Validation and Cross-Reactivity Data for Fentanyl Analogs With the Immunalysis Fentanyl ELISA

被引:26
作者
Guerrieri, Davide [1 ]
Kjellqvist, Fanny [2 ]
Kronstrand, Robert [1 ,2 ]
Green, Henrik [1 ,2 ]
机构
[1] Linkoping Univ, Div Drug Res, Dept Med & Hlth Sci, SE-58185 Linkoping, Sweden
[2] Natl Board Forens Med, Dept Forens Genet & Forens Toxicol, Artillerigatan 12, SE-58758 Linkoping, Sweden
关键词
POSTMORTEM; RECEPTOR;
D O I
10.1093/jat/bky060
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Every year new fentanyl analog compounds, or fentanyls, appear on the drug scene. Development of immunoassays dedicated for screening individual molecules is challenging due to the short-lived presence of these compounds on the recreational drug market. Therefore, we investigated the detecting capabilities of the immunalysis fentanyl direct enzyme-linked immunosorbent assay (ELISA) kit against fentanyl in whole blood, and determined the cross-reactivity of nine fentanyl analogs (2-fluorofentanyl, acetylfentanyl, acrylfentanyl, carfentanil, cyclopropylfentanyl, tetrahydrofuranylfentanyl, furanylfentanyl, ocfentanil, valerylfentanyl) to confirm its validity for the general screening of fentanyls. Immunalysis ELISA assay was used to test whole blood samples fortified with fentanyl on a TECAN Freedom EVOlyzer platform, according to manufacturer specifications. The kit successfully was validated for fentanyl screening with a cutoff set at 0.5 ng/mL, and all tested analogs, with the exclusion of carfentanil, were detected. The lowest cross-reactivity with the kit was obtained with furanylfentanyl (20% +/- 1, 95% confidence intervals (CI)) and 4-fluoroisobutyrfentanyl (25% +/- 1, 95% CI), while the highest was recorded using acetylfentanyl (99% +/- 11, 95% CI) and acrylfentanyl (94% +/- 10, 95% CI). Post-mortem samples containing fentanyl, acrylfentanyl, cyclopropylfentanyl, THF-fentanyl and 4-fluoroisobutyrfentanyl were screened, and sensitivity and specificity of each analog were calculated. Positive screening results were generated by all post-mortem cases containing fentanyl (n = 14), acrylfentanyl (n = 11), cyclopropylfentanyl (n = 14), tetrahydrofuranylfentanyl (n = 13) and 4-fluoroisobutyrfentanyl (n = 10). Concentration of post-mortem fentanyl samples ranged from 0.5 ng/mL (cutoff) to 230 ng/mL, while the range for analogs was 3.4-36 ng/mL (cyclopentylfentanyl), 0.76-370 ng/mL (4-fluoroisobutyrfentanyl), 0.02-12 ng/mL (acrylfentanyl) and 2-26 ng/mL (tetrahydrofuranylfentanyl). The immunalysis fentanyl direct ELISA kit was successfully validated and showed significant cross-reactivity for all tested fentanyls, except carfentanil, making it a suitable technique for fentanyl and fentanyl analogs screening.
引用
收藏
页码:18 / 24
页数:7
相关论文
共 20 条
[1]  
[Anonymous], 2015, European Drug Report - Trends and Developments
[2]  
[Anonymous], PRINCIPLES FORENSIC
[3]  
[Anonymous], 2012, US PHARM CONV
[4]  
EMEA, 2015, EMEACHMPEWP192217200
[5]   OPIATE RECEPTOR - MODEL EXPLAINING STRUCTURE-ACTIVITY-RELATIONSHIPS OF OPIATE AGONISTS AND ANTAGONISTS [J].
FEINBERG, AP ;
CREESE, I ;
SNYDER, SH .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1976, 73 (11) :4215-4219
[6]   Acetyl Fentanyl Toxicity: Two Case Reports [J].
Fort, Chelsea ;
Curtis, Byron ;
Nichols, Clay ;
Niblo, Cheryl .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2016, 40 (09) :754-757
[7]   Acrylfentanyl: Another new psychoactive drug with fatal consequences [J].
Guerrieri, Davide ;
Rapp, Emma ;
Roman, Markus ;
Thelander, Gunilla ;
Kronstrand, Robert .
FORENSIC SCIENCE INTERNATIONAL, 2017, 277 :E21-E29
[8]   Postmortem and Toxicological Findings in a Series of Furanylfentanyl-Related Deaths [J].
Guerrieri, Davide ;
Rapp, Emma ;
Roman, Markus ;
Druid, Henrik ;
Kronstrand, Robert .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2017, 41 (03) :242-249
[9]   An Acute Acetyl Fentanyl Fatality: A Case Report With Postmortem Concentrations [J].
McIntyre, Iain M. ;
Trochta, Amber ;
Gary, Ray D. ;
Malamatos, Mark ;
Lucas, Jonathan R. .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2015, 39 (06) :490-494
[10]   Equianalgesic dose ratios for opioids: A critical review and proposals for long-term dosing [J].
Pereira, J ;
Lawlor, P ;
Vigano, A ;
Dorgan, M ;
Bruera, E .
JOURNAL OF PAIN AND SYMPTOM MANAGEMENT, 2001, 22 (02) :672-687