Wide range force feedback for catheter insertion mechanism for use in minimally invasive mitral valve repair surgery

被引:0
作者
Ahmadi, Roozbeh [1 ]
Sokhanvar, Saeed [2 ]
Packirisamy, Muthukumaran [1 ]
Dargahi, Javad [1 ]
机构
[1] Concordia Univ, Opt Bio Microsyst Lab, Dept Mech & Ind Engn, 1515 St Catherine W, Montreal, PQ H3G 2W1, Canada
[2] MIT, BioInstrumentat Lab, Cambridge, MA 02139 USA
来源
OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS AND TREATMENT APPLICATIONS IX | 2009年 / 7173卷
关键词
Optical MEMS; Sensor; Finite Element Analysis; Minimally Invasive Surgery; Mitral Valve Annuloplasty Surgery;
D O I
10.1117/12.808308
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Mitral valve regurgitation (MR) is a condition in which heart's mitral valve does not close tightly, which allows blood to leak back into the left atrium. Restoring the dimension of the mitral-valve annulus by percutaneous intervention surgery is a common choice to treat MR. Currently, this kind of open heart annuloplasty surgery is being performed through sternotomy with cardiomyopathy bypass. In order to reduce trauma to the patient and also to eliminate bypass surgery, robotic assisted minimally invasive surgery (MIS) procedure, which requires small keyhole incisions, has a great potential. To perform this surgery through MIS procedure, an accurate computer controlled catheter with wide-range force feedback capabilities is required. There are three types of tissues at the site of operation: mitral leaflet, mitral annulus and left atrium. The maximum allowable applied force to these three types of tissue is totally different. For instance, leaflet tissue is the most sensitive one with the lowest allowable force capacity. For this application, therefore, a wide-range force sensing is highly required. Most of the sensors that have been developed for use in MIS applications have a limited range of sensing. Therefore, they need to be calibrated for different types of tissue. The present work, reports on the design, modeling and simulation of a novel wide-range optical force sensor for measurement of contact pressure between catheter tip and heart tissue. The proposed sensor offers a wide input range with a high resolution and sensitivity over this range. Using Micro-Electro-Mechanical-Systems (MEMS) technology, this sensor can be microfabricated and integrated with commercially available catheters.
引用
收藏
页数:8
相关论文
共 7 条
[1]  
Adams M., INTRO OPTICAL WAVEGU
[2]   Percutaneous Mitral Valve Repair [J].
Gillinov, A. Marc ;
Liddicoat, John R. .
SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2006, 18 (02) :115-121
[3]   SU-8 MEMS Fabry-Perot pressure sensor [J].
Hill, G. C. ;
Melamud, R. ;
Declercq, F. E. ;
Davenport, A. A. ;
Chan, I. H. ;
Hartwell, P. G. ;
Pruitt, B. L. .
SENSORS AND ACTUATORS A-PHYSICAL, 2007, 138 (01) :52-62
[4]  
Madou M. J., 2002, FUNDAMENTALS MICROFA
[5]  
MATHUR V, 2006, COMSOL US C BOST
[6]   Novel robotic catheter remote control system: Feasibility and safety of transseptal puncture and endocardial catheter navigation [J].
Saliba, Walid ;
Cummings, Jennifer E. ;
Oh, Seil ;
Zhang, Youhua ;
Mazgalev, Todor N. ;
Schweikert, Robert A. ;
Burkhardt, J. David ;
Natale, Andrea .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (10) :1102-1105
[7]   A multifunctional PVDF-based tactile sensor for minimally invasive surgery [J].
Sokhanvar, S. ;
Packirisamy, M. ;
Dargahi, J. .
SMART MATERIALS AND STRUCTURES, 2007, 16 (04) :989-998