved quantification for Tc-94m PET imaging

被引:14
作者
Barker, WC [1 ]
Szajek, LP [1 ]
Green, SL [1 ]
Carson, RE [1 ]
机构
[1] NIH, Warren Grant Magnuson Clin Ctr, PET Dept, Bethesda, MD 20892 USA
关键词
biomedical nuclear imaging; Compton scattering; poistron emission tomography (PET);
D O I
10.1109/23.940156
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The positron emitter Tc-94m (52.5-min half-life) has the potential to improve quantification of tracers currently labeled with Tc-99m. The successful production of Tc-94m and the development of an accurate multiisotope decay correction have improved quantification for Tc-94m positron emission tomography (PET) imaging. However, the presence of gamma rays emitted in coincidence with the positrons produces an apparent increase in scatter of 40% to 90%. Furthermore, the contaminant positron emitter Tc-94 (293-min half-life) produced with Tc-94m also has positron-coincident gammas, so the effect is time-dependent. In this study, modification of the kernel parameters of the Bergstrom two-dimensional (2-D) scatter correction is proposed as a simple way to improve technetium PET quantification. Parameter optimization was performed using line source scatter phantom data for F-18 and technetium for a variety of positions in cylindrical and elliptical phantoms over a 3-h period. The accuracy of the correction was evaluated with remnant scatter fraction measurements using uniformly filled phantoms with 5-cm cold cylindrical inserts. On average, remnant 2-D scatter fractions of 10 +/- 3% were observed when parameters optimized for F-18 were used. The use of parameters optimized for technetium reduced the remnant scatter fractions to 1 +/- 3%, showing that accurate quantification with Tc-94m can be achieved.
引用
收藏
页码:739 / 742
页数:4
相关论文
共 12 条
  • [1] CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS
    BERGSTROM, M
    ERIKSSON, L
    BOHM, C
    BLOMQVIST, G
    LITTON, J
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) : 42 - 50
  • [2] IMPROVING THE RADIONUCLIDIC PURITY OF TC-94M FOR PET IMAGING
    CHRISTIAN, BT
    NICKLES, RJ
    STONE, CK
    MULNIX, TL
    CLARK, J
    [J]. APPLIED RADIATION AND ISOTOPES, 1995, 46 (02) : 69 - 73
  • [3] DEGRADO TR, 1994, J NUCL MED, V35, P1398
  • [4] SCATTER CORRECTION IN THE TRANSAXIAL SLICES OF A WHOLE-BODY POSITRON EMISSION TOMOGRAPH
    HOVERATH, H
    KUEBLER, WK
    OSTERTAG, HJ
    DOLL, J
    ZIEGLER, SI
    KNOPP, MV
    LORENZ, WJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) : 717 - 728
  • [5] KOHLMYER SG, 1999, J NUCL MED, V40
  • [6] Lederer CM., 1978, Table of isotopes
  • [7] *NAT EL MAN ASSN, 1994, PERF MEAS POS EM TOM
  • [8] PENTLOW KS, 1999, J NUCL MED, V40
  • [9] Production of high purity 94mTc for positron emission tomography studies
    Qaim, SM
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 2000, 27 (04) : 323 - 328
  • [10] ROSCH F, 1994, J NUCL MED, V35, pP253