Complications after percutaneous internal fixator for anterior pelvic ring injuries

被引:65
作者
Fang, Christian [1 ]
Alabdulrahman, Hatem [2 ]
Pape, Hans-Christophe [3 ]
机构
[1] Univ Hong Kong, Dept Orthopaed & Traumatol, Queen Mary Hosp, 102 Pokfulam Rd, Hong Kong, Hong Kong, Peoples R China
[2] Rhein Westfal TH Aachen, Dept Orthopaed Trauma, 30 Pauwels St, D-52074 Aachen, Germany
[3] Univ Hosp Zurich, Dept Trauma, Raemistr 100, CH-8091 Zurich, Switzerland
关键词
Pelvis fracture; Internal fixation; Minimal invasive surgery; External fixator; Complications; Lateral femoral cutaneous nerve; EXTERNAL FIXATION; EXPERIENCE; REDUCTION; FRACTURES; SERIES; NERVE;
D O I
10.1007/s00264-017-3415-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives The aim of the study is to report on the observed incidence of complications following pedicle screw and rod internal fixator (INFIX) stabilization of pelvic ring fractures. Methods In a retrospective review of consecutive patient series conducted in a University level 1 trauma hospital, 43 patients (21 female and eight male), mean age 64.2 (range 16-87) with OTA/AO type B or C pelvic ring fractures received percutaneous pedicle screw and rod internal fixator (INFIX) anterior stabilization of pelvic ring fractures; 29 fulfilled inclusion criteria with three months' minimal follow-up or known complication. Mean follow-up was 7.2 months. Main outcome measure was the incidence of complications and adverse outcomes. Results Fourteen (48.3%) had injury to the lateral femoral cutaneous nerve (LFCN). Three (10.3%) had chronic pain of the pelvis not related to the LFCN. Five out of 29 (17%) of patients had unplanned removal before six weeks with two due to early loosening, one femoral nerve palsy, one deep infection and one painful implant impingement. There were no patients with deep vein thrombosis, no intra-abdominal violations or vascular complications; 22 (76%) returned to their premorbid walking status. Conclusion With high risk of LFCN injury, we caution against liberal use of the INFIX in patients with stable fractures where conservative treatment may be more appropriate. Most complications occurring from INFIX are self-limiting.
引用
收藏
页码:1785 / 1790
页数:6
相关论文
共 29 条
[1]   Anterior Subcutaneous Internal Fixation of the Pelvis: Placement of the INFIX [J].
Chaus, George W. ;
Weaver, Michael J. .
OPERATIVE TECHNIQUES IN ORTHOPAEDICS, 2015, 25 (04) :262-269
[2]   Minimally-invasive fixation for anterior pelvic ring disruptions [J].
Cole, Peter A. ;
Dyskin, Evgeny A. ;
Gilbertson, Jeffrey A. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 :S27-S34
[3]   Anterior Pelvic External Fixator Versus Subcutaneous Internal Fixator in the Treatment of Anterior Ring Pelvic Fractures [J].
Cole, Peter A. ;
Gauger, Erich M. ;
Anavian, Jack ;
Ly, Thuan V. ;
Morgan, Robert A. ;
Heddings, Archie A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (05) :269-277
[4]   Internal anterior fixators for pelvic ring injuries: Do monaxial pedicle screws provide more stiffness than polyaxial pedicle screws? [J].
Eagan, Michael ;
Kim, Hyunchul ;
Manson, Theodore T. ;
Gary, Joshua L. ;
Russell, Joseph P. ;
Hsieh, Adam H. ;
O'Toole, Robert V. ;
Boulton, Christina L. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (06) :996-1000
[5]   Epidemiology of pelvic ring injuries [J].
Gansslen, A ;
Pohlemann, T ;
Paul, C ;
Lobenhoffer, P ;
Tscherne, H .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1996, 27 :13-20
[6]   Anterior Pelvic Reduction and Fixation Using a Subcutaneous Internal Fixator [J].
Gardner, Michael J. ;
Mehta, Samir ;
Mirza, Amer ;
Ricci, William M. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (05) :314-321
[7]   Percutaneous fixation of the pelvic ring - An update [J].
Giannoudis, P. V. ;
Tzioupis, C. C. ;
Pape, H. -C. ;
Roberts, C. S. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (02) :145-154
[8]  
Hesse D, 2015, J ORTHOP TRAUMA, V29, P138, DOI 10.1097/BOT.0000000000000193
[9]   Surgical Technique: A Percutaneous Method of Subcutaneous Fixation for the Anterior Pelvic Ring: The Pelvic Bridge [J].
Hiesterman, Timothy G. ;
Hill, Brian W. ;
Cole, Peter A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (08) :2116-2123
[10]  
Kuttner M, 2009, UNFALLCHIRURG, V112, P661, DOI 10.1007/s00113-009-1623-0