In this study, we measured scintillating and Cerenkov lights generated in a scintillating fiber-optic sensor using high-energy electron beams and a spectrometer. The spectrum of Cerenkov light generated in a plastic optical fiber was measured with a spectrometer and the intensities of Cerenkov light were measured for irradiated lengths of the plastic optical fiber by integrating the photon counts at every wavelength of the spectrum. The intensities of Cerenkov and scintillating lights were also characterized as a function of the incident angle of an electron beam from a clinical linear accelerator (CLINAC). To minimize or remove Cerenkov light, a subtraction method was employed using a background optical fiber. (C) 2011 The Japan Society of Applied Physics
引用
收藏
页码:176 / 179
页数:4
相关论文
共 12 条
[1]   Radiation-induced light in optical fibers and plastic scintillators: Application to brachytherapy dosimetry [J].
Arnfield, MR ;
Gaballa, HE ;
Zwicker, RD ;
Islam, Q ;
SchmidtUllrich, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :2077-2084
[2]   CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :925-935
[3]   A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams [J].
Clift, MA ;
Johnston, PN ;
Webb, DV .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1421-1433
[4]   Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams [J].
Clift, MA ;
Sutton, RA ;
Webb, DV .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1165-1182
[5]   OPTICAL FILTERING AND SPECTRAL MEASUREMENTS OF RADIATION-INDUCED LIGHT IN PLASTIC SCINTILLATION DOSIMETRY [J].
DEBOER, SF ;
BEDDAR, AS ;
RAWLINSON, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07) :945-958
[6]  
Fontbonne JM, 2002, IEEE T NUCL SCI, V49, P2223, DOI 10.1109/TNS.2002.803680
[7]   Spectral discrimination of Cerenkov radiation in scintillating dosimeters [J].
Frelin, AM ;
Fontbonne, JM ;
Ban, G ;
Colin, J ;
Labalme, M ;
Batalla, A ;
Isambert, A ;
Vela, A ;
Leroux, T .
MEDICAL PHYSICS, 2005, 32 (09) :3000-3006
[8]  
Knoll G. F., 1999, RAD DETECTION MEASUR, P711
[9]   Measurements and elimination of Cherenkov light in fiber-optic scintillating detector for electron beam therapy dosimetry [J].
Lee, Bongsoo ;
Jang, Kyoung Won ;
Cho, Dong Hyun ;
Yoo, Wook Jae ;
Tack, Gye-Rac ;
Chung, Soon-Cheol ;
Kim, Sin ;
Choc, Hyosung .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01) :344-348
[10]   Fabrication and characterization of a miniature fiber-optic scintillating detector for electron-beam therapy Dosimetry [J].
Lee, Bongsoo ;
Jang, Kyoung Won ;
Cho, Dong Hyun ;
Chung, Soon-Cheol ;
Tack, Gye-Rae ;
Eom, Gwang-Moon ;
Yi, Jeong Han .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2007, 51 (05) :1645-1648