Two-dimensional GEM imaging detector with delay-line readout
被引:41
作者:
Guedes, GP
Breskin, A
Chechik, R
[1
]
Vartsky, D
Bar, D
Barbosa, AF
Marinho, PRB
机构:
[1] Weizmann Inst Sci, Dept Particle Phys, IL-76100 Rehovot, Israel
[2] Soreq NRC, Yavne, Israel
[3] CBPF, BR-15022918 Rio De Janeiro, RJ, Brazil
来源:
关键词:
GEM;
gaseous imaging detector;
delay-line readout;
2D readout;
D O I:
10.1016/S0168-9002(03)02017-5
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
A 100 x 100 mm(2) 2D imaging detector, based on a triple-GEM gaseous multiplier, striped x-y readout anode and discrete delay-line readout, is presented. The fast (2.1 ns tap(-1)) delay-line circuit was designed to match the anode-charge signal profile, namely its rise-time and length. The detector's imaging capability was systematically studied in Ar/CO2 (70/30) with 5.9 keV X-rays; x-y resolution of sigma = 0.05 and 0.1 mm for top and bottom anode strips, respectively, and integral non-linearity of similar to0.15% are demonstrated. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:473 / 483
页数:11