Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE

被引:32
作者
Schroder, David T. [2 ]
Kelly, Natalie H. [1 ]
Wright, Timothy M. [1 ]
Parks, Michael L. [2 ]
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Adult Reconstruct & Joint Replacement Div, New York, NY 10021 USA
关键词
TOTAL HIP; POLYETHYLENE;
D O I
10.1007/s11999-010-1552-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Highly crosslinked UHMWPE is associated with increased wear resistance in hip simulator and clinical studies. Laboratory and case studies, however, have described rim fracture in crosslinked acetabular liners. Controversy exists, therefore, on the relative merits of crosslinked liners over conventional liners in terms of wear performance versus resistance to fatigue cracking. Questions/purposes We asked whether crosslinked liners would show less surface damage than conventional liners but would be more susceptible to fatigue damage. Methods We examined 36 conventional UHMWPE and 39 crosslinked UHMWPE retrieved implants with similar patient demographics and identical design for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks. Results We observed no difference in wear damage scores for the two liners. Conventional liners more frequently impinged but were more often elevated with smaller head sizes. We observed creep in approximately 70% of both types of liners. Incipient rim cracks were found in five crosslinked liners, and one liner had a rim fracture. Only one conventional liner had an incipient rim crack. Conclusions Contrary to our expectation, damage was similar between crosslinked and conventional UHMWPE liners. Moreover, the 15% occurrence (six of 39) of incipient or complete fractures in crosslinked liners as compared with a 3% occurrence (one of 36) in conventional liners may have implications for the long-term performance of crosslinked liners. Longer-term studies will be necessary to establish the fate of rim cracks and thus the overall clinical fatigue performance of crosslinked liners.
引用
收藏
页码:387 / 394
页数:8
相关论文
共 29 条
[1]   Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners [J].
Bradford, L ;
Baker, DA ;
Graham, J ;
Chawan, A ;
Ries, MD ;
Pruitt, LA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1271-1282
[2]   Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up [J].
Bragdon, Charles R. ;
Barrett, Susan ;
Martell, John M. ;
Greene, Meridith E. ;
Malchau, Henrik ;
Harris, William H. .
JOURNAL OF ARTHROPLASTY, 2006, 21 (07) :935-943
[3]   A Double-Blind, Prospective, Randomized Controlled Trial Comparing Highly Cross-Linked and Conventional Polyethylene in Primary Total Hip Arthroplasty [J].
Calvert, George T. ;
Devane, Peter A. ;
Fielden, Jan ;
Adams, Kathryn ;
Horne, J. Geoffrey .
JOURNAL OF ARTHROPLASTY, 2009, 24 (04) :505-510
[4]   Gamma irradiation alters fatigue-crack behavior and fracture toughness in 1900H and GUR 1050 UHMWPE [J].
Cole, JC ;
Lemons, JE ;
Eberhardt, AW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05) :559-566
[5]   Comparison of cross-linked polyethylene materials for orthopaedic applications [J].
Collier, JP ;
Currier, BH ;
Kennedy, FE ;
Currier, JH ;
Timmins, GS ;
Jackson, SK ;
Brewer, RL .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (414) :289-304
[6]   The Otto Aufranc Award -: Highly cross-linked polyethylene in Total Hip Arthroplasty -: Randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :6-16
[7]   5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets -: Two randomized studies using radiostereometric analysis [J].
Digas, Georgios ;
Karrholm, Johan ;
Thanner, Jonas ;
Herberts, Peter .
ACTA ORTHOPAEDICA, 2007, 78 (06) :746-754
[8]   Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32-and 38-mm femoral heads [J].
Estok, Daniel M., II ;
Burroughs, Brian R. ;
Muratoglu, Orhun K. ;
Harris, William H. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (04) :581-589
[9]   Clinical fracture of cross-linked UHMWPE acetabular liners [J].
Furmanski, Jevan ;
Anderson, Martin ;
Bal, Sonny ;
Greenwald, A. Seth ;
Halley, David ;
Penenberg, Brad ;
Ries, Michael ;
Pruitt, Lisa .
BIOMATERIALS, 2009, 30 (29) :5572-5582
[10]   Fatigue crack propagation resistance of virgin and highly crosslinked, thermally treated ultra-high molecular weight polyethylene [J].
Gencur, SJ ;
Rimnac, CM ;
Kurtz, SM .
BIOMATERIALS, 2006, 27 (08) :1550-1557