Correction for the 17O interference in δ(13C) measurements when analyzing CO2 with stable isotope mass spectrometry (IUPAC Technical Report)

被引:293
作者
Brand, Willi A. [1 ]
Assonov, Sergey S. [2 ]
Coplen, Tyler B. [3 ]
机构
[1] Max Planck Inst Biogeochem, MPI BGC, D-07701 Jena, Germany
[2] European Commiss, Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
[3] US Geol Survey, Reston, VA 20192 USA
关键词
delta(C-13) correction; delta(C-18) correction; IUPAC Inorganic Chemistry Division; O-17; correction; oxygen isotope ratio; OXYGEN; FRACTIONATION; DELTA-O-17; O-18/O-16; ABUNDANCE; O-17/O-16; VALUES; CARBON; RATIOS;
D O I
10.1351/PAC-REP-09-01-05
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Measurements of delta(C-13) determined on CO2 with an isotope-ratio mass spectrometer (IRMS) must be corrected for the amount of O-17 in the CO2. For data consistency, this must be done using identical methods by different laboratories. This report aims at unifying data treatment for CO2 IRMS by proposing (i) a unified set of numerical values, and (ii) unified correction algorithm, based on a simple, linear approximation formula. Because the oxygen of natural CO2 is derived mostly from the global water pool, it is recommended that a value of 0.528 be employed for the factor lambda, which relates differences in O-17 and O-18 abundances. With the currently accepted N(C-13)/N(C-12) of 0.011 180(28) in VPDB (Vienna Peedee belemnite) reevaluation of data yields a value of 0.000 393(1) for the oxygen isotope ratio N(O-17)/N(O-16) of the evolved CO2. The ratio of these quantities, a ratio of isotope ratios, is essential for the O-17 abundance correction: [N(O-17)/N(O-16)/[N(C-13)/N(C-12)] = 0.035 16(8). The equation [delta(C-13) approximate to (45)delta(VPDB-CO2) + 2 R-17/R-13 ((45)delta(VPDB-Ca2) - lambda(46)delta(VPDB-CO2))]closely approximates delta(C-13) values with less than 0.010 % deviation for normal oxygenbearing materials and no more than 0.026 `700 in extreme cases. Other materials containing oxygen of non-mass-dependent isotope composition require a more specific data treatment. A similar linear approximation is also suggested for delta(O-18). The linear approximations are easy to implement in a data spreadsheet, and also help in generating a simplified uncertainty budget.
引用
收藏
页码:1719 / 1733
页数:15
相关论文
共 30 条
[1]  
ALLISON CE, 1995, RECOMMENDATIONS REPO
[2]   Reporting small Δ17O values:: existing definitions and concepts [J].
Assonov, SS ;
Brenninkmeijer, CAM .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (05) :627-636
[3]   On the 17O correction for CO2 mass spectrometric isotopic analysis [J].
Assonov, SS ;
Brenninkmeijer, CAM .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (10) :1007-1016
[4]   A redetermination of absolute values for 17RVPDB-CO2 and 17RVSMOW [J].
Assonov, SS ;
Brenninkmeijer, CAM .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (10) :1017-1029
[5]   A new method to determine the 17O isotopic abundance in CO2 using oxygen isotope exchange with a solid oxide [J].
Assonov, SS ;
Brenninkmeijer, CAM .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (24) :2426-2437
[6]   ABSOLUTE O-18 CONTENT OF STANDARD MEAN OCEAN WATER [J].
BAERTSCHI, P .
EARTH AND PLANETARY SCIENCE LETTERS, 1976, 31 (03) :341-344
[7]   High precision measurements of 17O/16O and 18O/16O ratios in H2O [J].
Barkan, E ;
Luz, B .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (24) :3737-3742
[8]   How well do we know VPDB? Variability of δ13C and δ18O in CO2 generated from NBS19-calcite [J].
Brand, Will A. ;
Huang, Lin ;
Mukai, Hitoshi ;
Chivulescu, Alina ;
Richter, Juergen M. ;
Rothe, Michael .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (06) :915-926
[9]   COMPARISON OF STABLE ISOTOPE REFERENCE SAMPLES [J].
COPLEN, TB ;
KENDALL, C ;
HOPPLE, J .