Ultrahigh-Magnitude Brain Magnetic Resonance Imaging Scan on Rhesus Monkeys With Implanted Deep Brain Stimulation Hardware

被引:1
作者
Li, Jun-Ju [1 ,2 ,3 ]
Shi, Lin [1 ,2 ]
Chen, Ying-Chuan [1 ,2 ]
Zhu, Guan-Yu [1 ,2 ]
Zhang, Jian-Guo [1 ,2 ]
机构
[1] Capital Med Univ, Beijing Tiantan Hosp, Dept Neurosurg, Beijing, Peoples R China
[2] Beijing Neurosurg Inst, Funct Neurosurg Lab, Beijing, Peoples R China
[3] Peoples Hosp Hainan Prov, Dept Neurosurg, Haikou, Hainan, Peoples R China
来源
NEUROMODULATION | 2018年 / 21卷 / 02期
基金
中国国家自然科学基金;
关键词
Deep brain stimulation; heating injury; magnetic resonance imaging; rhesus monkey; ultrahigh; EPILEPTIC MONKEYS; MRI; ELECTRODES; NUCLEUS; DISEASE; COIL; SAR;
D O I
10.1111/ner.12735
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
BackgroundPatients with implanted deep brain stimulation (DBS) hardware are prohibited from undergoing magnetic resonance imaging (MRI) scans at magnitudes greater than 1.5 T to avoid potential MRI-related heating injury. Whether DBS devices are compatible with higher field MRI scanning is unknown. This study aimed to investigate whether 7.0 T and 3.0 T MRI scans can be safely performed on rhesus monkeys with implanted DBS devices. MethodsEight male rhesus monkeys were included in this study and stereotactically implanted with DBS devices in the left anterior thalamus. Two weeks after DBS device implantation, 7.0 T and 3.0 T MRI scans were performed. The monkeys were observed for 72 hours. After explantation of the DBS system, 7.0 T MRI was repeated to determine potential lesions. Hematoxylin and eosin staining and transmission electron microscopy were conducted to assess pathological alterations. ResultsIn both groups, the monkeys exhibited no behavioral changes related to neurological deficits. Post-explantation MRI showed no malacia foci surrounding the DBS tracks. Additionally, neither hematoxylin and eosin staining nor transmission electron microscopy showed clear injury near the DBS leads. ConclusionThese results indicate that no obvious heating injury was induced in the tissue surrounding the DBS leads by the 7.0 T and 3.0 T MRI scans. Although the results of this study may not be generalizable, these data suggest that patients with implanted DBS devices can undergo even 7.0 T MRI without risk of brain injury.
引用
收藏
页码:168 / 175
页数:8
相关论文
共 21 条
[1]   Ultra-high magnetic resonance imaging (MRI): a potential examination for deep brain stimulation devices and the limitation study concerning MRI-related heating injury [J].
Chen, Ying-Chuan ;
Li, Jun-Ju ;
Zhu, Guan-Yu ;
Shi, Lin ;
Yang, An-Chao ;
Jiang, Yin ;
Zhang, Xin ;
Zhang, Jian-Guo .
NEUROLOGICAL SCIENCES, 2017, 38 (03) :485-488
[2]  
Finelli DA, 2002, AM J NEURORADIOL, V23, P1795
[3]   Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loops [J].
Golestanirad, Laleh ;
Angelone, Leonardo M. ;
Iacono, Maria Ida ;
Katnani, Husam ;
Wald, Lawrence L. ;
Bonmassar, Giorgio .
MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (04) :1558-1565
[4]   Feasibility of Using Linearly Polarized Rotating Birdcage Transmitters and Close-Fitting Receive Arrays in MRI to Reduce SAR in the Vicinity of Deep Brain Simulation Implants [J].
Golestanirad, Laleh ;
Keil, Boris ;
Angelone, Leonardo M. ;
Bonmassar, Giorgio ;
Mareyam, Azma ;
Wald, Lawrence L. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (04) :1701-1712
[5]   Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants [J].
Golestanirad, Laleh ;
Iacono, Maria Ida ;
Keil, Boris ;
Angelone, Leonardo M. ;
Bonmassar, Giorgio ;
Fox, Michael D. ;
Herrington, Todd ;
Adalsteinsson, Elfar ;
LaPierre, Cristen ;
Mareyam, Azma ;
Wald, Lawrence L. .
NEUROIMAGE, 2017, 147 :577-588
[6]   Measurements of RF heating during 3.0-T MRI of a pig implanted with deep brain stimulator [J].
Gorny, Krzysztof R. ;
Presti, Michael F. ;
Goerss, Stephan J. ;
Hwang, Sun C. ;
Jang, Dong-Pyo ;
Kim, Inyong ;
Min, Hoon-Ki ;
Shu, Yunhong ;
Favazza, Christopher P. ;
Lee, Kendall H. ;
Bernstein, Matt A. .
MAGNETIC RESONANCE IMAGING, 2013, 31 (05) :783-788
[7]   MRI-Related Heating near Deep Brain Stimulation Electrodes: More Data Are Needed [J].
Gupte, Akshay A. ;
Shrivastava, Devashish ;
Spaniol, Maggie A. ;
Abosch, Aviva .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2011, 89 (03) :131-140
[8]   Prevention of mechanical failures in implanted spinal cord stimulation systems [J].
Henderson, JM ;
Oakley, JC ;
Schade, CM ;
Sasaki, J ;
Caraway, DL .
NEUROSURGERY, 2005, 57 (02) :414-414
[9]   Mechanisms of deep brain stimulation [J].
Herrington, Todd M. ;
Cheng, Jennifer J. ;
Eskandar, Emad N. .
JOURNAL OF NEUROPHYSIOLOGY, 2016, 115 (01) :19-38
[10]   Magnetic resonance imaging of implanted deep brain stimulators: Experience in a large series [J].
Larson, Paul S. ;
Richardson, R. Mark ;
Starr, Philip A. ;
Martin, Alastair J. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2008, 86 (02) :92-100