Comparison of mandibular bone microarchitecture between micro-CT and CBCT images

被引:47
作者
Panmekiate, S. [1 ]
Ngonphloy, N. [1 ]
Charoenkarn, T. [1 ]
Faruangsaeng, T. [1 ]
Pauwels, R. [1 ,2 ]
机构
[1] Chulalongkorn Univ, Fac Dent, Dept Radiol, Bangkok, Thailand
[2] Katholieke Univ Leuven, Fac Med, Oral Imaging Ctr, OMFS IMPATH Res Grp,Dept Imaging & Pathol, Leuven, Belgium
关键词
cone-beam computed tomography; microcomputed tomography; bone; bone quality; computer-assisted image processing; BEAM COMPUTED-TOMOGRAPHY; TRABECULAR BONE; HISTOLOGICAL SECTIONS; FRACTAL DIMENSION; CANCELLOUS BONE; QUANTIFICATION; ARCHITECTURE; QUALITY;
D O I
10.1259/dmfr.20140322
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare microarchitecture parameters of bone samples scanned using micro-CT (RCT) to those obtained by using CBCT. Methods: A bone biopsy trephine bur (3 x 10 mm) was used to remove 20 cylindrical bone samples from 20 dry hemimandibles. Samples were scanned using mu CT (mu CT 35; SCANCO Medical, Briittisellen, Switzerland) with a voxel size of 20 mu m and CBCT (3D Accuitomo 170; J. Morita, Kyoto, Japan) with a voxel size of 80 mu m. All corresponding sample scans were aligned and cropped. Image analysis was carried out using BoneJ, including the following parameters: skeleton analysis, bone surface per total volume (BS/TV), bone volume per total volume (BV/TV), connectivity density, anisotropy, trabecular thickness and spacing, structure model index, plateness and fractal dimension. Pearson and Spearman correlation coefficients (R) were calculated. CBCT values were then calibrated using the slope of the linear fit with the mu CT values. The mean error after calibration was calculated and normalized to the standard deviation of the mu CT values. Results: R-values ranged between 0.05 (plateness) and 0.83 (BS/TV). Correlation was significant for both Spearman and Pearson's R for 8 out of 16 parameters. After calibration, the smallest normalized error was found for BV/TV (0.48). For other parameters, the error range was 0.58-2.10. Conclusions: Despite the overall correlation, this study demonstrates the uncertainty associated with using bone microarchitecture parameters on CBCT images. Although clinically relevant parameter ranges are not available, the errors found in this study may be too high for some parameters to be considered for clinical application.
引用
收藏
页数:7
相关论文
共 34 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]  
[Anonymous], P SIGGRAPH 87, DOI DOI 10.1145/37401.37422
[3]  
[Anonymous], 2002, FORMA
[4]   3D Reconstruction of Histological Sections: Application to Mammary Gland Tissue [J].
Arganda-Carreras, Ignacio ;
Fernandez-Gonzalez, Rodrigo ;
Munoz-Barrutia, Arrate ;
Ortiz-De-Solorzano, Carlos .
MICROSCOPY RESEARCH AND TECHNIQUE, 2010, 73 (11) :1019-1029
[5]   BoneJ Free and extensible bone image analysis in ImageJ [J].
Doube, Michael ;
Klosowski, Michal M. ;
Arganda-Carreras, Ignacio ;
Cordelieres, Fabrice P. ;
Dougherty, Robert P. ;
Jackson, Jonathan S. ;
Schmid, Benjamin ;
Hutchinson, John R. ;
Shefelbine, Sandra J. .
BONE, 2010, 47 (06) :1076-1079
[6]  
Dougherty R., 2007, MICROSC MICROANAL, V13, P1678, DOI [10.1017/S1431927607074430, DOI 10.1017/S1431927607074430, 10.1017/s1431927607074430]
[7]  
Drago C J, 1992, J Prosthodont, V1, P29, DOI 10.1111/j.1532-849X.1992.tb00423.x
[8]  
Engquist B, 1988, Int J Oral Maxillofac Implants, V3, P129
[9]  
Fazzalari NL, 1996, J PATHOL, V178, P100, DOI 10.1002/(SICI)1096-9896(199601)178:1<100::AID-PATH429>3.0.CO
[10]  
2-K