Correcting capacitive displacement measurements in metrology applications with cylindrical artifacts

被引:38
作者
Smith, PT
Vallance, RR
Marsh, ER
机构
[1] George Washington Univ, Precis Syst Lab, Washington, DC 20052 USA
[2] Univ Kentucky, Lexington, KY 40506 USA
[3] Penn State Univ, Machine Dynam Res Lab, University Pk, PA 16802 USA
来源
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY | 2005年 / 29卷 / 03期
关键词
displacement sensors; cylinder; artifact; metrology;
D O I
10.1016/j.precisioneng.2004.11.010
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Metrology applications commonly require non-contact, capacitive sensors for displacement measurements due to their nanometer resolution. In some metrology applications, for example, the measurement of roundness and spindle error motion, the displacements of stationary and rotating cylindrical artifacts are measured. Error from using a conventionally calibrated sensor with a non-flat (e.g., cylindrical) target is typically neglected, but these errors cannot be ignored for nanometer-level accuracy. The capacitance between a sensor and a cylindrical target is less than that of a sensor with a flat target, which causes four effects. As the diameter of the target shrinks, the sensitivity of the sensor increases, the sensing range decreases, the sensing range shifts towards the target, and the nonlinearity increases. These errors can be greatly reduced by either calibrating sensors with the correct target surface or by determining corrections for post-processing data. This paper quantifies and experimentally verifies these errors for a commonly used sensor, and a simulation of a nanometer-level measurement of out-of-roundness and spindle error motion demonstrates that measurement accuracy is improved with corrected sensitivities. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:324 / 335
页数:12
相关论文
共 48 条
[1]  
[Anonymous], 1907, B BUREAU STANDARDS, DOI DOI 10.6028/BULLETIN.070
[2]   Global model generation for a capacitive silicon accelerometer by finite-element analysis [J].
Ansel, Y ;
Romanowicz, B ;
Renaud, P ;
Schropfer, G .
SENSORS AND ACTUATORS A-PHYSICAL, 1998, 67 (1-3) :153-158
[3]   A high precision method for measuring very small capacitance changes [J].
Ashrafi, A ;
Golnabi, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (08) :3483-3487
[4]  
*ASME, 2000, B557 ANSIASME
[5]  
*ASME, 1992, B554 ANSIASME
[6]  
*ASME, 1985, B8934M ANSIASME
[7]  
*ASME ANSI, 1972, B8931 ASMEANSI
[8]  
ATHERTON PD, 1997, NANOPOSITIONING BOOK
[9]   FINITE-ELEMENT MODELING AS A TOOL FOR DESIGNING CAPACITIVE POSITION SENSORS [J].
BONSE, MHW ;
MUL, C ;
SPRONCK, JW .
SENSORS AND ACTUATORS A-PHYSICAL, 1995, 46 (1-3) :266-269
[10]   EFFECT OF TILT AND SURFACE DAMAGE ON PRACTICAL CAPACITANCE DISPLACEMENT TRANSDUCERS [J].
BROWN, MA ;
BULLEID, CE .