Determination of hexavalent chromium in solid waste hypochlorite treated leachates by headspace gas chromatography
被引:3
|作者:
Zhong, Guang-Zhi
[1
]
Hu, Zhan-Bo
[1
]
Tu, Xiao-Jie
[2
]
Chai, Xin-Sheng
[3
]
Chen, Gang
[3
]
机构:
[1] Guangxi Univ, Sch Resources Environm & Mat, Nanning, Peoples R China
[2] Guangxi Univ, Sch Forest, Nanning, Peoples R China
[3] South China Univ Technol, State Key Lab Pulp & Paper Engn, Guangzhou, Peoples R China
关键词:
Headspace;
Gas chromatography;
Hexavalent chromium;
Solid waste leachates;
Potassium hydrogen oxalate buffer solution;
CR(VI);
REMOVAL;
SOIL;
SPECTROMETRY;
OXIDATION;
CR(III);
OXALATE;
RICE;
D O I:
10.1016/j.microc.2019.104494
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
This article reported a new headspace gas chromatography (HS-GC) for the determination of hexavalent chromium Cr(VI) in solid waste hypochlorite treated leachates. The method was based on the reaction between Cr (VI) and an excess potassium hydrogen oxalate in a pH buffer solution in the headspace vial at 60 degrees C, followed by HS-GC measuring the carbon dioxide produced. To eliminate the interference from hypochlorite in the leachates, two reaction times, i.e., 6 and 7 min, were selected. At pH = 1.50, it was found that the reaction of oxalate with Cr(VI) was partly achieved in 6 min and fully achieved in 7 min, and the reaction of oxalate with hypochlorite was achieved in 6 min. Based on two separate HS-GC measurement on the CO2 signals at 6 and 7, respectively, the amount of Cr(VI) in the sample can be calculated. The results showed that the present method has a good measurement precision (RSD < 4.70%) and accuracy (the recoveries were in the range of 92.5% - 107% and the relative differences were within 8.66% when compared with the data measured by the reference method). The present method can become a valuable tool in the Cr(VI) containing solid waste treatment related applications.
引用
收藏
页数:5
相关论文